Peter Huber Kältemaschinenbau Unistat P515w Dynamisches — Temperature control units.

The Unistat P515w Dynamisches is a water-cooled, high-discharge-pressure circulating temperature control unit with a -50 to 250 °C working range and 6 kW heating capacity. Its magnetically coupled stainless-steel pump and M30x1.5 connections distinguish it from air-cooled and standard-pressure siblings in the Unistat 5xx series. It serves processes with high pressure-drop external circuits requiring precise, dynamic temperature control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P515w Dynamisches Current —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...250 °C
Dimensions, Weight & Mechanical Design
Weight 181 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 504 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Provide 400 V 3~ supply with 3x16 A protection and local plug

The unit requires a 400 V three-phase 50 Hz supply protected at 3x16 A. Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Unistat P515w Technical Data, page 2
Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential pressure

Cooling water intake temperature must stay between 5 °C and 20 °C, with minimum differential pressure 1 bar and maximum pressure 6 bar. Higher cooling water temperature reduces cooling power and may require increased flow rate. Pipes must be insulated if condensation can occur. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Unistat P515w Technical Data, page 2
Process Limits
Heat transfer fluid viscosity must not exceed 50 mm²/s

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s. For FL-rated (flammable) fluid, the fluid must have a fire point of ≥65 °C; if ethanol is used, only supervised operation is permitted. Thorough venting and correct fill level are mandatory to avoid low-level protection interrupting circulation.

Unistat P515w Technical Data, page 2; Operation Manual, Section 1.7
Mounting
Install a drip tray and exclude ATEX zones

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. The unit must NOT be installed or started up within an ATEX zone; explosion protection requires an Ex px cabinet. Ambient temperature must be 5–40 °C. Transport and store upright only; allow about 2 hours acclimation when moving between cold and warm environments before power-on.

Operation Manual, Section 1.4 / page 17; Section 1.3.3 / page 15-16; page 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power Failure Auto resumes temperature control automatically after a power outage if it was active; choose power-interruption behavior according to whether unattended restart is acceptable.

Operation Manual, Section 1.7.2 / page 20; page 19; page 20
Maintenance
Shut off cooling water even for overnight downtimes

Shut off the cooling water supply to the unit even for shorter downtimes (e.g. overnight) to prevent water damage and unnecessary load on the cooling water circuit. Check periodically for leaks and the quality of the hose and hose connections and replace as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors.

Operation Manual, page 29; Section 1.4.2 / page 19

Selection & Integration Q&A

When should I choose the Unistat P515w Dynamisches instead of a sibling model like the 510w or 520w?

Choose the P515w when the external circuit has high pressure drop requiring the high-discharge-pressure pump. The 510w and 520w differ in cooling capacity and noise level; select those if pressure drop is moderate and their capacity matches the load.

What are the requirements for connecting the cooling water circuit and the process loop?

Cooling water: G1/2 male connection, 5–20 °C intake, 1 bar min differential pressure, 6 bar max, 504 l/h at 15 °C. Process loop: M30x1.5 male pump connections with bore shape Y (60°) per DIN 3863; hoses must use ball socket and sleeve nut per DIN 3870.

What maintenance is required and what is not included in base delivery?

Periodically check hoses and connections for leaks and quality; shut off cooling water overnight; clean liquefier fins and strainer. A plug for the 400 V cable is not included; a drip tray and compatible heat transfer fluid must be provided on site.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P515w Technical Data
Datasheet ·83,8 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®