No photo available yet
Peter Huber Kältemaschinenbau Unistat P635w Dynamisches
Peter Huber Kältemaschinenbau Unistat P635w Dynamisches — Temperature control units.
The Unistat P635w Dynamisches is a water-cooled, high-pressure-drop temperature control unit with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 24 kW heating. Its M38x1.5 pump connection and 5.3 bar maximum delivery pressure distinguish it from standard Unistat 5xx siblings for processes with high pressure drop. It serves closed-loop temperature control of reactors and process equipment requiring dynamic response.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P635w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unistat P635w Dynamisches is a free-standing, water-cooled (w) unit in the Unistat 6xx series, configured with the 'P' designation for applications with high pressure drop. Unlike the air-cooled Unistat 510w/515w/520w/525 Dynamisches siblings, it uses a water-cooled refrigerant circuit and a magnetically coupled stainless-steel pump with adjustable speed. The pump delivers up to 201 l/min and 5.3 bar maximum delivery pressure (42 l/min at 5.0 bar), with M38x1.5 male pump connections. The unit requires a 400 V 3~ 50 Hz supply and, because current exceeds 63 A, is delivered without cable or plug for hard-wired installation. Cooling water connections are G3/4 male.
Working range and behavior
Temperature range is -60 to 200 °C with 24 kW heating capacity and 0.01 K display resolution. The controller is Pilot ONE with POKO ONE control output, Ethernet, USB (Host and Device) and RS232 interfaces, and a Pt100 external sensor connection. The unit is rated safety class III/FL per DIN 12876 and pressure equipment category I. Maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s. Minimum filling capacity is 8.9 l with a 19.9 l expansion tank. Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and drops as cooling water temperature rises.
Installation and operation specifics
Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, with inlet temperature between 5 °C and 20 °C; if cooling water is colder than the condensation point, internal cooling water pipes must be insulated. The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature. Voltage and frequency tolerances are ±5 % and ±2 %, but the combination of −5 % voltage with +2 % frequency is not permitted. The equipment is EMC class A, group 1 (EN55011), intended for industrial environments. For externally closed applications the expansion vessel must remain unlocked; for externally open applications it must be locked. The unit must not be installed in an ATEX zone. A drip tray must be installed below the unit.
Integration & Operating Notes
Because current consumption exceeds 63 A, the unit is supplied without cable or plug and must be hard-wired by an electrician. Verify that the site supply matches the rating plate and that upstream protection is sized for the full load; the ±5 % voltage / ±2 % frequency tolerance combination of −5 % voltage with +2 % frequency is not permitted.
The water-cooled refrigerant circuit requires a differential pressure of at least 1 bar and no more than 4 bar, with static pressure not exceeding 6 bar and inlet temperature between 5 °C and 20 °C. Below 1 bar the condenser cannot reject rated capacity; above 4 bar or 6 bar static, components may be overstressed. Insulate internal cooling water pipes if the supply is below the condensation point.
For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. Conversely, for externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Confirm the application type before commissioning and set the vessel accordingly.
The magnetically coupled stainless-steel pump is limited to a maximum kinematic viscosity of 100 mm²/s. At low temperatures, fluid viscosity rises; select a heat transfer fluid whose viscosity at the lowest operating temperature remains below this limit to avoid pump cavitation or motor overload.
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. After transport between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Transport and store upright only.
In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the protective cooling action is not available. Configure the Pilot ONE accordingly if exothermic risk exists.
Shut off the cooling water supply to the unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Choose the P635w when the external circuit has high pressure drop and requires up to 5.3 bar delivery pressure with 201 l/min max flow; the 5xx siblings lack the 'P' high-pressure configuration. Select a 5xx model for lower pressure-drop applications where the P-variant's pump performance is unnecessary.
Provide a 400 V 3~ 50 Hz hard-wired supply (no plug delivered, >63 A), cooling water at 1–4 bar differential and ≤6 bar static with 5–20 °C inlet, G3/4 male cooling water connections, M38x1.5 male pump connections, and a drip tray. A Pt100 external sensor and Ethernet/USB/RS232 interfaces are available for integration.
Base delivery excludes power cable and plug (hard-wired), heat transfer fluid (min. 8.9 l filling), and cooling water hoses. Periodically check hoses and replace as needed; shut off cooling water overnight. Refrigerant-circuit work requires approved contractors. No specific service interval is stated in the available documents.