Peter Huber Kältemaschinenbau Unistat 930w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 930w Dynamisches

Peter Huber Kältemaschinenbau Unistat 930w Dynamisches — Temperature control units.

The Unistat 930w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. Within the Unistat w series it is the high-capacity 9xx frame, distinguished by 32 kW heating and 66 A three-phase draw. It serves closed-loop process temperature control where high heating power and deep cooling are required.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 930w Dynamisches Current —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 29 l
Filling volume 12 l
Heating capacity 32 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...200 °C
Dimensions, Weight & Mechanical Design
Weight 984 kg
Electrical, Power & Drive Specifications
Current consumption 66 A (max.)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1300 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Power cable and plug are not supplied above 63 A

With a maximum current of 66 A the unit ships without cable and without plug; the customer must procure and terminate the three-phase feeder and provide external branch circuit protection according to UL 489. Panel sizing must account for 3x80 A fusing.

Unistat 930w Technical Data, page 2
Electrical
Leakage current above 3.5 mA constrains RCD choice

The datasheet states leakage current exceeds 3.5 mA, so a standard 30 mA residual current device may be unsuitable; the earthing arrangement and protective device selection must be engineered accordingly before commissioning.

Unistat 930w Technical Data, page 2
Cooling & Media
Cooling water must stay within 5–20 °C and 1–6 bar

Rated cooling output is quoted at 15 °C cooling water inlet and 1 bar differential pressure; intake temperature must remain between 5 and 20 °C and pressure must not exceed 6 bar. If the water can fall below the condensation point, the cooling water pipes must be insulated.

Unistat 930w Technical Data, page 2
Process Limits
Heat transfer fluid viscosity must not exceed 50 mm²/s

Fluid selection must keep kinematic viscosity at or below 50 mm²/s across the whole operating range, otherwise pump delivery and cooling performance suffer. For FL-rated fluids the fire point must be at least 65 °C, and ethanol use is permitted only under supervision.

Unistat 930w Technical Data, page 2; Operation Manual, Section 1.7
Mounting
Expansion vessel lock state depends on application type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. Verify the isolating sleeve or shut-off valve position before filling.

Unistat Operation Manual, page 33
Mounting
Unit must not be installed inside an ATEX zone

Without an Ex px cabinet the unit is not protected against explosion; installation and start-up within an ATEX zone are prohibited. Site selection must exclude hazardous areas unless the separate ATEX annex and Ex px cabinet rating plate apply.

Unistat Operation Manual, Section 1.3.3 / page 15-16
Maintenance
Shut off cooling water even for overnight downtime

The manual directs shutting off the cooling water supply even for shorter downtimes such as overnight, and periodic checks of hoses and hose connections for leaks and quality with replacement as required. Refrigerant-circuit work must be performed by approved refrigeration/air-conditioning contractors.

Unistat Operation Manual, page 29; Section 1.4.2 / page 19
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Power-interruption behavior is selectable between Off/Standby and automatic restart.

Unistat Operation Manual, page 19-21

Selection & Integration Q&A

When should I choose the Unistat 930w over the 815w or 1005w siblings?

Pick the 930w when 32 kW heating and the 9xx frame capacity are needed with water cooling; the 815w is a smaller frame, while the 1005w/1015w are larger. All share the 'Dynamisches' dynamic control concept, so selection turns on heating power and frame size rather than control architecture.

What upstream electrical and cooling-water provisions must be in place before the 930w can be commissioned?

A 460 V 3~ 60 Hz feeder sized for 66 A with 3x80 A fusing and UL 489 branch protection, plus a customer-supplied power cable and plug (not delivered above 63 A). Cooling water at G3/4 male with 1–6 bar differential pressure, 5–20 °C intake and 1300 l/h at 15 °C.

What site and duty-cycle limits apply to installing the 930w?

Ambient 5–40 °C, IP20, no ATEX zone, upright transport only, leveling feet and parking brakes set before start-up, and about 2 hours acclimation after a cold-to-warm move. A drip tray for condensation water or heat transfer fluid must be installed below the unit.

What maintenance items and accessories are not included in the base delivery?

The power cable and plug are excluded above 63 A, as are heat transfer fluid, cooling-water hoses and DIN 3863/3870 fittings, and the optional gas detection sensor and processing unit. Periodic tasks include hose and connection inspection, liquefier fin cleaning, strainer cleaning and fluid replacement.