No photo available yet
Peter Huber Kältemaschinenbau Unistat P525 Dynamisches
Peter Huber Kältemaschinenbau Unistat P525 Dynamisches — Temperature control units with Noise level 56 dB.
The Unistat P525 Dynamisches is a refrigerated heating circulator with a magnetically coupled stainless steel high-discharge-pressure pump, built for externally closed or open circuits with high pressure drop. Its -55 to 250 °C range and 6 kW heating capacity distinguish it from the 525 and water-cooled w siblings. It serves flow-through chemistry and semiconductor processes requiring stable temperature control under restrictive flow paths.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P525 Dynamisches Current | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 56 dB(A) |
| Control output | POKO ONE |
| Expansion volume | 18 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Power supply | 208V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P525 Dynamisches is the high-pressure-drop variant of the Unistat 525 series; the 'P' suffix denotes a circulating pump with elevated discharge pressure for restrictive external circuits such as flow-through chemistry or semiconductor applications.
- Unlike the water-cooled 'w' siblings (510w, 515w, 520w), this model is air-cooled; noise is 56 dB(A), matching the standard 525 and lower than the 60–61 dB(A) of the w variants.
- The pump is magnetically coupled and made of stainless steel; the unit is a refrigerated heating circulator with an air-cooled cooling machine and optical level indicator.
- Controller is Pilot ONE with POKO ONE control output, 0.01 K display resolution, and Ethernet, USB (host and device) and RS232 interfaces.
Working range and media behavior
- Temperature range is -55 to 250 °C; heating capacity is 6 kW; expansion volume is 18 l.
- Permissible ambient temperature is 5 to 40 °C; cooling capacity is rated at 20 °C room temperature and may drop as ambient rises.
- The unit is rated safety class III / FL; for FL-rated flammable heat transfer fluid the fluid must have a fire point of at least 65 °C, and ethanol use is permitted only under supervision.
- Viscosity limit is 50 mm²/s; external sensor connection is Pt100.
- The pump connection is M30x1.5 male with bore shape Y (60°) per DIN 3863, mating with ball socket and sleeve nut per DIN 3863/3870.
Installation, connection and control
- Power supply is 208 V 3~ 60 Hz; an external transformer supplied as standard provides the required voltage to the unit (output 440 V 3~ 60 Hz). The transformer is a separate component measuring 253x560x602 mm and weighing 135 kg.
- The transformer input requires 3x50 A fuses; maximum current of the temperature control unit with external transformer is 45 A. External branch circuit protection according to UL 489 is required and not included.
- Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for high leakage.
- Three-phase devices below 63 A are supplied with cable but without a plug.
- The Pilot ONE controller supports overtemperature protection settings (OT limit heating, OT expansion vessel, Process Safety), Delta T limiter, Pt100 monitoring, auto/expert mode, and filling/venting/degassing/draining procedures for externally closed and open applications.
- For externally closed applications the expansion vessel must remain unlocked; for externally open applications it must be locked.
Integration & Operating Notes
The transformer input requires 3x50 A fuses, and the maximum current of the temperature control unit with external transformer is 45 A. External branch circuit protection according to UL 489 is required and is not included in delivery. Leakage current exceeds 3.5 mA, so the protective earth and any RCD must be selected for high leakage; standard 30 mA RCDs may nuisance-trip.
For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. The correct state must be verified before commissioning, and the unit must be transported and stored upright only, with an emptied circuit before any move.
Published cooling capacity is valid for a room temperature of 20 °C; if the ambient temperature rises, cooling capacity may drop. The permissible ambient range is 5 to 40 °C, so installation rooms that exceed 20 °C will deliver less than rated heat extraction and must be accounted for in the process heat balance.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON, so the controller configuration must be checked before relying on this behavior for exothermic process protection.
Selection & Integration Q&A
Choose the P525 when the external circuit has high pressure drop, such as flow-through chemistry or semiconductor applications, because the P pump delivers high discharge pressure. The standard 525 has the same 56 dB(A) noise but lacks the high-pressure pump. The w variants (510w, 515w, 520w) are water-cooled and louder at 60–61 dB(A); select them only when water cooling is available and air cooling is not preferred.
The unit requires a 208 V 3~ 60 Hz supply feeding the supplied external transformer (output 440 V 3~ 60 Hz). Provide 3x50 A fuses on the transformer input and UL 489 listed branch protection; maximum unit current with transformer is 45 A. Leakage current exceeds 3.5 mA, so earth and RCD selection must accommodate high leakage. A plug is not supplied for three-phase devices below 63 A.
Install only outside ATEX zones, in an ambient range of 5 to 40 °C, with cooling capacity rated at 20 °C room temperature. Transport and store upright only, and move only an emptied unit. Allow about 2 hours acclimation when moving between cold and warm environments before power-on. Leveling feet must be activated and parking brakes engaged before commissioning.
Base delivery includes the external transformer but not branch protection, a plug, or a drip tray; the responsible body must install a condensation/fluid drip tray. Periodic checks of hoses and connections for leaks and quality are required, with replacement as needed. Cooling water supply should be shut off even overnight. Refrigerant circuit work must be performed by approved refrigeration contractors.