Peter Huber Kältemaschinenbau Unistat P815w Dynamisches

Peter Huber Kältemaschinenbau Unistat P815w Dynamisches — Temperature control units.

The Unistat P815w Dynamisches is a water-cooled, magnetically coupled stainless-steel pump temperature control unit spanning -85 to 250 °C with 2 kW heating, driven by the Pilot ONE controller. Its 'P' designation targets applications with high pressure drop, distinguishing it from the 815w Dynamisches sibling. It serves precise dynamic temperature control of externally closed circuits in process and laboratory systems.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P815w Dynamisches Current —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 2 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 231 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 130 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load

Control, installation and operation

Integration & Operating Notes

Electrical
Supply tolerance combinations are restricted, not independent

The ±5 % voltage and ±2 % frequency tolerances may not be combined freely: -5 % voltage with +2 % frequency is explicitly not allowed. Verify the site supply against this matrix before connection, because a compliant-looking voltage alone does not guarantee permissible operation.

Unistat P815w Technical Data, page 2
Electrical
Provide plug and RCD sized for >3.5 mA leakage current

Three-phase units below 63 A are delivered with a cable but without a plug, so the customer must supply and install it. Because leakage current exceeds 3.5 mA, a suitable residual current device and protective earthing must be provided; the electrician must confirm the rating-plate voltage and frequency match.

Unistat P815w Technical Data, page 2; Unistat Operation Manual, page 37
Cooling & Media
Cooling water must stay within 1–6 bar and 5–20 °C

The cooling water circuit requires a minimum differential pressure of 1 bar and tolerates a maximum of 6 bar, with intake temperature between 5 °C and 20 °C. Higher intake temperatures reduce cooling power and increase water flow, so facility water conditions directly cap achievable performance.

Unistat P815w Technical Data, page 2
Mounting
Never install or start inside an ATEX zone

Without an Ex px cabinet the unit is not protected against explosion, so site selection must exclude ATEX zones. Ex px operation requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.

Unistat Operation Manual, Section 1.3.3 / pages 15-16
Mounting
Transport upright, emptied, and acclimate 2 hours before power-on

Horizontal transport damages the compressor, and only an emptied unit may be moved to avoid thermal fluid overflow. Casters are not for transport and carry only 25 % of total mass symmetrically; use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on.

Unistat Operation Manual, pages 22 and 29
Process Limits
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. The wrong state in either direction produces damage or fluid loss, so the application type must be fixed before commissioning.

Unistat Operation Manual, page 33
Control & I/O
Process Safety mode needs the compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles, which requires the compressor to be set to Permanently ON. Configure this deliberately when protecting exothermic processes such as glass reactors.

Unistat Operation Manual, page 19
Maintenance
Refrigerant work is restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel cannot service the refrigerant circuit. The refrigerant charge carries a reportable CO2 equivalent (about 7.8 t total for R-452A and R-23), which must be managed under F-gas rules.

Unistat Operation Manual, Section 1.4.2 / page 19; Unistat P815w Technical Data, page 1

Selection & Integration Q&A

When should I choose the P815w Dynamisches rather than the plain 815w Dynamisches or a 705w/1005w Dynamisches?

Pick the P815w when the external circuit has high pressure drop: the 'P' modification is intended for applications with high pressure drop, and the pump still delivers 28 l/min at 2.5 bar. The plain 815w suits low-resistance circuits; 705w/1005w differ in class and capacity.

What must the facility provide for cooling water and power to integrate this unit?

Cooling water at 5–20 °C with 1–6 bar differential pressure through a G1/2 male connection (130 l/h at 15 °C water, 0 °C flow), and a 400 V 3~ 50 Hz supply fused at 3x16 A. The unit ships with cable but no plug; leakage current >3.5 mA requires a suitable RCD and earthing.

What mounting and duty-cycle limits apply on site?

Ambient must stay 5–40 °C, ATEX zones are excluded, and the unit must stand upright on activated leveling feet with casters braked. Allow 2 h acclimation after cold/warm moves. Shut off cooling water even overnight; the expansion vessel lock state must match open vs closed application.

What maintenance and accessories are needed beyond the base delivery?

The customer supplies the three-phase plug and a drip tray below the unit. Periodic checks cover hoses and connections, liquefier fins, the hat-type strainer/dirt trap, emergency-stop function test and heat transfer fluid inspection/replacement. Refrigerant-circuit work requires approved contractors.