Peter Huber Kältemaschinenbau Unichiller 030Tw
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 030Tw

Peter Huber Kältemaschinenbau Unichiller 030Tw — Recirculating chillers with Noise level 61 dB.

The Unichiller 030Tw is a water-cooled recirculating chiller for externally closed circuits, rated -20 to 40 °C with a 26 l/min pump and G3/4 male process connections. Its 61 dB(A) acoustic level distinguishes it from the 69 dB(A) 030T sibling, while the -H variant shares the same noise figure. It serves as a central cooling source for closed process loops requiring a stable temperature window.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 030Tw Current 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Expansion volume 4 l
Maximum delivery flow rate 26 l/min (pressure pump max; also 3 bar)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 137 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 25 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 65 л/час · at 15°C, flow 0°C: 99 l/h · at 15°C, flow 15°C: 99 l/h
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media conditions

Installation, control and operation

Integration & Operating Notes

Process Limits
Closed-loop process only; not for open baths

The hydraulic circuit must be externally closed because the unit uses an atmospheric open expansion tank and is designed for externally closed applications. Open baths or vented loops will not match the expansion and filling concept (2.5 l minimum fill, 4 l expansion volume).

Unichiller 030Tw Technical Data, Product Overview
Cooling & Media
Cooling water 5–20 °C and 1–6 bar differential

The facility cooling-water supply must deliver 5 to 20 °C intake at a differential pressure of at least 1 bar and no more than 6 bar at the G1/2 male connection. Warmer water lowers cooling power and can raise the required flow rate beyond the rated 99 l/h at 15 °C.

Unichiller 030Tw Technical Data, Technical Specifications
Cooling & Media
Cooling water circuit materials restrict water quality

The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only cooling water compatible with this material set may be used; aggressive or chloride-rich water risks corrosion of the copper and stainless components.

Unichiller 030Tw Technical Data, Technical Specifications
Electrical
Voltage and frequency tolerances are coupled

The 400 V 3~ 50 Hz supply must stay within ±5% voltage only when frequency is within ±2%. The combination of -5% voltage with +2% frequency is explicitly not allowed, so supply quality should be verified before commissioning to avoid invalid operating conditions.

Unichiller 030Tw Technical Data, Technical Specifications
Electrical
Cable supplied without plug; 3x10 A fusing required

Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug, so the installer must provide a suitable plug or hard-wire the connection on site. The supply must be fused at 3x10 A; maximum unit current is 4.5 A.

Unichiller 030Tw Technical Data, Technical Specifications
Mounting
Indoor only, 50–200 cm side clearance, level before start

Installation is restricted to indoors at illuminance of at least 300 lx and altitude up to 2000 m, with 50 to 200 cm wall clearance per side for maintenance. Leveling feet must be unscrewed or parking brakes applied before the unit is switched on to prevent movement and crushing injury.

Unichiller Operation Manual, Installation
Mounting
Not permitted in ATEX zones

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Refrigerant charge (0.7 kg R-449A) and room size must be checked at installation so the refrigerant limit per m³ of ambient air is not exceeded.

Unichiller Operation Manual, Installation
Mounting
Protect cables below water-cooled unit from condensate

Electrical lines routed directly below a water-cooled unit must be protected against liquid ingress from condensation. Pump connections must be counterheld and the maximum tightening torques must not be exceeded during hose installation to avoid leaks and thread damage.

Unichiller Operation Manual, Installation
Control & I/O
Configure power-failure response before unattended runs

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For unattended processes the chosen behavior determines whether the pump and cooling restart automatically after a mains outage, which must be matched to the process risk.

Unichiller Operation Manual, Normal Operation
Control & I/O
Process Safety mode releases max cooling on OT trigger

In Process Safety mode the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is relevant for exothermic processes where cutting cooling would cause runaway temperatures; the mode must be selected deliberately.

Unichiller Operation Manual, Normal Operation
Maintenance
Refrigerant circuit work only by approved contractors

Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration and air-conditioning system contractors. The refrigerant R-449A is classified A1 with GHS hazard H280, and the 0.7 kg charge corresponds to 1.0 t CO2 equivalent, which may trigger leak-checking obligations.

Unichiller Operation Manual, Maintenance
Maintenance
Shut off cooling water even for overnight downtime

The cooling-water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion during idle periods. This applies to the G1/2 cooling-water connection and should be part of the daily shutdown procedure.

Unichiller Operation Manual, Installation

Selection & Integration Q&A

When should I choose the 030Tw over the 030T or 030Tw-H?

Pick the 030Tw when a water-cooled unit at 61 dB(A) is required and the process loop is externally closed without high pressure drop. The 030T is air-cooled at 69 dB(A); the 030Tw-H shares the 61 dB(A) figure but is a different modification. For high pressure-drop loops, a 'P' model is specified instead.

What cooling-water supply and piping must be provided for the 030Tw?

Provide 5–20 °C cooling water at 1–6 bar differential pressure to the G1/2 male connection, sized for 99 l/h at 15 °C supply (65 l/h at -10 °C, 25 l/h at -20 °C process flow). Use only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, and insulate pipes if condensation can form.

What installation and duty-cycle limits apply from the manual?

Indoors only, at least 300 lx illuminance, altitude up to 2000 m, ambient 5–40 °C, with 50–200 cm wall clearance per side. Not permitted in ATEX zones. Leveling feet or parking brakes must be set before start-up, and cooling water must be shut off even overnight.

What is not included in base delivery and what service is required?

The three-phase cable is supplied without a plug, so a plug or hard-wiring must be provided on site; fusing is 3x10 A. Refrigerant-circuit work must be done by approved refrigeration contractors. Minimum filling capacity is 2.5 l with 4 l expansion volume, so the external loop must be filled and vented accordingly.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 030Tw Technical Data
Datasheet ·63,7 КБ ·2 pp.
View PDF