Peter Huber Kältemaschinenbau Unistat GL 560w Dynamisches
Peter Huber Kältemaschinenbau Unistat GL 560w Dynamisches — Temperature control units.
The Unistat GL 560w Dynamisches is a water-cooled, CO2 (R-744) refrigerated circulating thermostat with a magnetically coupled stainless-steel pump, covering -50 to 200 °C with 24 kW heating. Within the Unistat 5xx series it is the GL/water-cooled modification, requiring a 400 V three-phase supply and a 1–4 bar cooling-water circuit.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat GL 560w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration machine using natural refrigerant R-744 (GWP 1); no air-cooled option, so a cooling-water supply is mandatory.
- Magnetically coupled stainless-steel pump with M38x1.5 male pressure connection; pump speed adjustable.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; external Pt100 sensor input.
- Interfaces: Ethernet, USB (host and device), RS232.
- Safety class III / FL (DIN 12876 flammable-fluid rating).
- Sibling Unistat 5xx models (510w, 515w, 520w, 525) share the series architecture; this GL 560w is the CO2-refrigerant, water-cooled variant.
Working range and media
- Temperature range -50 to 200 °C; heating capacity 24 kW.
- Maximum permissible kinematic viscosity of the heat transfer fluid: 100 mm²/s.
- FL rating requires a heat transfer fluid with fire point ≥ 65 °C; ethanol only under supervised operation.
- Cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM — only compatible water may be used.
Installation and operation
- 400 V 3~ 50 Hz supply; three-phase units below 63 A are delivered with cable but without plug.
- Voltage tolerance ±5% with simultaneous frequency tolerance ±2%.
- IP20 protection; ambient 5–40 °C; indoor dry location, not in an ATEX zone.
- Cooling water: G3/4 male connection, 1–4 bar differential pressure, max 6 bar, intake 5–20 °C.
- Rated cooling data valid only at 20 °C room, 15 °C cooling-water inlet and 1 bar differential pressure.
- Pilot ONE power-interruption behavior selectable (Off/Standby or automatic restart); Process Safety mode available for exothermic protection.
Integration & Operating Notes
The building water system must deliver at least 1 bar and no more than 4 bar differential pressure across the cooling-water circuit, with static pressure never exceeding 6 bar. Below 1 bar the rated cooling capacity is not reached; above 6 bar the internal circuit can be damaged. Rated performance is quoted only at 15 °C inlet and 1 bar differential.
The unit requires 400 V 3~ 50 Hz and is delivered with a cable but without a plug because three-phase devices below 63 A ship that way. The electrical installer must provide a matching plug and verify that voltage stays within ±5% while frequency stays within ±2%; certain tolerance combinations are not permitted.
For an externally closed application the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For an externally open (bath) application the vessel must be locked to prevent fluid loss. The installation must also include a condensation-water/heat-transfer-fluid drip tray below the unit.
Heat transfer fluid with kinematic viscosity above 100 mm²/s overloads the magnetically coupled pump and degrades circulation. Because the unit is FL-rated, the fluid must also have a fire point of at least 65 °C, and ethanol may only be used under supervised operation.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This mode only works if the compressor is configured to Permanently ON; otherwise the protective cooling action is not available during exothermic events.
The cooling-water supply should be closed for any downtime, including overnight, to avoid water damage and unnecessary load on the circuit. Hoses and hose connections must be checked periodically for leaks and quality and replaced as required; refrigerant-circuit work is restricted to approved refrigeration contractors.
Selection & Integration Q&A
Pick the GL 560w when a cooling-water supply is available and CO2 refrigerant (R-744, GWP 1) is preferred for environmental compliance. Air-cooled siblings need no water but reject heat to the room; choose them only where water is unavailable and room heat load is acceptable.
A G3/4 male cooling-water connection fed at 1–4 bar differential (max 6 bar, intake 5–20 °C) with water compatible with copper, 1.4401 stainless, PA, PPE, PTFE and EPDM. The process side uses the M38x1.5 male pump connection, so matching hoses and fittings must be ordered separately.
Install indoors at 5–40 °C ambient, IP20, never inside an ATEX zone. Transport and store upright only, move only when emptied, and let the unit acclimate about 2 hours after a cold-to-warm move before power-on. Leveling feet or caster brakes must be set before commissioning.
The three-phase plug is not included (cable only). A drip tray below the unit is the operator's responsibility. Periodic tasks include checking hoses for leaks and quality, cleaning the dirt trap and liquefier fins, and replacing heat transfer fluid; refrigerant work requires approved contractors.