Peter Huber Kältemaschinenbau Unistat P925w Dynamisches
Peter Huber Kältemaschinenbau Unistat P925w Dynamisches — Temperature control units.
The Unistat P925w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. Its 12 kW heating and staged cascade refrigeration (R-452A/R-23) deliver 16 kW at -20 °C falling to 1.8 kW at -90 °C. It serves closed-loop process temperature control where high cooling demand at low temperature and 191 l/min pump flow are required.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P925w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Expansion volume | 29 l |
| Filling volume | 12 l (min) |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Weight | 994 kg (net) |
| Current consumption | 52 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1300 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing, water-cooled (w suffix) Unistat with a two-stage cascade refrigeration circuit: R-452A (5 kg, GWP 2141) in the first stage and R-23 (2.2 kg, GWP 14800) in the second.
- Magnetically coupled stainless-steel circulation pump, speed-adjustable, delivering up to 191 l/min and 5.5 bar; 22 l/min still available at 5.0 bar.
- Safety classification III / FL with adjustable overtemperature protection per DIN 12876; Pilot ONE controller with POKO ONE potential-free contact output.
- Within the Dynamisches series, the P925w is the largest water-cooled variant listed; siblings 705w, 815w, 1005w and 1015w cover lower capacities. The P925w is distinguished by its 12 kW heating capacity and 52 A maximum current draw.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 16 kW at -20 °C, 3.5 kW at -80 °C and 1.8 kW at -90 °C, so the unit must be sized against the actual heat load at the lowest required process temperature.
- The heat transfer fluid must have a kinematic viscosity no higher than 50 mm²/s; at low temperatures fluid selection is limited by viscosity rise.
- The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used to avoid corrosion and fouling.
- Cooling water consumption is 1300 l/h at 15 °C water and 0 °C flow, with a minimum differential pressure of 1 bar and maximum pressure of 6 bar.
Installation, connection and control
- Requires a 400 V 3~ 50 Hz supply with maximum current 52 A and 3x63 A fusing; three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be sourced on site.
- Leakage current exceeds 3.5 mA, requiring attention to protective earthing and RCD selection.
- Pump connection is M38x1.5 male; cooling water connection is G3/4 male.
- Ambient temperature must be 5–40 °C; protection rating IP20 restricts installation to dry indoor areas.
- Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces; an external Pt100 sensor connection is available for process control.
- The unit must not be installed or started within an ATEX zone; use with medical devices or direct foodstuff temperature control is not permissible.
Integration & Operating Notes
The 400 V 3~ 50 Hz supply must be dedicated and rated for the 52 A maximum current; the 3x63 A fuse rating defines upstream breaker and cable cross-section. The unit is delivered with cable but without plug, so the plug and any hard-wired connection must be provided by an electrician on site.
The cooling water circuit requires a minimum differential pressure of 1 bar and must not exceed 6 bar; the specified flow is 1300 l/h at 15 °C water and 0 °C flow. Only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM may be used, and pipes must be insulated if water temperature falls below the condensation point.
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s. At low process temperatures, viscosity rises and may exceed this limit, restricting fluid choice and potentially affecting pump performance and heat transfer.
The unit must only be transported and stored upright; horizontal transport damages the compressor. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.
In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the safety function will not operate as intended.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Selection & Integration Q&A
Choose the P925w when the process requires up to 12 kW heating and cooling capacity down to -90 °C with high pump flow (191 l/min, 5.5 bar). The 705w and 815w have lower capacities; select them only if the heat load and pressure drop are within their smaller envelopes.
Cooling water: 5–20 °C, 1300 l/h at 15 °C, minimum 1 bar differential pressure, maximum 6 bar, G3/4 male connection. Process loop: M38x1.5 male pump connection. Ensure water compatibility with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors only (IP20), ambient 5–40 °C, upright transport and storage, leveling feet activated, 2-hour acclimation after temperature changes. Do not install in an ATEX zone. A drip tray for condensation water/heat transfer fluid must be installed below the unit.
The unit is delivered with cable but without plug; a suitable plug must be sourced. Periodic checks of hoses and connections for leaks and quality are required, with replacement as needed. Cooling water supply should be shut off during downtimes. Refrigerant-circuit work must be by approved contractors.