Peter Huber Kältemaschinenbau Unistat 920w Dynamisches
Peter Huber Kältemaschinenbau Unistat 920w Dynamisches — Temperature control units with Noise level 71 dB.
The Unistat 920w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. It is the 920-size modification of the Unistat w series, distinguished by 12 kW heating and 42 A max current on a 400 V three-phase supply. It serves closed-loop process temperature control where deep-cold capacity is needed only at low load.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 920w Dynamisches Current | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Noise level | 71 dB(A) (sound pressure level) |
| Control output | POKO ONE (potential-free contact) |
| Expansion volume | 29 l |
| Filling volume | 12 l (min) |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Weight | 893 kg (net) |
| Current consumption | 42 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s (max permissible) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing, water-cooled (w) Unistat with 893 kg net weight and 954x1504x1622 mm footprint; leveling feet and casters are for positioning only, not transport.
- Magnetically coupled stainless-steel pump with M38x1.5 male connections and adjustable speed; 29 l expansion volume and 12 l minimum filling.
- Sits above the 815w and 705w siblings in the w series; the 1005w and 1015w are larger neighbors. The 815w is quoted at 60 dB(A) versus 71 dB(A) here, so acoustic planning differs across the series.
- Safety classification III / FL with adjustable overtemperature protection per DIN 12876; refrigerant R-1270 (A3, H220) requires flammable-refrigerant site assessment.
Behavior under load and media
- Temperature range -90...200 °C with 12 kW heating. Cooling power falls steeply at the cold end (11 kW at -20 °C, 1 kW at -90 °C), so deep-cold process loads must be small.
- Heat transfer fluid limited to 50 mm²/s maximum kinematic viscosity; FL-rated fluids need a fire point ≥65 °C, and ethanol use is supervised-only.
- Cooling water must be 5...20 °C intake, 1 bar minimum differential and 6 bar maximum; circuit materials include copper, 1.4401 stainless, MS, PA, PPE, PTFE and EPDM, so water chemistry must be compatible.
- Process Safety mode releases maximum cooling at the OT trigger instead of switching off all poles; it requires the compressor set to Permanently ON.
Installation, connection and control
- 400 V 3~ 50 Hz supply, 42 A max current, 63 A fuse; three-phase units below 63 A arrive with cable but without plug. Voltage tolerance ±5% only with simultaneous ±2% frequency tolerance.
- Pilot ONE controller with Ethernet, USB host/device and RS232; POKO ONE potential-free contact output; external Pt100 sensor input; display resolution 0.01 K.
- Cooling water connection G3/4 male; pump connection M38x1.5 male with DIN 3863 ball-socket geometry.
- Ambient 5...40 °C, IP20, 71 dB(A). Acclimation of about 2 hours is required after moving between cold and warm environments before power-on.
Integration & Operating Notes
The 42 A maximum current consumption requires a 63 A fuse and a three-phase feed sized for it. The unit is delivered with cable but without a plug, so the mating plug and upstream protection must be provided on site before commissioning.
Below 1 bar differential the rated cooling performance is not reached; above 6 bar the circuit is at risk. Intake temperature above 20 °C or below 5 °C invalidates the design basis, and pipes must be insulated where condensation can form.
Cooling power drops from 11 kW at -20 °C to 1 kW at -90 °C. Process loads at the cold end must be sized against this curve, otherwise the setpoint will not be held and the unit will run continuously at maximum output.
Horizontal transport damages the compressor, and a filled unit risks thermal fluid overflow. Casters carry only 25% of total mass symmetrically and must not be used for relocation; use an industrial truck and engage parking brakes or leveling feet before start-up.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open bath applications it must be locked to prevent fluid loss. The wrong state at commissioning is a direct damage risk.
In Process Safety mode the controller releases maximum cooling at the overtemperature trigger instead of switching off all poles. This only works if the compressor is configured to Permanently ON; otherwise the protective strategy does not deliver cooling when triggered.
Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active, and Power Failure Auto resumes temperature control automatically if it was active. Choose according to whether unattended restart is acceptable for the process.
The cooling water supply should be closed for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Hoses and connections must be checked periodically for leaks and quality and replaced as required.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The circuit is factory-tested to 10-4 Pa m3/s, but general maintenance personnel may not open it; plan a service contract accordingly.
The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet with its own annex, and only the model/serial number on the Ex px rating plate is suitable. Site selection must exclude hazardous areas.
Selection & Integration Q&A
Pick the 920w when 12 kW heating and 42 A on 400 V three-phase match your load and you need the 920 frame. The 815w is quieter at 60 dB(A) versus 71 dB(A) here; the 1005w is a larger neighbor for higher duty. Confirm cold-end load against the 1 kW at -90 °C figure before sizing up or down.
A 400 V 3~ 50 Hz feed with 63 A fuse and a mating plug, since the unit ships with cable but no plug. Cooling water at 5–20 °C, 1–6 bar differential, connected to G3/4 male. Pump ports are M38x1.5 male with DIN 3863 ball-socket geometry, so matching hoses and fittings must be ordered.
Transport upright and emptied only; casters carry 25% of mass and are not for moving. Leveling feet or parking brakes must be set before start-up, ambient must stay 5–40 °C, and about 2 hours acclimation is required after a cold/warm move. A drip tray must be installed below the unit.
Heat transfer fluid (minimum 12 l fill, 29 l expansion), cooling water hoses and pump fittings, the power plug, and a drip tray are not included. Periodic hose and connection inspection, cooling-water shutoff for overnight downtime, and refrigerant-circuit service by approved contractors must be planned separately.