Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches

Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches — Temperature control units.

The Unistat GL 550w Dynamisches is a water-cooled, CO2 (R-744) free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 200 °C. Its 24 kW heating and up to 41 kW cooling with thermooil distinguish it from air-cooled siblings in the Unistat 5xx series. It serves closed-loop process temperature control in industrial automation where high heat-transfer fluid flow (196 l/min max) and three-phase power are available.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat GL 550w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...200 °C
Dimensions, Weight & Mechanical Design
Weight 800 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 3100 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be sized for 61.5 A and 3x63 A fusing

The unit draws up to 61.5 A from a 400 V 3~ 50 Hz supply, requiring a 3x63 A fuse and corresponding cable. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must comply with local regulations. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is not allowed.

Datasheet, page 2
Cooling & Media
Cooling water must be 5–20 °C and 1–4 bar differential

Cooling water intake temperature must be between 5 °C and 20 °C, with a differential pressure of 1–4 bar and maximum pressure 6 bar. Consumption is 3100 l/h at 15 °C. Pipes must be insulated if water can drop below the dew point to prevent condensation. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Datasheet, page 2
Mounting
Unit must not be installed in an ATEX zone

Installation or start-up within an ATEX zone is prohibited; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or condensing environments.

Operation Manual, Section 1.3.3 / page 15-16
Process Limits
Heat transfer fluid must have fire point ≥65 °C for FL rating

For FL-rated (flammable) heat transfer fluid, the fluid must have a fire point of at least 65 °C; if ethanol is used, only supervised operation is permitted. The unit is rated safety class III/FL. Use with medical devices (e.g. in vitro diagnostic procedures) or for direct foodstuff temperature control is not permissible.

Operation Manual, Section 1.7 / page 19
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active. This keeps heat extraction available under exothermic conditions, reducing risk from glass breakage or uncontrolled phase transitions.

Operation Manual, Section 1.7.2 / page 20
Maintenance
Shut off cooling water even for short downtimes

Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors.

Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat GL 550w Dynamisches instead of an air-cooled sibling like the Unistat 510w or 520w?

Choose the GL 550w when you need CO2 refrigerant (GWP 1) and high heating capacity (24 kW) with water cooling. Air-cooled siblings lack the GL suffix and have lower heating capacity; the 510w and 520w are air-cooled and suited where cooling water is unavailable. The 550w requires a cooling water supply of 5–20 °C at 1–4 bar differential.

What are the requirements for the cooling water circuit and electrical supply when integrating this unit?

Cooling water: G3/4 male connections, 5–20 °C intake, 1–4 bar differential, max 6 bar, 3100 l/h at 15 °C. Electrical: 400 V 3~ 50 Hz, max 61.5 A, 3x63 A fuse, cable supplied without plug. Leakage current >3.5 mA requires suitable earthing and RCD. Voltage/frequency tolerance ±5%/±2%, but -5% voltage with +2% frequency is not allowed.

What maintenance is required and what accessories are not included in the base delivery?

Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water even for short downtimes. Clean liquefier fins and hat-type strainer. The base delivery includes a power cable but no plug (for <63 A three-phase). Cooling water hoses and adapters for G3/4 male and pump M38x1.5 male are not included and must be procured. A drip tray must be installed by the responsible body.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat GL 550w Technical Data
Datasheet ·98,5 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF