The Unistat P630w Dynamisches is a water-cooled, high-pressure-drop temperature control unit with a magnetically coupled stainless-steel pump delivering up to 201 l/min and 5.3 bar, covering -60 to 200 °C with 24 kW heating. Its 'P' designation targets processes with high pressure losses, distinguishing it from standard-flow siblings in the Unistat 5xx/6xx series. It serves reactor and semiconductor cooling/heating loops requiring stable circulation against restrictive flow paths.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P630w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Hydraulics and process behavior

Control, connection and operation

Integration & Operating Notes

Cooling & Media
Cooling water must meet pressure and temperature window

Cooling water is required at 1–4 bar differential pressure and no more than 6 bar, with intake temperature between 5 °C and 20 °C. If cooling water is colder than the condensation point, the internal cooling water pipes must be insulated to prevent condensation and corrosion. Shut off cooling water even for short downtimes such as overnight.

Unistat P630w Technical Data, page 2; Unistat Operation Manual, page 29
Mounting
Install level, upright and outside any ATEX zone

The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet. Transport and store upright only, and allow about 2 hours acclimation when moving between cold and warm environments before power-on. Leveling feet must be activated and caster parking brakes engaged before commissioning; casters are not suitable for transport.

Unistat Operation Manual, pages 15, 22, 29
Process Limits
Expansion vessel lock state depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Confirm the application type before filling and venting, and respect the 100 mm²/s maximum kinematic viscosity when selecting heat transfer fluid.

Unistat Operation Manual, page 33; Unistat P630w Technical Data, page 2
Control & I/O
Process safety mode needs compressor set to Permanently ON

In 'Process safety' mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, which protects against exothermic events. This mode requires the compressor to be set to Permanently ON. Separately, the power-interruption response can be set to Off/Standby (default) or ON/Temperature control active, so choose according to whether unattended restart is acceptable.

Unistat Operation Manual, pages 19–21

Selection & Integration Q&A

When should I choose the Unistat P630w Dynamisches instead of a standard-flow sibling such as the Unistat 520w Dynamisches?

Choose the P630w when the process circuit has high pressure losses: its pump still delivers 42 l/min at 5.0 bar against a 5.3 bar maximum. Standard-flow siblings suit low-pressure-drop loops where maximum flow matters more than discharge pressure.

What mating fittings and site services are required to connect the P630w to my process loop and cooling water?

Pump connections are M38x1.5 male and require ball sockets per DIN 3863 with sleeve nuts per DIN 3870. Cooling water connects via G3/4 male at 1–4 bar differential pressure, max 6 bar, 5–20 °C inlet. Electrical supply is 400 V 3~ 50 Hz with no plug supplied.

What maintenance and consumables are not included in base delivery for the P630w?

Base delivery excludes the power plug, mating pump fittings, heat transfer fluid, cooling water hoses and drip tray. Operator maintenance covers periodic hose inspection and replacement, fluid inspection/replacement and circuit rinsing; refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P630w Technical Data
Datasheet ·18,8 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF