No photo available yet
Peter Huber Kältemaschinenbau Unistat P525w Dynamisches
Peter Huber Kältemaschinenbau Unistat P525w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat P525w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C with 6 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from the standard Unistat 525 Dynamisches. It serves as a precise, dynamically controlled source for externally closed or open processes requiring three-phase power and cooling water.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P525w Dynamisches Current | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 61 dB(A) |
| Control output | POKO ONE |
| Expansion volume | 18 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 208 kg |
| Current consumption | 14 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P525w is a water-cooled (w) free-standing Unistat 5xx model with a 400 V 3~ 50 Hz supply, 14 A max current, and a 3x16 A fuse requirement. The 'P' suffix denotes a unit configured for applications with high pressure drop, while 'Dynamisches' indicates dynamic temperature control behavior.
- Unlike the air-cooled Unistat 525 Dynamisches (56 dB(A)), this water-cooled variant is rated at 61 dB(A) and requires a cooling water supply. Sibling models 510w, 515w, and 520w share the water-cooled architecture but differ in capacity and noise level (60–61 dB(A)).
- The pump is magnetically coupled and stainless steel, with M30x1.5 male connections, delivering up to 82 l/min and 3 bar max pressure. The cooling water connection is G1/2 male.
Working range and media behavior
- Temperature range is -55 to 250 °C with 0.01 K display resolution. Heating capacity is 6 kW. The expansion vessel holds 18 l, and minimum filling capacity is 5.1 l.
- Cooling capacity depends on cooling water temperature: higher cooling water temperature reduces cooling power and may require increased flow. The unit is water-cooled and CFC/HCFC-free, using refrigerant R-452A (GWP 2141, CO2 equivalent 3.1 t).
- The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM; only compatible cooling water may be used. If cooling water temperature can fall below the dew point, internal cooling water pipes must be insulated to prevent condensation.
Installation, control, and operation
- The Pilot ONE controller provides Ethernet, USB (Host and Device), and RS232 interfaces, with POKO ONE control output. External Pt100 sensor connection is supported. Safety class is III/FL.
- Installation requires a 400 V three-phase supply with 3x16 A fuse; three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site. Leakage current exceeds 3.5 mA, affecting earthing and RCD selection.
- Cooling water must be 5–20 °C with at least 1 bar differential pressure and maximum 6 bar pressure. Ambient temperature must remain 5–40 °C. The unit must not be installed in an ATEX zone. A drip tray for condensation water/heat transfer fluid must be installed below the unit.
- For externally closed applications, the expansion vessel must remain unlocked; for externally open applications, it must be locked. Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off all poles, and requires the compressor set to Permanently ON.
Integration & Operating Notes
The unit is delivered with a cable but without a plug for three-phase devices below 63 A. The electrical installation must be sized for 14 A max current and a 3x16 A fuse, and a suitable plug must be provided on site. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this.
Cooling capacity depends on cooling water temperature; higher temperatures reduce cooling power and may require higher flow. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible water may be used. If cooling water can fall below the dew point, internal pipes must be insulated to prevent condensation.
In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active under exothermic conditions.
The cooling water supply should be shut off even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Without an Ex px cabinet, the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.
Selection & Integration Q&A
Choose the P525w when your application has high pressure drop and requires water cooling; the 'P' designates suitability for high pressure drop. The standard 525 is air-cooled (56 dB(A)), while the P525w is water-cooled (61 dB(A)) and needs a cooling water supply. Smaller 510w/515w/520w models offer lower capacity; select based on required heating/cooling power and flow.
Cooling water: 5–20 °C, minimum 1 bar differential pressure, maximum 6 bar, G1/2 male connection. Electrical: 400 V 3~ 50 Hz, 14 A max, 3x16 A fuse; cable delivered without plug. Leakage current >3.5 mA requires appropriate earthing and RCD selection. Ambient 5–40 °C.
Must not be installed in an ATEX zone. Ambient temperature 5–40 °C. Unit must be transported and stored upright only; allow 2 hours acclimation when moving between cold and warm environments before power-on. Leveling feet must be activated and parking brakes engaged. A drip tray must be installed below the unit.
Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water even for overnight downtimes. The base delivery includes a cable but no plug for three-phase connection; a suitable plug must be provided on site. A drip tray is not included and must be installed by the responsible body. Refrigerant-circuit work requires approved contractors.