Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches
Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches — Temperature control units.
The Unistat GL 550w Dynamisches is a water-cooled, CO2 (R-744) free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 200 °C. Its 24 kW heating and up to 41 kW cooling with thermooil distinguish it from air-cooled siblings in the Unistat 5xx series. It serves closed-loop process temperature control in industrial automation where high heat-transfer fluid flow (196 l/min max) and three-phase power are available.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat GL 550w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...200 °C |
| Weight | 800 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 3100 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing, water-cooled (w suffix) unit with CO2 refrigerant (GL suffix), 800 kg, 918x963x1771 mm; requires 400 V 3~ 50 Hz supply with 3x63 A fuse and max current 61.5 A.
- Magnetically coupled stainless-steel pump with M38x1.5 male connection; cooling water circuit uses G3/4 male connections.
- Pilot ONE controller with POKO ONE control output; interfaces Ethernet, USB (host/device), RS232; external Pt100 sensor input.
- Safety class III/FL; IP20; ambient 5–40 °C.
- Sibling models in the Unistat 5xx series (510w, 515w, 520w, 525) differ in cooling capacity and noise level; this GL 550w adds CO2 refrigerant and higher heating capacity (24 kW).
Behavior under load and media
- Temperature range -50 to 200 °C; cooling capacity is strongly fluid- and temperature-dependent: 41 kW with thermooil at 20–200 °C, 6 kW with ethanol at -50 °C.
- Pump delivers 196 l/min max and 86 l/min at 2.0 bar; max delivery pressure 2.5 bar; system pressure drop must be checked against this curve.
- Minimum filling capacity 8.9 l; expansion tank 19.9 l; viscosity limit 100 mm²/s.
- Cooling water consumption 3100 l/h at 15 °C, flow 0 °C; differential pressure 1 bar; max pressure 6 bar.
Installation, control and operation
- Requires 400 V three-phase supply; delivered with cable but without plug (current <63 A). Leakage current >3.5 mA requires appropriate earthing and RCD selection.
- Voltage/frequency tolerance ±5%/±2%, but -5% voltage with +2% frequency is not allowed.
- Cooling water intake 5–20 °C, 1–4 bar differential, max 6 bar; insulate pipes if below dew point.
- EMC Class A, Group 1 to EN55011; may need filtering in sensitive environments.
- Pilot ONE offers Process Safety mode (max cooling on OT trigger), power-failure auto-restart option, and Delta T limiter.
- Must not be installed in ATEX zone; use with medical devices or direct foodstuff is not permissible.
Integration & Operating Notes
The unit draws up to 61.5 A from a 400 V 3~ 50 Hz supply, requiring a 3x63 A fuse and corresponding cable. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must comply with local regulations. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is not allowed.
Cooling water intake temperature must be between 5 °C and 20 °C, with a differential pressure of 1–4 bar and maximum pressure 6 bar. Consumption is 3100 l/h at 15 °C. Pipes must be insulated if water can drop below the dew point to prevent condensation. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
Installation or start-up within an ATEX zone is prohibited; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or condensing environments.
For FL-rated (flammable) heat transfer fluid, the fluid must have a fire point of at least 65 °C; if ethanol is used, only supervised operation is permitted. The unit is rated safety class III/FL. Use with medical devices (e.g. in vitro diagnostic procedures) or for direct foodstuff temperature control is not permissible.
In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active. This keeps heat extraction available under exothermic conditions, reducing risk from glass breakage or uncontrolled phase transitions.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors.
Selection & Integration Q&A
Choose the GL 550w when you need CO2 refrigerant (GWP 1) and high heating capacity (24 kW) with water cooling. Air-cooled siblings lack the GL suffix and have lower heating capacity; the 510w and 520w are air-cooled and suited where cooling water is unavailable. The 550w requires a cooling water supply of 5–20 °C at 1–4 bar differential.
Cooling water: G3/4 male connections, 5–20 °C intake, 1–4 bar differential, max 6 bar, 3100 l/h at 15 °C. Electrical: 400 V 3~ 50 Hz, max 61.5 A, 3x63 A fuse, cable supplied without plug. Leakage current >3.5 mA requires suitable earthing and RCD. Voltage/frequency tolerance ±5%/±2%, but -5% voltage with +2% frequency is not allowed.
Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water even for short downtimes. Clean liquefier fins and hat-type strainer. The base delivery includes a power cable but no plug (for <63 A three-phase). Cooling water hoses and adapters for G3/4 male and pump M38x1.5 male are not included and must be procured. A drip tray must be installed by the responsible body.