No photo available yet
Peter Huber Kältemaschinenbau Unistat P610 Dynamisches
Peter Huber Kältemaschinenbau Unistat P610 Dynamisches — Temperature control units.
The Unistat P610 Dynamisches is a free-standing, air-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -60 to 200 °C with 6 kW heating. Its P designation targets high-pressure-drop loops, distinguishing it from standard 5xx siblings. It serves precise process temperature control in closed circuits.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P610 Dynamisches Current | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 26 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 208V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 6xx-class chassis with Pilot ONE touchscreen controller and POKO ONE control output; IP20, indoor use only.
- The 'P' suffix denotes the high-pressure-drop variant: the pump delivers up to 3 bar and 82 l/min, but only 36 l/min at 2.5 bar, so it is selected when the external loop has significant pressure loss.
- Sibling models (Unistat 510w, 515w, 520w, 525 Dynamisches) differ in cooling configuration and noise level; the P610 is the air-cooled high-pressure-drop member of the family.
- Pump is magnetically coupled and stainless steel, with adjustable speed; pump connection is M30x1.5 male, bore shape Y (60°) per DIN 3863.
Thermal behavior and media
- Working range -60 to 200 °C; heating capacity 6 kW. Cooling capacity is strongly temperature-dependent: 2.6 kW at -40 °C falls to 0.05 kW at -60 °C, so the cold end is for holding low heat loads, not pull-down.
- Cooling capacity is rated at 20 °C ambient and derates as room temperature rises; ambient range is 5–40 °C.
- Heat transfer fluid must stay below 50 mm²/s kinematic viscosity at the lowest operating temperature; minimum filling capacity is 6.5 l with a 26 l expansion vessel.
- Refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, requiring A3 zoning and ignition-source assessment at the installation site.
Electrical and control architecture
- Supplied for 460 V 3~ 60 Hz operation via an external transformer fed from 208 V 3~ 60 Hz; maximum current with transformer is 90 A, and the transformer cabinet measures 364x740x682 mm.
- External branch circuit protection per UL 489 is required; leakage current exceeds 3.5 mA, so a high-leakage-current protective device or reinforced earthing is needed.
- Three-phase devices above 63 A are delivered without cable and plug; the customer must supply and terminate the power cable.
- Interfaces: Ethernet, USB host and device, RS232; external Pt100 sensor connection; display resolution 0.01 K.
- Safety class III/FL; overtemperature protection and Process Safety mode are configured via the Pilot ONE controller.
Integration & Operating Notes
The unit is configured for 460 V 3~ 60 Hz but requires an external transformer fed from 208 V 3~ 60 Hz. The transformer cabinet (364x740x682 mm) and its 3x100 A input fusing must be accommodated, and the feeder, disconnect and protection sized for at least 90 A. Three-phase devices above 63 A ship without cable or plug.
Standard 30 mA RCDs may nuisance-trip. The protection scheme must use a high-leakage-current protective device or reinforced earthing per local code, and external branch circuit protection according to UL 489 must be specified in the panel.
Cooling capacity collapses from 2.6 kW at -40 °C to 0.05 kW at -60 °C. The -60 °C end of the range is suitable only for low-heat-load holding, not for pull-down. The process heat load must be checked against the actual curve, and capacity must be derated for ambient temperatures above 20 °C.
In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Choose the power-interruption behavior (Off/Standby or ON/Temperature control active) according to whether unattended restart is acceptable.
Selection & Integration Q&A
Choose the P610 when the external loop has high pressure drop: its pump delivers 36 l/min at 2.5 bar (max 3 bar, 82 l/min). Siblings like the 525 are standard-flow variants; select them for low-pressure-drop loops where maximum flow matters more than head.
It is configured for 460 V 3~ 60 Hz via an external transformer fed from 208 V 3~ 60 Hz. Max current with transformer is 90 A; feeder and UL 489 branch protection must be sized accordingly. Leakage current exceeds 3.5 mA, so reinforced earthing or a high-leakage-current device is required.
Three-phase devices above 63 A ship without power cable and plug; the customer supplies and terminates the cable. The external transformer is a separate cabinet. Cooling water hoses, heat transfer fluid, and the optional gas detection sensor/processing unit are not included and must be engineered by the operating company.