Peter Huber Kältemaschinenbau Unistat 630w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 630w Dynamisches

Peter Huber Kältemaschinenbau Unistat 630w Dynamisches — Temperature control units.

The Unistat 630w Dynamisches is a water-cooled, free-standing temperature control system with a -60 to 200 °C range and 24 kW heating capacity, positioned as the high-capacity member of the Unistat 5xx/6xx series. Its magnetically coupled stainless-steel pump and Pilot ONE controller support dynamic temperature control of externally closed circuits. It is intended for precise thermal conditioning of industrial processes requiring high heating power and deep cooling.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 630w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 715 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Requires 400 V 3~ supply with 3x50 A fuses and no plug

The unit draws up to 48.7 A and is delivered with a cable but without a plug. The site must provide a suitable plug and protection devices rated for 3x50 A. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is not allowed. Leakage current exceeds 3.5 mA, so RCD selection and protective earthing must be engineered accordingly.

Datasheet, page 2
Cooling & Media
Cooling water must be 5–20 °C at 1–4 bar differential

The cooling water circuit requires a differential pressure of 1–4 bar and a static pressure no higher than 6 bar, with G3/4 male connections. Cooling capacity drops as the cooling water intake temperature rises above 5 °C. If the cooling water can fall below the condensation point, the internal cooling water pipes must be insulated to prevent condensation and corrosion.

Datasheet, page 2
Process Limits
Heat transfer fluid viscosity must not exceed 100 mm²/s

The magnetically coupled stainless-steel pump can handle fluids up to 100 mm²/s kinematic viscosity. At low temperatures, many fluids become too viscous and can overload the pump or reduce circulation. The minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l; ensure the fluid volume and expansion volume match the application.

Datasheet, page 2
Mounting
Activate leveling feet and parking brakes before commissioning

The unit weighs 715 kg and must be transported upright only. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. Before operation, unscrew/activate leveling feet and engage parking brakes to prevent creeping. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.

Operation Manual, Section 1.4 / page 22
Control & I/O
Process safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power failure auto setting can resume temperature control automatically after a power outage if it was active; choose this behavior according to whether unattended restart is acceptable.

Operation Manual, Section 1.7.2 / page 19-20
Maintenance
Shut off cooling water even for overnight downtimes

The cooling water supply should be shut off during short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.

Operation Manual, Section 1.4.2 / page 19 and page 29

Selection & Integration Q&A

When should I choose the Unistat 630w Dynamisches over the smaller 510w or 520w models?

Choose the 630w when you need 24 kW heating capacity and a -60 to 200 °C range with high cooling power (21 kW at 0 °C with ethanol). The 510w and 520w have lower capacities and are air-cooled; the 630w is water-cooled and weighs 715 kg, requiring cooling water supply and a 400 V 3~ 48.7 A supply.

What are the requirements for the cooling water and electrical connections for this unit?

Cooling water: G3/4 male connections, 1–4 bar differential pressure, max 6 bar static, intake 5–20 °C. Electrical: 400 V 3~ 50 Hz, max 48.7 A, 3x50 A fuses, cable delivered without plug. Leakage current >3.5 mA requires suitable RCD and earthing.

What maintenance and accessories are required that are not included in the base delivery?

The base delivery includes a cable but no plug; a suitable 3x50 A plug must be provided on site. A drip tray for condensation water/heat transfer fluid must be installed by the responsible body. Cooling water hoses and fittings (ball sockets to DIN 3863, sleeve nuts to DIN 3870) are not included. Periodic checks of hoses and fluid replacement are required.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 630w Temperature Control System
Datasheet ·18,9 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF