Peter Huber Kältemaschinenbau Unichiller 030T
Peter Huber Kältemaschinenbau Unichiller 030T — Recirculating chillers with Noise level 69 dB.
The Unichiller 030T is a recirculating chiller with a Pilot ONE controller, rated for -10 to 40 °C and 460 V 3~ 60 Hz supply. It distinguishes itself within the series by its 69 dB(A) noise level, while the 030Tw variant operates at 61 dB(A). It is intended for externally closed cooling circuits in industrial processes.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 030T Current | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 69 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...40 °C |
| Weight | 164 kg |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and Series Positioning
The Unichiller 030T is a recirculating chiller in the Umwaelzkuehler Bis 10 Kw series. It uses refrigerant R-449A (GHS H280, GWP 1397) and is classified under Article 4.3 PED. The unit has IP20 protection and is designed for indoor installation only. Compared to its siblings, the 030T shares the 69 dB(A) noise level with the 040T and 040T-H, while the 030Tw and 030Tw-H are quieter at 61 dB(A). The 030T is therefore positioned as a standard-noise model, with quieter alternatives available in the same capacity class.
Hydraulic and Thermal Behavior
The chiller operates over a temperature range of -10 to 40 °C. The pump delivers a maximum of 31 l/min at up to 4 bar, with flow decreasing as pressure rises; the external circuit pressure drop must be matched to this pump curve. The minimum filling capacity is 3.5 l, and the expansion volume is also 3.5 l, so the external loop volume and expansion allowance must be verified. Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient temperatures, requiring derating for warm rooms. The unit is intended only for externally closed applications.
Control, Installation, and Operation
The Pilot ONE controller provides Ethernet, USB (Host and Device), and RS232 interfaces. Power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached, rather than switching off all poles. Installation requires a three-phase 460 V 60 Hz supply with a 3x10 A fuse and external branch circuit protection according to UL 489. Three-phase devices under 63 A are supplied with cable but without a plug. Voltage and frequency tolerances are ±5% and ±2% respectively, but must not be combined in opposite directions. The pump connection is G3/4 male, and maximum tightening torques must not be exceeded. A wall clearance of 50 to 200 cm per side is required for maintenance. The unit must not be installed in an ATEX zone, and refrigerant charge versus room size must be checked.
Integration & Operating Notes
The chiller requires a three-phase 460 V 60 Hz supply with a 3x10 A fuse. External branch circuit protection according to UL 489 is mandatory, so a suitable breaker must be included in the installation. Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions; verify supply quality before commissioning.
The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit must not be installed or started up inside an ATEX zone because it is not explosion-protected. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.
The pump delivers a maximum of 31 l/min at up to 4 bar, with flow dropping as pressure rises. The minimum filling capacity is 3.5 l and expansion volume is 3.5 l, so the external circuit volume and expansion allowance must be checked. The unit is intended only for externally closed applications, which constrains the hydraulic circuit design.
The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE. Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles. This is critical for exothermic processes where cutting cooling would cause runaway temperatures.
Selection & Integration Q&A
Choose the 030T when standard noise (69 dB(A)) is acceptable and you need the -10 to 40 °C range with 460 V 3~ 60 Hz supply. The 030Tw offers 61 dB(A) for noise-sensitive areas. The 040T provides the same noise level but is a different capacity class; select it only if higher cooling capacity is required.
The 030T requires a three-phase 460 V 60 Hz supply with a 3x10 A fuse. External branch circuit protection according to UL 489 is mandatory. Three-phase devices under 63 A are supplied with cable but without a plug, so the installer must provide the final connection. Voltage and frequency tolerances are ±5% and ±2%, but must not be combined in opposite directions.
Install indoors only, at 5–40 °C ambient temperature, with illuminance at least 300 lx and altitude up to 2000 m. Maintain 50–200 cm wall clearance per side for maintenance. Do not install in an ATEX zone. Ensure refrigerant charge per room volume is within limits. Leveling feet must be activated and parking brakes applied before operation.
Refrigerant circuit work must be performed by approved refrigeration contractors. Monitor fluid level via the digital indicator and maintain the expansion tank. Shut off cooling water during downtimes. The base delivery includes a cable but no plug for three-phase connection; mating hoses and fittings for G3/4 male pump connection are not included. External branch circuit protection (UL 489 breaker) must be provided by the installer.