Peter Huber Kältemaschinenbau Unichiller 100Tw
Peter Huber Kältemaschinenbau Unichiller 100Tw — Recirculating chillers with Noise level 65 dB.
The Unichiller 100Tw is a water-cooled recirculating chiller with a 96 l/min, 5.6 bar pump and a -20…40 °C working range, controlled by the Pilot ONE. Within the Unichiller series it is the water-cooled variant, requiring a 1–6 bar cooling-water supply instead of an air-cooled condenser. It serves externally closed process loops that need stable temperature control with heat rejection to a facility water circuit.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 100Tw Current | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 65 dB(A) |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 227 kg |
| Current consumption | 12 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 276 l/h (cooling water 15°C, flow -20°C) cooling water 15°C, flow -10°C: 375 l/h · cooling water 15°C, flow 0°C: 510 l/h · cooling water 15°C, flow 15°C: 510 l/h |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled condenser: the unit rejects heat to a facility cooling-water circuit (G3/4 male connection) rather than to room air, so it does not add condenser heat load to the installation room.
- Process circuit: C3 circulation pump, non-speed-adjustable, with G1 1/4 male pump connection; maximum delivery 96 l/min at 5.6 bar, with flow falling as external pressure drop rises.
- Fluid inventory: minimum filling capacity 6.5 l, expansion volume 5.5 l.
- Refrigerant R-449A, charge 1.75 kg (2.4 t CO2 equivalent), safety group A1.
- Compared with air-cooled siblings (e.g. Unichiller 030T, 030Tw, 030Tw-H, 040T), this model trades the air-cooled condenser for a water-cooled one, which is the decisive selection criterion when room heat rejection or ambient conditions are constrained.
Behavior across the working range
- Temperature range -20…40 °C; cooling-water consumption rises as the process flow temperature falls: 276 l/h at -20 °C, 375 l/h at -10 °C, 510 l/h at 0 °C and 510 l/h at +15 °C, all at 15 °C cooling-water inlet.
- Cooling power decreases as cooling-water temperature rises, and the required cooling-water flow rate may increase accordingly.
- The unit is intended for externally closed applications only; the hydraulic circuit must be designed as a closed loop.
Installation, control and operation
- Controller: Pilot ONE with Ethernet, USB (host and device) and RS232 interfaces; external Pt100 sensor connection available.
- Electrical: 400 V 3~ 50 Hz, 12 A current consumption; three-phase devices below 63 A are delivered with cable but without a plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and not all combinations are permitted (e.g. -5% voltage with +2% frequency is not allowed).
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto; Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
- Protection rating IP20; indoor use only, illuminance at least 300 lx, altitude up to 2000 m.
Integration & Operating Notes
The water-cooled condenser requires a facility cooling-water circuit with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar at the G3/4 male connection. Size the supply for the flow demand at the coldest process temperature (510 l/h at 0 °C and +15 °C, 276 l/h at -20 °C, all at 15 °C cooling-water inlet) and account for the rise in flow when cooling-water temperature increases.
The unit is designed for cooling-water intake temperatures between 5 °C and 20 °C. When the cooling-water temperature falls below the condensation point, the cooling-water pipes inside the unit must be insulated to prevent condensate formation. The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
The 400 V 3~ 50 Hz supply may deviate by ±5% voltage only with a simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed (e.g. -5% voltage with +2% frequency). Verify the site power quality against this envelope before connection; the unit is delivered with a cable but without a plug because current consumption is below 63 A.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access. The unit must be installed indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m. Before startup, activate the parking brakes on the casters and/or unscrew the leveling feet so the unit cannot move during operation.
The chiller is designed exclusively for externally closed applications, so the process loop must be hydraulically closed. It is not explosion-protected and must not be installed or started up inside an ATEX zone. Check the refrigerant charge (1.75 kg R-449A) against the room volume to stay within the permissible refrigerant limit per m³ of ambient air.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For processes that must resume automatically after a mains interruption, select Power Failure Auto; for exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Close the cooling-water supply to the unit during any downtime, including overnight, to prevent water damage and corrosion. Work on the refrigerant circuit must only be carried out by approved refrigeration/air-conditioning contractors. Electrical lines routed directly below a water-cooled unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Choose the 100Tw when heat must be rejected to a facility water circuit rather than to room air, or when ambient conditions make air cooling impractical. The air-cooled siblings (030T, 030Tw, 030Tw-H, 040T) avoid the cooling-water supply but release condenser heat into the installation room.
A cooling-water circuit with 1–6 bar differential pressure at a G3/4 male connection, sized for up to 510 l/h at 15 °C inlet. The process loop connects to the G1 1/4 male pump connection; the pump delivers up to 96 l/min at 5.6 bar, with flow decreasing as external pressure drop rises.
Indoor use only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, IP20. Keep 50–200 cm clearance per side for maintenance, activate parking brakes or leveling feet before startup, and do not install in an ATEX zone. Cooling-water intake must be 5–20 °C.
The unit ships with a cable but without a plug (three-phase, <63 A). No process hoses or fittings are included; the pump connection is G1 1/4 male and cooling water G3/4 male. Shut off cooling water during downtimes, and have refrigerant-circuit work performed only by approved contractors.