Peter Huber Kältemaschinenbau Unistat 920w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 920w Dynamisches

Peter Huber Kältemaschinenbau Unistat 920w Dynamisches — Temperature control units with Noise level 71 dB.

The Unistat 920w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. It is the 920-size modification of the Unistat w series, distinguished by 12 kW heating and 42 A max current on a 400 V three-phase supply. It serves closed-loop process temperature control where deep-cold capacity is needed only at low load.

Noise level
71 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 920w Dynamisches Current 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Noise level 71 dB(A) (sound pressure level)
Control output POKO ONE (potential-free contact)
Expansion volume 29 l
Filling volume 12 l (min)
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...200 °C
Dimensions, Weight & Mechanical Design
Weight 893 kg (net)
Electrical, Power & Drive Specifications
Current consumption 42 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be 400 V three-phase with 63 A protection

The 42 A maximum current consumption requires a 63 A fuse and a three-phase feed sized for it. The unit is delivered with cable but without a plug, so the mating plug and upstream protection must be provided on site before commissioning.

Unistat 920w Technical Data, page 2
Cooling & Media
Cooling water must stay within 5–20 °C and 1–6 bar

Below 1 bar differential the rated cooling performance is not reached; above 6 bar the circuit is at risk. Intake temperature above 20 °C or below 5 °C invalidates the design basis, and pipes must be insulated where condensation can form.

Unistat 920w Technical Data, page 2
Process Limits
Deep-cold duty is limited to about 1 kW at -90 °C

Cooling power drops from 11 kW at -20 °C to 1 kW at -90 °C. Process loads at the cold end must be sized against this curve, otherwise the setpoint will not be held and the unit will run continuously at maximum output.

Unistat 920w Technical Data, page 1
Mounting
Transport upright and emptied; casters are not for moving

Horizontal transport damages the compressor, and a filled unit risks thermal fluid overflow. Casters carry only 25% of total mass symmetrically and must not be used for relocation; use an industrial truck and engage parking brakes or leveling feet before start-up.

Unistat Operation Manual, page 22
Process Limits
Expansion vessel lock state depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open bath applications it must be locked to prevent fluid loss. The wrong state at commissioning is a direct damage risk.

Unistat Operation Manual, page 33
Control & I/O
Process Safety needs the compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling at the overtemperature trigger instead of switching off all poles. This only works if the compressor is configured to Permanently ON; otherwise the protective strategy does not deliver cooling when triggered.

Unistat Operation Manual, page 19
Control & I/O
Decide restart behavior after power failure before handover

Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active, and Power Failure Auto resumes temperature control automatically if it was active. Choose according to whether unattended restart is acceptable for the process.

Unistat Operation Manual, page 20
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be closed for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Hoses and connections must be checked periodically for leaks and quality and replaced as required.

Unistat Operation Manual, page 29
Maintenance
Refrigerant circuit work requires approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The circuit is factory-tested to 10-4 Pa m3/s, but general maintenance personnel may not open it; plan a service contract accordingly.

Unistat Operation Manual, Section 1.4.2, page 19
Mounting
No installation inside an ATEX zone

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet with its own annex, and only the model/serial number on the Ex px rating plate is suitable. Site selection must exclude hazardous areas.

Unistat Operation Manual, Section 1.3.3, page 15-16

Selection & Integration Q&A

When should I choose the Unistat 920w Dynamisches rather than the 815w or 1005w siblings?

Pick the 920w when 12 kW heating and 42 A on 400 V three-phase match your load and you need the 920 frame. The 815w is quieter at 60 dB(A) versus 71 dB(A) here; the 1005w is a larger neighbor for higher duty. Confirm cold-end load against the 1 kW at -90 °C figure before sizing up or down.

What must be provided on site for the electrical and cooling-water connections?

A 400 V 3~ 50 Hz feed with 63 A fuse and a mating plug, since the unit ships with cable but no plug. Cooling water at 5–20 °C, 1–6 bar differential, connected to G3/4 male. Pump ports are M38x1.5 male with DIN 3863 ball-socket geometry, so matching hoses and fittings must be ordered.

What mounting and duty-cycle limits does the manual impose on this unit?

Transport upright and emptied only; casters carry 25% of mass and are not for moving. Leveling feet or parking brakes must be set before start-up, ambient must stay 5–40 °C, and about 2 hours acclimation is required after a cold/warm move. A drip tray must be installed below the unit.

What maintenance and consumables are not included in base delivery?

Heat transfer fluid (minimum 12 l fill, 29 l expansion), cooling water hoses and pump fittings, the power plug, and a drip tray are not included. Periodic hose and connection inspection, cooling-water shutoff for overnight downtime, and refrigerant-circuit service by approved contractors must be planned separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 920w Technical Data
Datasheet ·31,4 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF