Peter Huber Kältemaschinenbau Unichiller P100-H
Peter Huber Kältemaschinenbau Unichiller P100-H — Recirculating chillers.
The Unichiller P100-H is a recirculating chiller with a high-discharge-pressure E2 pump, built for externally closed circuits with high pressure drop. Its -20...100 °C range and 2 kW heating distinguish it from standard-pump siblings. It serves as a closed-loop cooling source for industrial processes requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P100-H Current | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 299 kg |
| Current consumption | 15 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P100-H is a free-standing, air-cooled recirculating chiller with an atmospheric open expansion tank (48 l) and a minimum fill volume of 18 l.
- The 'P' designation denotes a high-discharge-pressure circulating pump (type E2, non-adjustable speed) for circuits with high pressure drop; standard Unichiller models (e.g., 030T, 040T) use lower-pressure pumps.
- Pump connection is G1 1/4 AG male; the unit is not explosion-protected and must not be installed in ATEX zones.
- Refrigerant is R-449A (A1, H280), charge 1.1 kg, GWP 1397, equivalent to 1.5 t CO2.
Operating behavior
- Temperature range is -20...100 °C with 2 kW heating capacity; cooling capacity is rated at 20 °C room temperature and derates as ambient rises.
- The unit is intended only for externally closed applications; open or vented external circuits are not acceptable.
- The Pilot ONE controller supports Ethernet, USB host/device, and RS232 for remote monitoring and data transmission.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles.
Installation and operation
- Requires 400 V 3~ 50 Hz supply, max current 15 A, fusing 3x16 A; supplied with cable but without plug.
- Voltage and frequency tolerances are linked: ±5 % voltage only with ±2 % frequency; -5 % voltage with +2 % frequency is not allowed.
- IP20 protection; indoor use only, illuminance ≥300 lx, altitude up to 2000 m, ambient 5...40 °C.
- Wall clearance of 50–200 cm per side is required for maintenance; leveling feet or parking brakes must be engaged before start-up.
- Cooling water supply (if water-cooled) must be shut off even for short downtimes; electrical lines below the unit must be protected against condensate.
Integration & Operating Notes
The chiller is designed exclusively for externally closed applications; an open or vented external circuit will not work and may cause fluid loss or poor temperature control. Ensure the external loop is sealed and compatible with the pump's high discharge pressure.
The unit draws up to 15 A per phase and requires a 400 V 3~ 50 Hz supply with 3x16 A fusing. It is delivered with a cable but without a plug, so the installer must provide a suitable 3-phase plug and ensure the site supply meets the linked voltage/frequency tolerance rules.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side for maintenance. The unit must be placed indoors with illuminance at least 300 lx and at altitudes up to 2000 m. Leveling feet or parking brakes must be engaged before operation to prevent movement.
The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, instead of switching off all poles. Choose the mode that matches the process hazard analysis.
Any work on the refrigerant circuit must be performed only by approved refrigeration/air-conditioning system contractors. The unit contains 1.1 kg of R-449A (GWP 1397, 1.5 t CO2 equivalent), so refrigerant handling and leak-check obligations apply. The expansion tank has a digital level indicator for routine monitoring.
Published cooling capacity is valid at 20 °C room temperature; higher ambient temperatures reduce available cooling. Size the chiller with margin if the installation room cannot be held near 20 °C. Also shut off cooling water supply (if water-cooled) during downtimes to prevent condensate damage.
Selection & Integration Q&A
Choose the P100-H when your external circuit has high pressure drop that would stall a standard pump; its 'P' pump delivers higher discharge pressure. For low-pressure-drop loops or lower noise needs, the 030T/040T series may suffice, but verify cooling capacity and temperature range for your duty.
It requires 400 V 3~ 50 Hz, max 15 A per phase, with 3x16 A fusing. The unit comes with a cable but no plug, so you must provide a suitable 3-phase plug. Ensure voltage and frequency tolerances are met: ±5 % voltage only with ±2 % frequency; -5 % voltage with +2 % frequency is not allowed.
Install indoors only, at 5...40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, and not in an ATEX zone. Provide 50–200 cm wall clearance per side for maintenance. Engage leveling feet or parking brakes before start-up. IP20 requires a dry, protected location.
Refrigerant circuit work must be by certified contractors; monitor fluid level via the digital indicator. The base delivery includes a cable but no plug, and no external hoses or fittings are mentioned. Regular checks of the expansion tank level and cooling water shutoff during downtimes are necessary.