Peter Huber Kältemaschinenbau Unichiller 070T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 070T-H

Peter Huber Kältemaschinenbau Unichiller 070T-H — Recirculating chillers with Noise level 66 dB.

The Unichiller 070T-H is an air-cooled recirculating chiller with a 7 kW cooling capacity at 15 °C and a -10…100 °C operating range. It is the higher-capacity modification in the Unichiller series, distinguished by its 2 kW heating stage and Pilot ONE controller. It serves externally closed process loops requiring precise temperature control.

Noise level
66 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 070T-H Current 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Noise level 66 dB(A)
Heating capacity 2 kW
Minimum fill volume 5 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...100 °C
Dimensions, Weight & Mechanical Design
Weight 234 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and Series Positioning

The Unichiller 070T-H is a free-standing, air-cooled recirculating chiller using R-449A refrigerant. Within the Unichiller series, it is positioned above the 030T and 040T siblings, offering greater cooling capacity (7 kW at 15 °C) and an integrated 2 kW heating stage. The 'H' suffix denotes heating capability, while the 'T' indicates the standard tower configuration. The unit is intended for externally closed applications only, as specified in the operation manual.

Performance and Media Behavior

Cooling capacity is strongly temperature-dependent: 7 kW at 15 °C, 4 kW at 0 °C, and 2.3 kW at -10 °C, measured at a 20 °C room temperature. The unit uses a C3 pump with G1 1/4 male connections and a fixed pump speed. The minimum filling volume is 5 l, and the expansion volume is 7.5 l. The temperature range of -10 to 100 °C is managed by the Pilot ONE controller, which supports Pt100 external sensor connection.

Installation, Control and Operation

The chiller requires a 400 V three-phase 50 Hz supply and is delivered with a cable but without a plug for final connection. It is rated IP20 and must be installed indoors in a dry location with at least 300 lx illuminance. Wall clearance of 50 to 200 cm per side is required for maintenance. The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces. The power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto), and a Process Safety mode is available for exothermic processes.

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for service access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access and adequate air flow for the air-cooled condenser. Insufficient clearance can lead to overheating and reduced cooling capacity.

Unichiller Operation Manual, Section on Installation
Electrical
Verify supply voltage and frequency tolerances separately

The unit accepts ±5% voltage and ±2% frequency tolerance, but these must not be combined in opposite directions. Ensure the site supply meets both conditions simultaneously to avoid compressor damage or invalid operation.

Unichiller 070T-H Technical Data, Specifications
Electrical
Three-phase units below 63 A are supplied without plug

The chiller is delivered with a cable but no plug. The customer must provide a suitable three-phase connection (400 V, 50 Hz) with a current rating below 63 A, and ensure the final connection complies with local electrical codes.

Unichiller 070T-H Technical Data, Specifications
Process Limits
Cooling capacity derates at lower temperatures and higher ambient

Cooling capacity is 7 kW at 15 °C, 4 kW at 0 °C, and 2.3 kW at -10 °C, based on a 20 °C room temperature. If the ambient temperature exceeds 20 °C, the available cooling capacity may drop further. Size the unit for the actual duty point and expected ambient conditions.

Unichiller 070T-H Technical Data, Specifications
Process Limits
Use only for externally closed applications

The unit is designed exclusively for externally closed loops. Do not connect it to open tanks or systems where the thermal fluid is exposed to atmosphere, as this can cause contamination, evaporation, and poor temperature control.

Unichiller Operation Manual, Function Description
Safety
Configure power-failure response for safe process restart

The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For processes where automatic restart is hazardous, select OFF/Standby; for critical cooling, consider Process Safety mode which releases maximum cooling capacity on overtemperature.

Unichiller Operation Manual, Setup Mode
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (GHS H280), and handling requires appropriate ventilation and safety precautions.

Unichiller Operation Manual, Maintenance

Selection & Integration Q&A

When should I choose the Unichiller 070T-H over the 030T or 040T siblings?

Choose the 070T-H when you need higher cooling capacity (7 kW at 15 °C) and integrated heating (2 kW). The 030T and 040T offer lower capacity and no heating. The 070T-H is also quieter (66 dB(A)) than the 030T (69 dB(A)). Select siblings for lower loads or where heating is not required.

What are the requirements for connecting the process loop to the 070T-H?

The pump connection is G1 1/4 male, so mating hoses or fittings must have a G1 1/4 female thread. Use hoses rated for the temperature range (-10 to 100 °C) and compatible with the thermal fluid. Ensure the external loop volume plus the 5 l minimum fill and 7.5 l expansion volume are accommodated.

What installation environment and duty-cycle limits apply to the 070T-H?

Install indoors only, with illuminance at least 300 lx and altitude up to 2000 m. Ambient temperature must be 5–40 °C. Maintain 50–200 cm side clearance for maintenance. The unit is IP20, so keep it dry. It is air-cooled, so ensure adequate ventilation for heat rejection.

What maintenance and accessories are required, and what is not included?

The unit is delivered with a cable but without a plug; the customer must provide the three-phase connection. Refrigerant circuit work must be by certified contractors. Regular maintenance includes checking refrigerant charge and room size per manual. No specific maintenance intervals are given in the provided facts; consult the manual for details.