Peter Huber Kältemaschinenbau Unistat 525 Dynamisches
Peter Huber Kältemaschinenbau Unistat 525 Dynamisches — Temperature control units with Noise level 56 dB.
The Unistat 525 Dynamisches is a free-standing, air-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C. Within the Unistat 5xx Dynamisches series it is the non-water-cooled 525 variant, distinguished from the 525w by air cooling and a 56 dB(A) sound level. It serves closed-loop heating and cooling of externally closed or open applications in industrial processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 525 Dynamisches Current | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 56 dB(A) (sound pressure level) |
| Control output | POKO ONE (potential-free contact) |
| Expansion volume | 18 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 417 kg (net) |
| Current consumption | 16 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s (max) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 5xx Dynamisches chassis with casters/leveling feet provisions, 417 kg net, IP20 enclosure for dry indoor installation.
- Air-cooled refrigerant circuit using R-452A (A1, H280), 3.9 kg charge, GWP 2141 (8.3 t CO2 equivalent); the water-cooled sibling 525w Dynamisches is the alternative when heat rejection to air is not acceptable.
- Magnetically coupled stainless-steel pump with M30x1.5 male connections and adjustable speed; Pilot ONE controller with POKO ONE potential-free contact, Ethernet/USB/RS232, Pt100 external sensor input, 0.01 K display resolution.
- Heating capacity 6 kW; expansion volume 18 l; minimum filling capacity 5.1 l.
Behavior under load and media
- Temperature range -55 to 250 °C; cooling capacity with ethanol falls from 7 kW at 0 °C to 0.65 kW at -50 °C, so low-temperature duty must be checked against the actual fluid and temperature.
- Pump delivers up to 79 l/min and 1.5 bar; flow falls to 29 l/min at 1.25 bar, so external circuit pressure drop must be verified against the pump curve.
- Maximum permissible kinematic viscosity 50 mm²/s; heat-transfer fluid must remain below this limit at the lowest operating temperature.
- FL safety class III/FL: for flammable fluids the fire point must be ≥65 °C; ethanol only under supervised operation.
- Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off all poles; requires compressor set to Permanently ON.
Installation, connection and control
- 400 V 3~ 50 Hz supply, 16 A maximum current, 3x20 A fusing; delivered with cable but without plug (three-phase <63 A). Leakage current >3.5 mA requires protective earthing and RCD selection on site.
- Voltage/frequency tolerance ±5%/±2%, but -5% voltage with +2% frequency is not allowed. EMC Class A Group 1 for industrial environments.
- Ambient 5–40 °C; output data valid at 20 °C room temperature, cooling capacity may drop above 20 °C.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications; locking a closed loop causes overpressure damage.
- Power-interruption response configurable via Pilot ONE: Off/Standby (default) or ON/Temperature control active with automatic restart.
- Acclimation about 2 hours when moving between cold and warm environments before power-on; transport upright only, emptied.
Integration & Operating Notes
The unit must only be transported upright and emptied; horizontal transport damages the compressor and a filled unit risks thermal fluid overflow. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Casters are not for transport and carry only 25% of total mass symmetrically.
Selection & Integration Q&A
Choose the air-cooled 525 when no cooling water is available and 56 dB(A) is acceptable; the 525w is water-cooled at 61 dB(A). Smaller 510w/515w/520w siblings offer lower capacity; select the 525 when its 6 kW heating and cooling curve fit the load.
The pump connection is M30x1.5 male with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be specified. Verify external pressure drop against the pump curve (79 l/min max, 1.5 bar max; 29 l/min at 1.25 bar) and provide at least 5.1 l filling capacity with 18 l expansion volume.
Ambient must be 5–40 °C, dry indoor, IP20; cooling capacity may drop above 20 °C. The unit must not be installed or started within an ATEX zone without an Ex px cabinet. Transport upright and emptied, acclimate 2 hours before power-on, and provide a drip tray per local regulations.
The base delivery includes a cable but no plug; the site must fit the plug and provide 3x20 A fusing. Periodic checks of hoses and connections, cleaning of liquefier fins and strainer, and heat-transfer fluid inspection/replacement are required. Refrigerant work must be by approved contractors; a float switch in the sight glass is optional for extended level security.