Peter Huber Kältemaschinenbau Unichiller 030Tw
Peter Huber Kältemaschinenbau Unichiller 030Tw — Recirculating chillers with Noise level 61 dB.
The Unichiller 030Tw is a water-cooled recirculating chiller for externally closed circuits, rated -20 to 40 °C with a 26 l/min pump and G3/4 male process connections. Its 61 dB(A) acoustic level distinguishes it from the 69 dB(A) 030T sibling, while the -H variant shares the same noise figure. It serves as a central cooling source for closed process loops requiring a stable temperature window.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 030Tw Current | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 61 dB(A) |
| Expansion volume | 4 l |
| Maximum delivery flow rate | 26 l/min (pressure pump max; also 3 bar) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 137 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 25 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 65 л/час · at 15°C, flow 0°C: 99 l/h · at 15°C, flow 15°C: 99 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-449A (A1, H280; 0.7 kg charge, 1.0 t CO2 equivalent) with a plate-type evaporator and an atmospheric open expansion tank (4 l expansion volume, 2.5 l minimum filling capacity).
- Process side: pump type B, non-adjustable speed, maximum delivery 26 l/min (8 l/min at 2.0 bar), G3/4 AG male pump connections; cooling-water side: G1/2 male connections.
- Within the Unichiller series, the 030Tw is the water-cooled 61 dB(A) variant; the 030T is air-cooled at 69 dB(A), and the 030Tw-H shares the 61 dB(A) figure. The 040T and 040T-H are larger siblings at 69 dB(A).
- Controller: Pilot ONE with Ethernet, USB host/device and RS232; external Pt100 sensor input; protection IP20; safety class I / NFL.
Behavior under load and media conditions
- Temperature range -20 to 40 °C; ambient 5 to 40 °C. Cooling water intake must be 5 to 20 °C; warmer water reduces cooling power and may increase the required cooling-water flow rate.
- Cooling-water consumption at 15 °C supply: 99 l/h at 0 °C and at 15 °C process flow, 65 l/h at -10 °C, 25 l/h at -20 °C. Minimum cooling-water differential pressure 1 bar, maximum 6 bar.
- The unit is designed for externally closed applications only; models with 'P' in the name are the ones suited to high pressure-drop loops, so the 030Tw should not be applied where the process side demands high pressure drop.
Installation, control and operation
- Three-phase 400 V 3~ 50 Hz supply, maximum current 4.5 A, fusing 3x10 A; cable supplied without plug for devices below 63 A. Voltage tolerance ±5% only with ±2% frequency; the combination -5% voltage with +2% frequency is not permitted.
- Power-failure response configurable via Pilot ONE as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
- Indoor installation only, illuminance at least 300 lx, altitude up to 2000 m; wall clearance 50 to 200 cm per side for maintenance. Leveling feet or parking brakes must be set before start-up. Cooling water must be shut off even for overnight downtime.
Integration & Operating Notes
The hydraulic circuit must be externally closed because the unit uses an atmospheric open expansion tank and is designed for externally closed applications. Open baths or vented loops will not match the expansion and filling concept (2.5 l minimum fill, 4 l expansion volume).
The facility cooling-water supply must deliver 5 to 20 °C intake at a differential pressure of at least 1 bar and no more than 6 bar at the G1/2 male connection. Warmer water lowers cooling power and can raise the required flow rate beyond the rated 99 l/h at 15 °C.
The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only cooling water compatible with this material set may be used; aggressive or chloride-rich water risks corrosion of the copper and stainless components.
The 400 V 3~ 50 Hz supply must stay within ±5% voltage only when frequency is within ±2%. The combination of -5% voltage with +2% frequency is explicitly not allowed, so supply quality should be verified before commissioning to avoid invalid operating conditions.
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug, so the installer must provide a suitable plug or hard-wire the connection on site. The supply must be fused at 3x10 A; maximum unit current is 4.5 A.
Installation is restricted to indoors at illuminance of at least 300 lx and altitude up to 2000 m, with 50 to 200 cm wall clearance per side for maintenance. Leveling feet must be unscrewed or parking brakes applied before the unit is switched on to prevent movement and crushing injury.
The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Refrigerant charge (0.7 kg R-449A) and room size must be checked at installation so the refrigerant limit per m³ of ambient air is not exceeded.
Electrical lines routed directly below a water-cooled unit must be protected against liquid ingress from condensation. Pump connections must be counterheld and the maximum tightening torques must not be exceeded during hose installation to avoid leaks and thread damage.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For unattended processes the chosen behavior determines whether the pump and cooling restart automatically after a mains outage, which must be matched to the process risk.
In Process Safety mode the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is relevant for exothermic processes where cutting cooling would cause runaway temperatures; the mode must be selected deliberately.
Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration and air-conditioning system contractors. The refrigerant R-449A is classified A1 with GHS hazard H280, and the 0.7 kg charge corresponds to 1.0 t CO2 equivalent, which may trigger leak-checking obligations.
The cooling-water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion during idle periods. This applies to the G1/2 cooling-water connection and should be part of the daily shutdown procedure.
Selection & Integration Q&A
Pick the 030Tw when a water-cooled unit at 61 dB(A) is required and the process loop is externally closed without high pressure drop. The 030T is air-cooled at 69 dB(A); the 030Tw-H shares the 61 dB(A) figure but is a different modification. For high pressure-drop loops, a 'P' model is specified instead.
Provide 5–20 °C cooling water at 1–6 bar differential pressure to the G1/2 male connection, sized for 99 l/h at 15 °C supply (65 l/h at -10 °C, 25 l/h at -20 °C process flow). Use only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, and insulate pipes if condensation can form.
Indoors only, at least 300 lx illuminance, altitude up to 2000 m, ambient 5–40 °C, with 50–200 cm wall clearance per side. Not permitted in ATEX zones. Leveling feet or parking brakes must be set before start-up, and cooling water must be shut off even overnight.
The three-phase cable is supplied without a plug, so a plug or hard-wiring must be provided on site; fusing is 3x10 A. Refrigerant-circuit work must be done by approved refrigeration contractors. Minimum filling capacity is 2.5 l with 4 l expansion volume, so the external loop must be filled and vented accordingly.