Peter Huber Kältemaschinenbau Unistat P635w Dynamisches

Peter Huber Kältemaschinenbau Unistat P635w Dynamisches — Temperature control units.

The Unistat P635w Dynamisches is a water-cooled, high-pressure-drop temperature control unit with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 24 kW heating. Its M38x1.5 pump connection and 5.3 bar maximum delivery pressure distinguish it from standard Unistat 5xx siblings for processes with high pressure drop. It serves closed-loop temperature control of reactors and process equipment requiring dynamic response.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P635w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unistat P635w Dynamisches is a free-standing, water-cooled (w) unit in the Unistat 6xx series, configured with the 'P' designation for applications with high pressure drop. Unlike the air-cooled Unistat 510w/515w/520w/525 Dynamisches siblings, it uses a water-cooled refrigerant circuit and a magnetically coupled stainless-steel pump with adjustable speed. The pump delivers up to 201 l/min and 5.3 bar maximum delivery pressure (42 l/min at 5.0 bar), with M38x1.5 male pump connections. The unit requires a 400 V 3~ 50 Hz supply and, because current exceeds 63 A, is delivered without cable or plug for hard-wired installation. Cooling water connections are G3/4 male.

Working range and behavior

Temperature range is -60 to 200 °C with 24 kW heating capacity and 0.01 K display resolution. The controller is Pilot ONE with POKO ONE control output, Ethernet, USB (Host and Device) and RS232 interfaces, and a Pt100 external sensor connection. The unit is rated safety class III/FL per DIN 12876 and pressure equipment category I. Maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s. Minimum filling capacity is 8.9 l with a 19.9 l expansion tank. Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and drops as cooling water temperature rises.

Installation and operation specifics

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, with inlet temperature between 5 °C and 20 °C; if cooling water is colder than the condensation point, internal cooling water pipes must be insulated. The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature. Voltage and frequency tolerances are ±5 % and ±2 %, but the combination of −5 % voltage with +2 % frequency is not permitted. The equipment is EMC class A, group 1 (EN55011), intended for industrial environments. For externally closed applications the expansion vessel must remain unlocked; for externally open applications it must be locked. The unit must not be installed in an ATEX zone. A drip tray must be installed below the unit.

Integration & Operating Notes

Electrical
Hard-wired 400 V supply required; no plug delivered

Because current consumption exceeds 63 A, the unit is supplied without cable or plug and must be hard-wired by an electrician. Verify that the site supply matches the rating plate and that upstream protection is sized for the full load; the ±5 % voltage / ±2 % frequency tolerance combination of −5 % voltage with +2 % frequency is not permitted.

Datasheet, Technical Specifications; Manual, Section on power connection
Cooling & Media
Cooling water differential pressure must stay within 1–4 bar

The water-cooled refrigerant circuit requires a differential pressure of at least 1 bar and no more than 4 bar, with static pressure not exceeding 6 bar and inlet temperature between 5 °C and 20 °C. Below 1 bar the condenser cannot reject rated capacity; above 4 bar or 6 bar static, components may be overstressed. Insulate internal cooling water pipes if the supply is below the condensation point.

Datasheet, Technical Specifications
Process Limits
Expansion vessel must be unlocked for closed applications

For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. Conversely, for externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Confirm the application type before commissioning and set the vessel accordingly.

Operation Manual, Preparations for Operation
Process Limits
Heat transfer fluid viscosity must not exceed 100 mm²/s

The magnetically coupled stainless-steel pump is limited to a maximum kinematic viscosity of 100 mm²/s. At low temperatures, fluid viscosity rises; select a heat transfer fluid whose viscosity at the lowest operating temperature remains below this limit to avoid pump cavitation or motor overload.

Datasheet, Technical Specifications
Mounting
Install a drip tray and acclimate unit before power-on

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. After transport between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Transport and store upright only.

Operation Manual, Commissioning
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the protective cooling action is not available. Configure the Pilot ONE accordingly if exothermic risk exists.

Operation Manual, Setup Mode
Maintenance
Shut off cooling water even for overnight downtimes

Shut off the cooling water supply to the unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.

Operation Manual, Commissioning and Maintenance

Selection & Integration Q&A

When should I choose the P635w Dynamisches instead of a Unistat 5xx Dynamisches sibling?

Choose the P635w when the external circuit has high pressure drop and requires up to 5.3 bar delivery pressure with 201 l/min max flow; the 5xx siblings lack the 'P' high-pressure configuration. Select a 5xx model for lower pressure-drop applications where the P-variant's pump performance is unnecessary.

What adjacent system parts and utilities must be provisioned for this unit?

Provide a 400 V 3~ 50 Hz hard-wired supply (no plug delivered, >63 A), cooling water at 1–4 bar differential and ≤6 bar static with 5–20 °C inlet, G3/4 male cooling water connections, M38x1.5 male pump connections, and a drip tray. A Pt100 external sensor and Ethernet/USB/RS232 interfaces are available for integration.

What maintenance and consumables are required, and what is not included in base delivery?

Base delivery excludes power cable and plug (hard-wired), heat transfer fluid (min. 8.9 l filling), and cooling water hoses. Periodically check hoses and replace as needed; shut off cooling water overnight. Refrigerant-circuit work requires approved contractors. No specific service interval is stated in the available documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P635w Technical Data
Datasheet ·18,8 КБ ·2 pp.
View PDF