Peter Huber Kältemaschinenbau Unistat 530w Dynamisches
Peter Huber Kältemaschinenbau Unistat 530w Dynamisches — Temperature control units with Noise level 59 dB.
Water-cooled Unistat 530w Dynamisches is a free-standing, magnetically coupled stainless-steel pump temperature control unit covering -55 to 250 °C with 12 kW heating and 21 kW cooling on Thermo-oil. Inside the Unistat 5xx w series it is the mid-capacity water-cooled variant, positioned between the 510w/515w/520w and the air-cooled 525 Dynamisches, intended for precise external closed or open process loops requiring high flow and 0.01 K stability.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 530w Dynamisches Current | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 59 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 443 kg |
| Current consumption | 25 A |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 890 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing water-cooled unit (suffix w) with dimensions 730 x 804 x 1738 mm and 443 kg net weight; requires an indoor, dry, ventilated location (IP20, ambient 5–40 °C).
- Magnetically coupled stainless-steel pump with M38x1.5 AG male process connections and G3/4 male cooling-water connections; pump speed adjustable.
- Controller is Pilot ONE with POKO ONE control output, 0.01 K display resolution, Pt100 external sensor input, and Ethernet, USB (host and device) and RS232 interfaces.
- Within the Unistat 5xx w Dynamisches family, the 530w is the larger water-cooled modification; the 525 Dynamisches is air-cooled and the 510w/515w/520w are smaller water-cooled siblings.
Behavior under load and media
- Operating range -55 to 250 °C with 0.01 K stability at -10 °C; heating capacity 12 kW.
- Cooling capacity is strongly fluid-dependent: 21 kW with Thermo-oil from 20 to 250 °C, but with ethanol 16 kW at 0 °C, 9 kW at -20 °C, 3 kW at -40 °C and 1 kW at -50 °C — low-temperature duty must be sized on the ethanol curve, not the oil rating.
- Pump delivers max 196 l/min and max 2.5 bar; at 2.0 bar backpressure 86 l/min remains, so the external circuit pressure drop must be checked against this curve.
- Cooling water: 5–20 °C intake, 1–4 bar differential pressure, max 6 bar, 890 l/h consumption at 15 °C with 0 °C flow; circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Refrigerant R-1270 (A3, H220), GWP 0, 1.35 kg charge — A3 flammability rules apply.
Installation, control and operation
- Electrical: 460 V 3~ 60 Hz, max 25 A, 3x25 A fusing; three-phase units under 63 A are supplied with cable but without plug; leakage current >3.5 mA requires external branch circuit protection per UL 489.
- Voltage/frequency tolerances ±5% / ±2% must not combine in the same direction.
- Pilot ONE supports Process Safety mode (compressor Permanently ON) and configurable power-failure restart (Off/Standby default or ON/Temperature control active).
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications; low-level protection via pressure sensor requires thorough venting and correct fill level.
Integration & Operating Notes
Voltage tolerance is ±5% and frequency tolerance ±2%, but the manual explicitly forbids the combination of -5% voltage and +2% frequency in the same supply; verify site supply quality before commissioning to avoid compressor and electronics damage.
The facility water system must deliver cooling water within 5–20 °C intake and 1–4 bar differential pressure at no more than 6 bar static pressure; 890 l/h is consumed at 15 °C water with 0 °C flow, so the supply header must be sized for that flow without starving other users.
When cooling water temperature can drop below the condensation point, the internal cooling-water pipes must be insulated to prevent condensation; this is a site-specific decision based on the facility water temperature and ambient humidity.
Selection & Integration Q&A
Choose the 530w when you need the higher cooling and heating capacity of the larger water-cooled frame (21 kW oil / 12 kW heating) and can supply 5–20 °C cooling water. The 525 Dynamisches is air-cooled (no water connection) and the 510w/515w/520w are smaller water-cooled siblings with lower capacity.
460 V 3~ 60 Hz, max 25 A, 3x25 A fusing, with external branch circuit protection per UL 489 because leakage current exceeds 3.5 mA. Voltage/frequency tolerances are ±5% / ±2% but must not combine as -5% voltage with +2% frequency. Cable supplied, plug not included.
Indoor, dry, ventilated location only: IP20, ambient 5–40 °C, no ATEX zone. Transport upright and emptied; allow 2 h acclimation before power-on. Leveling feet must be activated and casters are not for transport. A drip tray is required below the unit.
The base delivery excludes the final power plug, cooling-water hoses and insulation, drip tray, and heat transfer fluid. Periodic tasks include checking hoses and connections for leaks, cleaning liquefier fins and the hat-type strainer, inspecting the mechanical seal, and replacing heat transfer fluid as needed.