Peter Huber Kältemaschinenbau Unichiller GL 075
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 075

Peter Huber Kältemaschinenbau Unichiller GL 075 — Recirculating chillers.

The Unichiller GL 075 is a recirculating chiller with a 7.5 kW cooling capacity at 20 °C, rated for -20 to 40 °C and a 460 V three-phase 60 Hz supply. It uses a fixed-speed pressure pump delivering up to 50 l/min at 3.5 bar through G1 1/4 male connections. It is intended for externally closed cooling circuits in industrial processes.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller GL 075 Current —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must match 460 V three-phase 60 Hz within tight tolerances

The chiller requires a 460 V three-phase 60 Hz supply. Voltage tolerance is ±5% and frequency tolerance ±2%, but not all combinations are allowed. Verify site power quality before installation; a transformer or different supply may be needed.

Datasheet, page 2, 'Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%'
Electrical
External branch circuit protection per UL 489 required

The electrical design must include a UL 489 listed breaker for branch circuit protection. Three-phase devices below 63 A are delivered with cable but without a plug, so the installer must provide a suitable connection.

Datasheet, page 2, 'External branch circuit protection according to UL 489 required.'
Mounting
Maintain 50–200 cm wall clearance per side for maintenance

A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be unscrewed/activated and parking brakes applied before switching on the unit.

Operation Manual, page 25, 'a wall clearance of 50 to 200 cm is required on either side.'
Control & I/O
Configure power-failure response to match process restart needs

The Pilot ONE controller allows power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For processes where automatic restart is unsafe, select OFF/Standby; for continuous cooling, Power Failure Auto may be appropriate.

Operation Manual, page 19, 'Power Failure Auto'
Control & I/O
Process Safety mode releases maximum cooling on overtemperature

In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. This is critical for exothermic processes where cutting cooling would cause runaway temperatures.

Operation Manual, page 18, 'The operating mode Process safety releases the maximum cooling capacity as soon as the set trigger value of the overtemperature protection is detected by the controller.'
Environment
Not explosion-protected; ATEX zones prohibited

The unit must not be installed or started up inside an ATEX zone because it is not explosion-protected. Ensure the installation location is classified as non-hazardous.

Operation Manual, page 13, 'Do NOT install or start up the temperature control unit within an ATEX zone.'
Environment
Class A EMC may require measures in sensitive environments

The equipment is Class A, Group 1 for electromagnetic disturbance to EN55011. In residential or sensitive environments, additional EMC measures may be necessary.

Datasheet, page 2, 'Classification (disturbance) to EN55011: Class A, Group 1'

Selection & Integration Q&A

When should I choose the Unichiller GL 075 over sibling models like the 030T or 030Tw?

Choose the GL 075 when you need 7.5 kW cooling at 20 °C and have a 460 V three-phase 60 Hz supply. Siblings like the 030T offer lower capacity and different noise levels (69 dB(A) vs 61 dB(A) for 030Tw); select based on required cooling capacity and electrical configuration.

What are the requirements for connecting the GL 075 to my process cooling loop?

The pump connection is G1 1/4 male; mating hoses or fittings must match this thread. The external circuit must be closed and its pressure drop checked against the pump's maximum delivery of 50 l/min at 3.5 bar. Do not exceed maximum tightening torques and counterhold connections.

What maintenance and service items are required, and what is not included in base delivery?

Refrigerant circuit work must be by approved contractors. Monitor fluid level via digital indicator. Three-phase devices below 63 A are delivered with cable but without a plug; installer must provide a UL 489 breaker and suitable plug. No other accessories are specified in the provided facts.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 075 — Technical Data
Datasheet ·30,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF