Peter Huber Kältemaschinenbau Unistat 525 Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 525 Dynamisches

Peter Huber Kältemaschinenbau Unistat 525 Dynamisches — Temperature control units with Noise level 56 dB.

The Unistat 525 Dynamisches is a free-standing, air-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C. Within the Unistat 5xx Dynamisches series it is the non-water-cooled 525 variant, distinguished from the 525w by air cooling and a 56 dB(A) sound level. It serves closed-loop heating and cooling of externally closed or open applications in industrial processes.

Noise level
56 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 525 Dynamisches Current 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 56 dB(A) (sound pressure level)
Control output POKO ONE (potential-free contact)
Expansion volume 18 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 417 kg (net)
Electrical, Power & Drive Specifications
Current consumption 16 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Mounting
Transport upright and emptied; acclimate 2 hours before power-on

The unit must only be transported upright and emptied; horizontal transport damages the compressor and a filled unit risks thermal fluid overflow. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Casters are not for transport and carry only 25% of total mass symmetrically.

Unistat Operation Manual, page 22 and page 29

Selection & Integration Q&A

When should I choose the Unistat 525 Dynamisches instead of the 525w Dynamisches or a smaller 5xx sibling?

Choose the air-cooled 525 when no cooling water is available and 56 dB(A) is acceptable; the 525w is water-cooled at 61 dB(A). Smaller 510w/515w/520w siblings offer lower capacity; select the 525 when its 6 kW heating and cooling curve fit the load.

What must I provide on the process side to connect the pump and external circuit?

The pump connection is M30x1.5 male with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be specified. Verify external pressure drop against the pump curve (79 l/min max, 1.5 bar max; 29 l/min at 1.25 bar) and provide at least 5.1 l filling capacity with 18 l expansion volume.

What ambient and installation limits apply, and can the unit be placed in a hazardous area?

Ambient must be 5–40 °C, dry indoor, IP20; cooling capacity may drop above 20 °C. The unit must not be installed or started within an ATEX zone without an Ex px cabinet. Transport upright and emptied, acclimate 2 hours before power-on, and provide a drip tray per local regulations.

What maintenance and accessories are required beyond the base delivery?

The base delivery includes a cable but no plug; the site must fit the plug and provide 3x20 A fusing. Periodic checks of hoses and connections, cleaning of liquefier fins and strainer, and heat-transfer fluid inspection/replacement are required. Refrigerant work must be by approved contractors; a float switch in the sight glass is optional for extended level security.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 525 Technical Datasheet
Datasheet ·170,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF