Peter Huber Kältemaschinenbau Unichiller 070Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 070Tw-H

Peter Huber Kältemaschinenbau Unichiller 070Tw-H — Recirculating chillers.

The Unichiller 070Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a fixed-speed C3 pump, rated for -10 to 100 °C in externally closed circuits. It is the 'Tw-H' modification of the 070 platform, adding heating and water cooling for processes needing both heat rejection and controlled warm-up. It serves as a central temperature-control unit for closed process loops with heat loads up to 7 kW at 15 °C.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 070Tw-H Current —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Heating capacity 2 kW
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...100 °C
Dimensions, Weight & Mechanical Design
Weight 212 kg
Electrical, Power & Drive Specifications
Current consumption 12 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 264 л/час (at 15°C, flow -10°C)
at 15°C, flow 0°C: 330 l/h · at 15°C, flow 15°C: 474 l/h
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and positioning

Behavior under load

Installation, control and operation

Integration & Operating Notes

Electrical
Three-phase supply needs a site-fitted plug and 3x16 A protection

The unit draws 12 A on a 400 V 3~ 50 Hz supply and is delivered with cable but without a plug because it is below 63 A. The installer must provide a matching plug and a 3x16 A protective device; voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not permitted.

Unichiller 070Tw-H Datasheet, Specifications
Cooling & Media
Cooling water must stay within 5–20 °C and 1.5–6 bar differential

Inlet temperature outside 5–20 °C degrades capacity or causes condensation; pipes below the dew point must be insulated. Differential pressure must be at least 1.5 bar for flow and must not exceed 6 bar. The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water chemistry must be compatible to avoid corrosion and fouling.

Unichiller 070Tw-H Datasheet, Specifications
Process Limits
Closed-loop only; open baths are not supported

The chiller is designed for externally closed applications and uses an atmospheric open expansion tank. Minimum filling capacity is 5.9 l with 7.5 l expansion volume, which sets the minimum external circuit volume. Open vessels or unsealed circuits will not work with this hydraulic configuration.

Unichiller 070Tw-H Datasheet, Overview; Unichiller Operation Manual, Function Description
Maintenance
Refrigerant circuit work restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit is not explosion-protected and must not be installed or started inside an ATEX zone. Refrigerant charge and room size must be checked at installation to avoid exceeding the limit per m³ of ambient air.

Unichiller Operation Manual, Maintenance and Installation

Selection & Integration Q&A

When should I choose the 070Tw-H instead of a sibling such as the 030Tw or 040T?

Pick the 070Tw-H when the process needs both heating (2 kW) and water cooling with up to 7 kW at 15 °C, and when a water-cooled condenser is acceptable. The 030T/030Tw/040T siblings are air-cooled and have no heater; they suit lower heat loads where a cooling-water supply is unavailable.

What must be verified in the cooling-water supply and hydraulic loop before connecting this chiller?

Provide 5–20 °C cooling water at 1.5–6 bar differential through G1/2 male connections, with water chemistry compatible with copper, 1.4401 stainless, PA, PPE, PTFE and EPDM. The process loop must be externally closed with at least 5.9 l filling volume and 7.5 l expansion; pump connections are G1 1/4 AG male and must be counterheld at the stated torque.

What installation environment and duty-cycle limits apply to the 070Tw-H?

Indoors only, 5–40 °C ambient, ≥300 lx, ≤2000 m altitude, IP20, with 50–200 cm side clearance for maintenance. It is not ATEX-rated. Shut off cooling water even overnight; set leveling feet or parking brakes before start-up. Cooling capacity falls from 7 kW at 15 °C to 2.5 kW at -10 °C, so duty must be rated at the actual setpoint.

What is not included in base delivery and what service/consumables should be planned?

The three-phase cable is supplied without a plug, so a site plug and 3x16 A protection are needed. Refrigerant-circuit service must be by approved contractors; no consumables or accessories are listed in the verified facts. Plan cooling-water shutoff and water treatment as ongoing maintenance items.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 070Tw-H Datasheet
Datasheet ·79,5 КБ ·2 pp.
View PDF