No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P525w Dynamisches

Peter Huber Kältemaschinenbau Unistat P525w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat P525w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C with 6 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from the standard Unistat 525 Dynamisches. It serves as a precise, dynamically controlled source for externally closed or open processes requiring three-phase power and cooling water.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P525w Dynamisches Current 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Expansion volume 18 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 208 kg
Electrical, Power & Drive Specifications
Current consumption 14 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, control, and operation

Integration & Operating Notes

Electrical
Three-phase supply requires 3x16 A fuse and site-provided plug

The unit is delivered with a cable but without a plug for three-phase devices below 63 A. The electrical installation must be sized for 14 A max current and a 3x16 A fuse, and a suitable plug must be provided on site. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this.

Datasheet, page 2; Manual, Section 1.4
Cooling & Media
Cooling water must be 5–20 °C with ≥1 bar differential pressure

Cooling capacity depends on cooling water temperature; higher temperatures reduce cooling power and may require higher flow. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible water may be used. If cooling water can fall below the dew point, internal pipes must be insulated to prevent condensation.

Datasheet, page 2
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active under exothermic conditions.

Operation Manual, page 19
Maintenance
Shut off cooling water even for overnight downtimes

The cooling water supply should be shut off even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Operation Manual, page 29
Mounting
Unit must not be installed in an ATEX zone

Without an Ex px cabinet, the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.

Operation Manual, Section 1.3.3 / page 15-16

Selection & Integration Q&A

When should I choose the Unistat P525w Dynamisches over the standard Unistat 525 Dynamisches or the smaller 510w/515w/520w water-cooled models?

Choose the P525w when your application has high pressure drop and requires water cooling; the 'P' designates suitability for high pressure drop. The standard 525 is air-cooled (56 dB(A)), while the P525w is water-cooled (61 dB(A)) and needs a cooling water supply. Smaller 510w/515w/520w models offer lower capacity; select based on required heating/cooling power and flow.

What are the requirements for the cooling water supply and the electrical connection when integrating this unit?

Cooling water: 5–20 °C, minimum 1 bar differential pressure, maximum 6 bar, G1/2 male connection. Electrical: 400 V 3~ 50 Hz, 14 A max, 3x16 A fuse; cable delivered without plug. Leakage current >3.5 mA requires appropriate earthing and RCD selection. Ambient 5–40 °C.

What are the mounting and environmental limits for the Unistat P525w Dynamisches?

Must not be installed in an ATEX zone. Ambient temperature 5–40 °C. Unit must be transported and stored upright only; allow 2 hours acclimation when moving between cold and warm environments before power-on. Leveling feet must be activated and parking brakes engaged. A drip tray must be installed below the unit.

What maintenance is required and what accessories are not included in the base delivery?

Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water even for overnight downtimes. The base delivery includes a cable but no plug for three-phase connection; a suitable plug must be provided on site. A drip tray is not included and must be installed by the responsible body. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P525w Technical Data
Datasheet ·76,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF