Peter Huber Kältemaschinenbau Unistat P950w Dynamisches
Peter Huber Kältemaschinenbau Unistat P950w Dynamisches — Temperature control units.
The Unistat P950w Dynamisches is a water-cooled, free-standing temperature control unit with a 36 kW heating capacity and a -90 to 200 °C working range, driven by a magnetically coupled stainless-steel pump and Pilot ONE controller. Its dynamic modification targets high-pressure-drop circuits, delivering up to 260 l/min at 4.8 bar through DN40 PN10/40 flanges. It serves as a high-capacity heat-transfer source for externally closed industrial processes requiring deep cooling and rapid temperature transitions.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P950w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 90 l |
| Filling volume | 30 l (min) |
| Heating capacity | 36 kW |
| Pump connection | flange DN40 PN10/40 EN1092-1 |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1 1/4 male |
| Cooling water consumption | 5000 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The P950w is a water-cooled (w suffix) free-standing Unistat with a two-stage refrigeration circuit using R-507 and R-23. Within the Unistate Bis 120C series, the Dynamisches modification is distinguished by its high-delivery pump (260 l/min, 4.8 bar max) and DN40 PN10/40 pump flanges, making it suitable for circuits with significant pressure drop. Sibling models such as the Unistat 705w, 815w, 1005w and 1015w are air-cooled or water-cooled variants with different capacities; the P950w is the largest water-cooled dynamic unit in this set. The unit measures 2630 x 1300 x 1980 mm and requires a 400 V 3~ 50 Hz supply.
Behavior under load and media
Cooling capacity is strongly temperature-dependent: 36 kW at -40 °C, 25 kW at -60 °C and 10 kW at -80 °C, so process heat loads must be checked against the actual operating temperature. The pump curve shows flow falling as pressure rises, with a maximum of 260 l/min at low pressure and 4.8 bar at low flow; external circuit pressure drop must be matched to avoid insufficient flow. The heat transfer fluid must have a kinematic viscosity no higher than 100 mm²/s at the operating temperature. The unit has a minimum filling volume of 30 l and a 90 l expansion tank. The temperature range is -90 to 200 °C, with a display resolution of 0.01 K.
Installation, control and operation
The Pilot ONE controller with POKO ONE control output provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 connection for an external sensor. The cooling water circuit requires a supply of 5–20 °C at 3–6 bar with a minimum differential pressure of 3 bar; consumption is 5000 l/h at 15 °C and 0 °C flow. The unit has IP20 protection and must be installed indoors at 5–40 °C ambient. The electrical supply must be 400 V 3~ 50 Hz with ±5 % voltage and ±2 % frequency tolerance, but the combination of -5 % voltage and +2 % frequency is not allowed. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this. Three-phase devices under 63 A are delivered with cable but without a plug. The unit must not be installed in an ATEX zone. For externally closed applications, the expansion vessel must remain unlocked; for externally open applications, it must be locked. A drip tray must be installed below the unit. The refrigerant charge includes R-23 with 103.6 t CO2 equivalent, triggering F-gas leak monitoring obligations.
Integration & Operating Notes
Insufficient cooling water pressure prevents the water-cooled refrigeration machine from operating correctly. The site cooling water system must be verified before commissioning. Cooling water pipes must be insulated if the water temperature can fall below the condensation point. Shut off cooling water during downtimes, even overnight.
In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger. This reduces risk from exothermic conditions such as glass breakage or uncontrolled phase transitions.
Selection & Integration Q&A
Choose the P950w when you need the highest cooling capacity in the water-cooled dynamic series (36 kW at -40 °C) and a high-delivery pump (260 l/min, 4.8 bar) for circuits with significant pressure drop. Smaller siblings like the 815w or 1015w have lower capacities and are better suited to lower heat loads or where space and power are limited.
Cooling water must be supplied at 5–20 °C, 3–6 bar, with a minimum differential pressure of 3 bar and a flow of 5000 l/h at 15 °C. Process piping must mate with DN40 PN10/40 EN1092-1 flanges on the pump. The heat transfer fluid must have a viscosity ≤100 mm²/s at operating temperature.
Install indoors at 5–40 °C ambient, IP20, not in an ATEX zone. Allow 2 hours acclimation after moving between cold and warm environments before power-on. Transport upright only and empty. Leveling feet must be activated and casters are not for transport. Shut off cooling water during downtimes.
Refrigerant-circuit work requires approved contractors; F-gas leak checks apply due to 103.6 t CO2 equivalent. Hoses and connections must be periodically checked and replaced as needed. The base delivery includes cable but no plug for three-phase units under 63 A. A drip tray must be installed by the customer. Heat transfer fluid is not included and must be selected per viscosity and fire-point limits.