Peter Huber Kältemaschinenbau Unichiller GL 100w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 100w

Peter Huber Kältemaschinenbau Unichiller GL 100w — Recirculating chillers.

The Unichiller GL 100w is a water-cooled recirculating chiller with a 20 kW cooling capacity at 10 °C and a -20…40 °C temperature range, controlled by the Pilot ONE. It is designed for externally closed applications and requires a 400 V three-phase supply and cooling water at 1–6 bar differential pressure. This modification is distinguished within the series by its high capacity and water-cooled condenser, making it suitable for large heat loads in industrial processes.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller GL 100w Current —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Expansion volume 10 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Current consumption 7 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 20 kW (at 10 °C)
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and positioning

Behavior under load and media

Installation, connection, and control

Integration & Operating Notes

Electrical
Three-phase supply with 3x16 A fuse required

The chiller requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse. The unit is delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%; certain combinations are not allowed.

Datasheet, page 1 and 2
Cooling & Media
Cooling water inlet must stay between 5 °C and 20 °C

The cooling water inlet temperature must be kept between 5 °C and 20 °C. Higher temperatures reduce cooling capacity and may require increased flow. Pipes must be insulated if below the condensation point. Cooling water differential pressure must be 1–6 bar.

Datasheet, page 2
Mounting
Maintain 50–200 cm wall clearance per side

A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit must be installed indoors with at least 300 lx illuminance and up to 2000 m altitude. Leveling feet must be activated and parking brakes applied before operation.

Operation Manual, page 25 and 31
Process Limits
Only for externally closed applications

The unit is designed exclusively for externally closed applications. It is not explosion-protected and must not be installed or started up inside an ATEX zone. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.

Operation Manual, page 34 and 13
Maintenance
Refrigerant circuit work requires certified contractors

Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The unit uses R-744 refrigerant (A1, H280), requiring appropriate handling precautions. Cooling water supply must be shut off even for short downtimes such as overnight.

Operation Manual, page 17 and 28
Control & I/O
Configure power-failure response for process safety

The Pilot ONE controller allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Operation Manual, page 18 and 19

Selection & Integration Q&A

When should I choose the Unichiller GL 100w over a sibling model like the Unichiller 030T?

Choose the GL 100w when you need high cooling capacity (20 kW at 10 °C) and a water-cooled condenser, and when a cooling water supply is available. The 030T is air-cooled and has lower capacity, making it suitable for smaller loads or where cooling water is not available.

What are the requirements for the cooling water supply and electrical connection when integrating this chiller?

Cooling water must be supplied at 1–6 bar differential pressure, 5–20 °C inlet, with G1/2 male connections. Electrical supply is 400 V 3~ 50 Hz with a 3x16 A fuse; the unit comes with a cable but no plug, so a suitable plug must be provided. Voltage tolerance is ±5% with frequency tolerance ±2%.

What are the mounting and environmental limits for installing the Unichiller GL 100w?

Install indoors only, with at least 300 lx illuminance and up to 2000 m altitude. Maintain 50–200 cm wall clearance per side for maintenance. Ambient temperature must be 5–40 °C. The unit is not explosion-protected and must not be installed in an ATEX zone.

What maintenance and service items are required, and what is not included in the base delivery?

Refrigerant circuit work must be performed by approved refrigeration contractors. Cooling water supply must be shut off during downtimes. The base delivery includes a cable but no plug; a suitable plug must be provided on site. No other accessories or consumables are specified in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 100w Technical Data
Datasheet ·30,4 КБ ·2 pp.
View PDF