Peter Huber Kältemaschinenbau Unichiller 070Tw-H
Peter Huber Kältemaschinenbau Unichiller 070Tw-H — Recirculating chillers.
The Unichiller 070Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a fixed-speed C3 pump, rated for -10 to 100 °C in externally closed circuits. It is the 'Tw-H' modification of the 070 platform, adding heating and water cooling for processes needing both heat rejection and controlled warm-up. It serves as a central temperature-control unit for closed process loops with heat loads up to 7 kW at 15 °C.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 070Tw-H Current | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Heating capacity | 2 kW |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 212 kg |
| Current consumption | 12 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 264 л/час (at 15°C, flow -10°C) at 15°C, flow 0°C: 330 l/h · at 15°C, flow 15°C: 474 l/h |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and positioning
- Water-cooled refrigeration circuit using R-449A (A1, H280; charge 1.35 kg, CO2 equivalent 1.9 t) with an atmospheric open expansion tank.
- Fixed-speed pump type C3, maximum delivery 84 l/min, falling to 7 l/min at 5.5 bar; pump connection G1 1/4 AG male.
- Within the Unichiller series, the 070Tw-H is the heated, water-cooled variant; sibling 030T/030Tw/040T models are air-cooled and lack the 2 kW heater. The 'P' suffix models (not this unit) are intended for high-pressure-drop loops.
- IP20 protection, 212 kg, Pilot ONE controller.
Behavior under load
- Cooling capacity is strongly temperature-dependent: 7 kW at 15 °C process temperature, 2.5 kW at -10 °C. Heat load must be matched at the actual setpoint, not at nominal conditions.
- Cooling water consumption rises as process temperature falls: 474 l/h at 15 °C flow, 330 l/h at 0 °C, 264 l/h at -10 °C (all at 15 °C cooling-water inlet).
- The 2 kW heater supports warm-up and hold at temperatures above ambient; the unit is not a steam or high-temperature bath.
Installation, control and operation
- Requires 400 V 3~ 50 Hz supply, 12 A current consumption, 3x16 A fusing; three-phase devices below 63 A are delivered with cable but without a plug, so a site plug must be provided.
- Cooling water: G1/2 male connections, inlet temperature 5–20 °C, differential pressure 1.5–6 bar; shut off supply even for overnight downtime.
- Pilot ONE controller with Ethernet, USB host/device and RS232; external Pt100 sensor input; power-failure response configurable as OFF/Standby, ON or Power Failure Auto; Process Safety mode releases maximum cooling on overtemperature trigger.
- Indoor installation only, 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m; 50–200 cm wall clearance per side for maintenance; leveling feet/parking brakes must be set before start-up.
Integration & Operating Notes
The unit draws 12 A on a 400 V 3~ 50 Hz supply and is delivered with cable but without a plug because it is below 63 A. The installer must provide a matching plug and a 3x16 A protective device; voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not permitted.
Inlet temperature outside 5–20 °C degrades capacity or causes condensation; pipes below the dew point must be insulated. Differential pressure must be at least 1.5 bar for flow and must not exceed 6 bar. The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water chemistry must be compatible to avoid corrosion and fouling.
The chiller is designed for externally closed applications and uses an atmospheric open expansion tank. Minimum filling capacity is 5.9 l with 7.5 l expansion volume, which sets the minimum external circuit volume. Open vessels or unsealed circuits will not work with this hydraulic configuration.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit is not explosion-protected and must not be installed or started inside an ATEX zone. Refrigerant charge and room size must be checked at installation to avoid exceeding the limit per m³ of ambient air.
Selection & Integration Q&A
Pick the 070Tw-H when the process needs both heating (2 kW) and water cooling with up to 7 kW at 15 °C, and when a water-cooled condenser is acceptable. The 030T/030Tw/040T siblings are air-cooled and have no heater; they suit lower heat loads where a cooling-water supply is unavailable.
Provide 5–20 °C cooling water at 1.5–6 bar differential through G1/2 male connections, with water chemistry compatible with copper, 1.4401 stainless, PA, PPE, PTFE and EPDM. The process loop must be externally closed with at least 5.9 l filling volume and 7.5 l expansion; pump connections are G1 1/4 AG male and must be counterheld at the stated torque.
Indoors only, 5–40 °C ambient, ≥300 lx, ≤2000 m altitude, IP20, with 50–200 cm side clearance for maintenance. It is not ATEX-rated. Shut off cooling water even overnight; set leveling feet or parking brakes before start-up. Cooling capacity falls from 7 kW at 15 °C to 2.5 kW at -10 °C, so duty must be rated at the actual setpoint.
The three-phase cable is supplied without a plug, so a site plug and 3x16 A protection are needed. Refrigerant-circuit service must be by approved contractors; no consumables or accessories are listed in the verified facts. Plan cooling-water shutoff and water treatment as ongoing maintenance items.