Peter Huber Kältemaschinenbau Unistat P905 Dynamisches
Peter Huber Kältemaschinenbau Unistat P905 Dynamisches — Temperature control units.
The Unistat P905 Dynamisches is a free-standing refrigerated heating circulator with a magnetically coupled stainless-steel high-discharge-pressure pump, built for externally closed or open loops with high pressure drop. Its -90 to 250 °C range and 6 kW heating stage distinguish it from the w-suffixed water-cooled siblings in the Unistat 5xx/7xx/8xx/10x5 family. It serves flow-through chemistry and semiconductor process temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P905 Dynamisches Current | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...250 °C |
| Weight | 272 kg |
| Current consumption | 19 A (max) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...35 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled cascade refrigeration (R-452A 2.5 kg, R-23 second stage 0.75 kg) in a 272 kg free-standing frame; IP20, safety class III/FL.
- The 'P' designation denotes the high-discharge-pressure pump variant, intended for high pressure-drop circuits such as flow-through chemistry and semiconductor applications; the 'w' siblings (705w, 815w, 1005w, 1015w) are water-cooled and do not share this pump emphasis.
- Pump: magnetically coupled, stainless steel, speed-adjustable, M30x1.5 male connection.
- Controller: Pilot ONE with POKO ONE control output; 0.01 K display resolution; Ethernet, USB host/device and RS232 interfaces; external Pt100 sensor input.
Behavior under load and media
- Working range -90 to 250 °C with 6 kW heating; cooling capacity is rated at 20 °C room temperature and falls as ambient rises.
- Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity; FL-rated fluids require a fire point ≥65 °C, and ethanol use is supervised-only.
- Process Safety mode releases maximum cooling at the OT trigger instead of tripping all poles, and requires the compressor set to Permanently ON.
Installation, connection and control
- 460 V 3~ 60 Hz supply, 19 A max, delivered with cable but without plug; external branch circuit protection to UL 489 required; leakage current >3.5 mA.
- Piping must use ball sockets per DIN 3863 and sleeve nuts per DIN 3870; expansion vessel unlocked for externally closed loops, locked for open bath applications.
- Power-interruption response selectable between Off/Standby and automatic restart; Power Failure Auto resumes an active process after outage.
Integration & Operating Notes
The 460 V three-phase unit ships with cable but no plug, so a matching plug must be installed on site, and the upstream panel must include branch circuit protection per UL 489. Because leakage current exceeds 3.5 mA, RCD selection and protective-earth bonding need review before energizing.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open bath applications it must be locked to prevent fluid loss. Confirm the loop type before commissioning, since the wrong setting risks equipment damage or fluid spill.
Rated cooling output assumes 20 °C ambient; above that the available cooling capacity drops, so warm rooms need derating or ventilation planning. Ambient must stay within 5 to 35 °C, and the fluid's kinematic viscosity must not exceed 50 mm²/s or circulation and heat transfer degrade.
The unit must be transported and stored upright only and moved only when emptied; horizontal transport damages the compressor. Casters are not for transport and carry only 25% of total mass. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation faults.
In Process Safety mode the controller releases maximum cooling at the OT trigger rather than switching off all poles, which protects exothermic processes. This mode only works if the compressor is configured to Permanently ON; otherwise the safety behavior is not delivered as intended.
The cascade holds R-452A (2.5 kg, GWP 2141, 5.4 t CO2e) and R-23 (0.75 kg, GWP 14800, 11.1 t CO2e). Any work on the coolant circuit must be performed by approved refrigeration/air-conditioning contractors, so service contracts and leak-check intervals should be planned accordingly.
Selection & Integration Q&A
Pick the P905 when your circuit has high pressure drop (flow-through chemistry, semiconductor) and you want air cooling without a cooling-water supply. Choose a 'w' sibling when water cooling is available and lower heat rejection to room air is required.
Piping must use ball sockets per DIN 3863 and sleeve nuts per DIN 3870 on the M30x1.5 male pump connection. The supply is 460 V 3~ 60 Hz, 19 A max, with a field-installed plug and UL 489 branch protection; leakage current exceeds 3.5 mA.
Install outside ATEX zones, within 5–35 °C ambient, on leveling feet with brakes engaged, and allow 2 h acclimation after temperature changes. Restart after power loss is selectable between Off/Standby and automatic resume via Power Failure Auto.
No plug, drip tray, or heat transfer fluid is supplied; the responsible body must install a drip tray. Periodic hose/connection inspection and fluid replacement are required, and refrigerant-circuit service must be done by approved contractors.