Peter Huber Kältemaschinenbau Unistat P545w Dynamisches
Peter Huber Kältemaschinenbau Unistat P545w Dynamisches — Temperature control units.
The Unistat P545w Dynamisches is a water-cooled refrigerated heating circulator with a magnetically coupled stainless-steel pump built for high pressure drop circuits, covering -55 to 250 °C with 24 kW heating. Its 'P' pump (201 l/min, 5.3 bar max) distinguishes it from standard-pressure siblings for flow-through chemistry and semiconductor applications.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P545w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1630 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigerated heating circulator with electronically indicated level, 24 kW heating and up to 35 kW cooling; suitable for externally closed and open circuits.
- The 'P' designation denotes a circulating pump with high discharge pressure for applications with high pressure drops (flow-through chemistry, semiconductor industry); the 'w' denotes water cooling. Siblings 510w/515w/520w/525 Dynamisches are the lower-capacity air/water-cooled variants of the same series.
- Pump: magnetically coupled, stainless steel, speed-adjustable, max delivery 201 l/min and max pressure 5.3 bar; pump connection M38x1.5 AG male. Cooling-water connection G3/4 male.
- Control via Pilot ONE with POKO ONE control output; display resolution 0.01 K; external Pt100 sensor input; interfaces Ethernet, USB host/device, RS232.
Behavior under load and media
- Temperature range -55...250 °C; heating capacity 24 kW. Cooling capacity falls as cooling-water temperature rises, and cooling-water flow rate may need to increase accordingly.
- Cooling water: intake 5...20 °C, differential pressure 1...4 bar, max pressure 6 bar; consumption 1630 l/h at 15 °C, flow 0 °C.
- Heat transfer fluid max kinematic viscosity 100 mm²/s. Safety class III / FL; adjustable overtemperature protection per DIN 12876; 'Process safety' mode releases maximum cooling at the OT trigger instead of switching off all poles (requires compressor set to Permanently ON).
- Refrigerant R-452A (A1, H280), 3.6 kg, CO2 equivalent 7.7 t.
Installation, connection and operation
- Power supply 400 V 3~ 50 Hz; three-phase units below 63 A are delivered with cable but without plug. Voltage tolerance ±5%, frequency ±2%, but -5% voltage with +2% frequency is not allowed. Protection rating IP20; EN55011 Class A, Group 1.
- Ambient 5...40 °C. Must not be installed in an ATEX zone. Transport and store upright only; move only an emptied unit; allow ~2 h acclimation after cold/warm transfer before power-on.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications. Shut off cooling water even for overnight downtimes. Insulate cooling-water pipes if water temperature can fall below the condensation point.
Integration & Operating Notes
The unit is designed for cooling-water intake between 5 °C and 20 °C; warmer water reduces cooling power and may require higher flow. Insulate cooling-water pipes where water temperature can fall below the condensation point to avoid condensation inside the unit.
Voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against this envelope before connection; the three-phase unit is delivered with cable but without a plug, so the installer must provide the connection.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. Confirm the circuit type before commissioning.
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s; selecting a fluid that exceeds this at operating temperature overloads the pump and circuit. For FL-rated fluids the fire point must be ≥65 °C, and ethanol use permits only supervised operation.
In 'Process safety' mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Configure the restart behavior (Off/Standby vs ON/Temperature control active) according to whether unattended restart is acceptable.
Close the cooling-water supply during short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Periodically check hoses and connections for leaks and quality and replace as required; work on the refrigerant circuit must be done by approved refrigeration/air-conditioning contractors.
The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. Site selection must exclude hazardous areas.
Selection & Integration Q&A
Choose the P545w when the external circuit has high pressure drop (flow-through chemistry, semiconductor) and needs the 'P' high-discharge pump (201 l/min, 5.3 bar) plus water cooling; the 5xx Dynamisches siblings are lower-capacity alternatives for less demanding circuits.
G3/4 male connection; intake 5–20 °C, differential pressure 1–4 bar, max pressure 6 bar; consumption 1630 l/h at 15 °C, flow 0 °C. Insulate pipes where water can fall below the condensation point and shut off supply during downtimes.
Three-phase units below 63 A ship with cable but without a plug; the site must provide the plug/connection and a drip tray. Periodic hose/connection checks and cooling-water shutoff are operator duties; refrigerant work requires approved contractors.