No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P540w Dynamisches

Peter Huber Kältemaschinenbau Unistat P540w Dynamisches — Temperature control units.

The Unistat P540w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C. It is the high-capacity 'P' variant for applications with high pressure drop, delivering up to 201 l/min and 5.3 bar. It serves precise temperature control of externally closed or open circuits in industrial processes.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P540w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 473 kg
Electrical, Power & Drive Specifications
Current consumption 56 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1480 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unistat P540w Dynamisches is a free-standing, water-cooled unit (suffix 'w') with a 'P' designation for applications with high pressure drop. It uses a magnetically coupled stainless-steel pump with adjustable speed, a Pilot ONE controller, and POKO ONE control output. The cooling water circuit is built from copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. With a net weight of 473 kg and dimensions 730x804x1738 mm, it requires planned floor space and handling equipment. Unlike air-cooled siblings, it relies on a cooling water supply of 5–20 °C at 1–4 bar differential pressure (max 6 bar).

Performance and behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ 50 Hz with 3x63 A fusing

The unit draws up to 56 A per phase and requires 3x63 A fuses. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must be engineered accordingly. A plug is not included; the installer must provide a suitable connection.

Datasheet, page 2
Cooling & Media
Cooling water must be 5–20 °C at 1–4 bar differential

The cooling water circuit requires a differential pressure between 1 and 4 bar and a maximum pressure of 6 bar. Consumption is 1480 l/h at 15 °C. Higher water temperatures reduce cooling power. If water temperature falls below the condensation point, insulate the cooling water pipes in the unit. Shut off cooling water during downtimes, even overnight.

Datasheet, page 2; Operation Manual, page 29
Process Limits
Heat transfer fluid viscosity must not exceed 100 mm²/s

The magnetically coupled pump can be overloaded by viscous fluids. The maximum permissible kinematic viscosity is 100 mm²/s. For FL-rated fluids, the fire point must be ≥65 °C; ethanol requires supervised operation. Low-temperature cooling capacity is limited: 1.5 kW at -50 °C with ethanol.

Datasheet, page 2; Operation Manual, page 19
Mounting
Transport upright and allow 2 hours acclimation before power-on

Horizontal transport damages the compressor. Only move an emptied unit. Casters are not for transport and carry only 25% of total mass. Leveling feet must be activated and parking brakes engaged before commissioning. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Operation Manual, page 22 and page 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power Failure Auto can resume temperature control automatically after a power outage if it was active; choose this behavior according to whether unattended restart is acceptable.

Operation Manual, page 19 and page 20
Maintenance
Periodically inspect hoses and replace as needed

Hoses and connections must be checked periodically for leaks and quality and replaced as required. The mechanical seal should be inspected. The hat-type strainer/dirt trap requires cleaning, including draining the cooling circuit. Refrigerant-circuit work must be carried out by approved refrigeration/air-conditioning contractors.

Operation Manual, page 29; Maintenance section

Selection & Integration Q&A

When should I choose the Unistat P540w Dynamisches over other Unistat 5xx Dynamisches models?

Choose the P540w when the external circuit has high pressure drop and requires up to 201 l/min at 5.3 bar, and when water cooling is available. Smaller siblings like the 510w or 520w offer lower capacity; select them for lower heat loads or where cooling water is limited.

What are the requirements for the cooling water supply and the pump connection?

Cooling water: 5–20 °C, differential pressure 1–4 bar, max 6 bar, connection G3/4 male. Pump connection is M38x1.5 male; use ball socket per DIN 3863 and sleeve nut per DIN 3870. Insulate pipes if water temperature is below the condensation point.

What maintenance is required and what is not included in the base delivery?

Periodically check hoses and connections for leaks and replace as needed; clean the strainer; inspect mechanical seal. A plug is not included (cable only). A drip tray must be installed by the operator. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P540w Technical Data
Datasheet ·86,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF