No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P525 Dynamisches

Peter Huber Kältemaschinenbau Unistat P525 Dynamisches — Temperature control units with Noise level 56 dB.

The Unistat P525 Dynamisches is a refrigerated heating circulator with a magnetically coupled stainless steel high-discharge-pressure pump, built for externally closed or open circuits with high pressure drop. Its -55 to 250 °C range and 6 kW heating capacity distinguish it from the 525 and water-cooled w siblings. It serves flow-through chemistry and semiconductor processes requiring stable temperature control under restrictive flow paths.

Noise level
56 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P525 Dynamisches Current 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 56 dB(A)
Control output POKO ONE
Expansion volume 18 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Electrical, Power & Drive Specifications
Power supply 208V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Branch protection must be UL 489 listed and sized for 45 A

The transformer input requires 3x50 A fuses, and the maximum current of the temperature control unit with external transformer is 45 A. External branch circuit protection according to UL 489 is required and is not included in delivery. Leakage current exceeds 3.5 mA, so the protective earth and any RCD must be selected for high leakage; standard 30 mA RCDs may nuisance-trip.

Unistat P525 Technical Data Sheet, page 2
Mounting
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. The correct state must be verified before commissioning, and the unit must be transported and stored upright only, with an emptied circuit before any move.

Unistat Operation Manual, Section on Preparations for Operation, page 33
Process Limits
Cooling capacity is rated at 20 °C room temperature

Published cooling capacity is valid for a room temperature of 20 °C; if the ambient temperature rises, cooling capacity may drop. The permissible ambient range is 5 to 40 °C, so installation rooms that exceed 20 °C will deliver less than rated heat extraction and must be accounted for in the process heat balance.

Unistat P525 Technical Data Sheet, page 2
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON, so the controller configuration must be checked before relying on this behavior for exothermic process protection.

Unistat Operation Manual, Section 1.7.2, page 20

Selection & Integration Q&A

When should I choose the P525 Dynamisches instead of the standard 525 or the water-cooled w variants?

Choose the P525 when the external circuit has high pressure drop, such as flow-through chemistry or semiconductor applications, because the P pump delivers high discharge pressure. The standard 525 has the same 56 dB(A) noise but lacks the high-pressure pump. The w variants (510w, 515w, 520w) are water-cooled and louder at 60–61 dB(A); select them only when water cooling is available and air cooling is not preferred.

What electrical infrastructure and protection must be provided for this unit?

The unit requires a 208 V 3~ 60 Hz supply feeding the supplied external transformer (output 440 V 3~ 60 Hz). Provide 3x50 A fuses on the transformer input and UL 489 listed branch protection; maximum unit current with transformer is 45 A. Leakage current exceeds 3.5 mA, so earth and RCD selection must accommodate high leakage. A plug is not supplied for three-phase devices below 63 A.

What are the mounting and ambient limits for installing this unit?

Install only outside ATEX zones, in an ambient range of 5 to 40 °C, with cooling capacity rated at 20 °C room temperature. Transport and store upright only, and move only an emptied unit. Allow about 2 hours acclimation when moving between cold and warm environments before power-on. Leveling feet must be activated and parking brakes engaged before commissioning.

What maintenance and accessories are required beyond the base delivery?

Base delivery includes the external transformer but not branch protection, a plug, or a drip tray; the responsible body must install a condensation/fluid drip tray. Periodic checks of hoses and connections for leaks and quality are required, with replacement as needed. Cooling water supply should be shut off even overnight. Refrigerant circuit work must be performed by approved refrigeration contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P525 Technical Data Sheet
Datasheet ·66,4 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF