Peter Huber Kältemaschinenbau Unichiller 040Tw-H
Peter Huber Kältemaschinenbau Unichiller 040Tw-H — Recirculating chillers with Noise level 70 dB.
The Unichiller 040Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a temperature range of -10 to 100 °C, distinguished from air-cooled siblings by its G1/2 cooling-water connection and 70 dB(A) sound level. It serves externally closed process loops requiring precise temperature control with Pilot ONE electronics.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 040Tw-H Current | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 70 dB(A) |
| Expansion volume | 4 l |
| Heating capacity | 2 kW |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 141 kg |
| Current consumption | 9 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-449A (A1, H280) |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 32 l/h (cooling water 15°C, flow -10°C) cooling water 15°C, flow 0°C: 70 l/h · cooling water 15°C, flow 15°C: 132 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
The 040Tw-H is a water-cooled variant within the Unichiller series, using R-449A refrigerant and a fixed-speed pump (type B) with G3/4 male process connections. Unlike the air-cooled 030T and 040T siblings (69 dB(A)), this model requires an external cooling-water supply (G1/2 male) and operates at 70 dB(A). The 'w' denotes water cooling; the 'H' indicates an integrated 2 kW heater, enabling active heating as well as cooling.
Hydraulic and Thermal Behavior
The unit is designed for externally closed applications only. The cooling water circuit consumes 32–132 l/h depending on the flow temperature (at 15 °C cooling water), with a minimum differential pressure of 1 bar and maximum pressure of 6 bar. Cooling power decreases as cooling water temperature rises, and flow demand may increase. The internal expansion volume is 4 l, with a minimum filling capacity of 2.5 l. The temperature range spans -10 to 100 °C, and the Pilot ONE controller manages both cooling and heating.
Installation and Control
Installation requires 50–200 cm wall clearance per side for maintenance, indoor use only (illuminance ≥300 lx, altitude ≤2000 m), and protection from ATEX zones. Electrical supply is 400 V 3~ 50 Hz with max 9 A and 3x10 A fusing; a cable is provided but no plug. The Pilot ONE offers Ethernet, USB, and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling if the overtemperature trigger is reached.
Integration & Operating Notes
Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance access to the refrigerant circuit and components. Insufficient clearance can prevent servicing and may lead to overheating or restricted airflow.
Operating with cooling water below the condensation point requires insulation of the cooling water pipes to prevent condensation. Exceeding 20 °C reduces cooling capacity and may increase flow demand, affecting process stability.
The cooling water circuit is rated for a maximum pressure of 6 bar and a minimum differential pressure of 1 bar. Exceeding 6 bar can damage the heat exchanger or connections, while insufficient differential pressure may cause inadequate cooling flow.
Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A (GWP 1397) requires proper handling and recovery per regulations; unauthorized intervention voids safety compliance.
The Pilot ONE allows selecting OFF/Standby, ON, or Power Failure Auto after a power outage. For exothermic processes, Power Failure Auto or ON may be necessary to prevent runaway, while OFF/Standby is safer for unattended restart hazards.
Selection & Integration Q&A
Choose 040Tw-H when a cooling-water supply (1–6 bar, 5–20 °C) is available and lower ambient heat rejection is needed; it adds a 2 kW heater. Air-cooled 040T suits sites without water; 030Tw-H offers lower capacity but similar water cooling.
Provide cooling water at 5–20 °C with differential pressure ≥1 bar and static pressure ≤6 bar. Flow demand is 32–132 l/h at 15 °C water, depending on load. Use materials compatible with copper, stainless steel 1.4401, brass, PA, PPE, PTFE, and EPDM; insulate pipes if below dew point.
Install indoors only, at 5–40 °C ambient, ≤2000 m altitude, illuminance ≥300 lx, and not in ATEX zones. Maintain 50–200 cm side clearance. The unit is IP20, so keep it dry. Shut off cooling water during downtimes. Leveling feet or parking brakes must be engaged before operation.
The base delivery includes a power cable but no plug; the installer must provide a suitable three-phase plug or hardwire. Refrigerant circuit service requires certified contractors. Regular maintenance includes checking cooling water quality and clearing dust. No external sensor is included; a Pt100 can be connected.