No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P527w Dynamisches

Peter Huber Kältemaschinenbau Unistat P527w Dynamisches — Temperature control units with Noise level 57 dB.

The Unistat P527w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C with 12 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from standard 5xx siblings. It serves precise temperature control of externally closed or open circuits in industrial processes.

Noise level
57 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P527w Dynamisches Current 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 57 dB(A)
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 444 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 880 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ with 3x32 A fusing and no plug supplied

The unit draws up to 31.62 A and requires a 3x32 A fuse; three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug and connection. Voltage tolerance is ±5% only with simultaneous frequency tolerance ±2%, and opposite-sign deviations are not allowed.

Datasheet, page 2; Operation Manual, Section 1.8
Cooling & Media
Cooling water must stay within 5–20 °C and 1–4 bar differential

The cooling water circuit requires 880 l/h at 15 °C water and 0 °C flow, with minimum differential pressure 1 bar and maximum 4 bar, and absolute pressure not exceeding 6 bar. Inlet temperature must be between 5 °C and 20 °C; if water can fall below the dew point, pipes must be insulated to prevent condensation.

Datasheet, page 2
Process Limits
Fluid viscosity above 100 mm²/s is not permissible

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s. At low temperatures viscosity rises, so fluid selection must ensure this limit is not exceeded at the lowest operating temperature; otherwise circulation and heat transfer degrade.

Datasheet, page 2
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the protective cooling function is not available.

Operation Manual, page 19
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Operation Manual, page 29

Selection & Integration Q&A

When should I choose the P527w Dynamisches instead of a sibling like the 525 Dynamisches or 520w Dynamisches?

Choose the P527w when the external circuit has high pressure drop and water cooling is available; the 'P' modification is designed for high pressure drop. The 525 Dynamisches is air-cooled and quieter (56 dB(A)), while the 520w is water-cooled with 61 dB(A). Select based on cooling capacity, pressure drop and cooling water availability.

What are the requirements for the cooling water supply and the electrical connection?

Cooling water: G3/4 male, 880 l/h at 15 °C water/0 °C flow, differential pressure 1–4 bar, max. pressure 6 bar, inlet 5–20 °C. Electrical: 400 V 3~ 50 Hz, max. 31.62 A, 3x32 A fuse; cable supplied but no plug, so provide a suitable plug and connection.

What are the mounting and ambient limits for this unit?

Install upright and empty, with leveling feet activated and parking brakes engaged. Ambient temperature 5–40 °C, IP20, not in ATEX zones. Allow 2 h acclimation after temperature changes. Cooling water inlet 5–20 °C; insulate pipes if below dew point.

What maintenance is required and what accessories are not included in base delivery?

Periodically check hoses and connections for leaks and quality; shut off cooling water during downtimes. The plug for the power cable is not included. A drip tray for condensation/fluid must be installed by the responsible body. Refrigerant work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P527w Technical Data
Datasheet ·86,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.