Peter Huber Kältemaschinenbau Unistat 515w Dynamisches
Peter Huber Kältemaschinenbau Unistat 515w Dynamisches — Temperature control units.
The Unistat 515w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 250 °C with 6 kW heating. It sits between the 510w and 520w siblings in the Unistat 5xx series, distinguished by its water-cooled condenser and dynamic temperature control capability. It is intended for precise temperature control of externally closed or open applications in industrial processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 515w Dynamisches Current | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...250 °C |
| Weight | 181 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 504 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The Unistat 515w is a water-cooled (w suffix) free-standing unit in the Unistat 5xx series; siblings include the 510w, 520w, 525 and 525w, all sharing the dynamic temperature control concept.
- The pump is magnetically coupled and made of stainless steel, with adjustable speed and a maximum delivery of 112 l/min (54 l/min at 1.0 bar).
- The cooling water circuit uses G1/2 male connections and requires a differential pressure of at least 1 bar and no more than 6 bar.
- The unit weighs 181 kg and measures 460x554x1455 mm, requiring adequate floor space and handling equipment.
Thermal performance and media
- The temperature range is -50 to 250 °C, with a heating capacity of 6 kW.
- Cooling power depends strongly on temperature and heat transfer fluid: with ethanol, it drops from 5.3 kW at 0 °C to 0.25 kW at -50 °C.
- The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 3.1 t, relevant for environmental compliance and leak reporting.
- The unit is rated safety class III / FL, allowing flammable heat transfer fluids with a fire point of at least 65 °C; ethanol requires supervised operation.
- Cooling water inlet temperature must be between 5 °C and 20 °C; pipes must be insulated if below the condensation point.
Control, installation and operation
- The Pilot ONE controller with POKO ONE control output provides a display resolution of 0.01 K and interfaces: Ethernet, USB (Host and Device), RS232, and a Pt100 external sensor connection.
- Power supply is 400 V 3~ 50 Hz with a 3x16 A fuse; three-phase devices under 63 A are delivered with cable but without a plug.
- Leakage current exceeds 3.5 mA, so residual current protection must be selected accordingly.
- The unit must not be installed in an ATEX zone; ambient temperature must be 5–40 °C, and protection rating is IP20.
- For externally closed applications, the expansion vessel must remain unlocked; for externally open applications, it must be locked to prevent fluid loss.
- A condensation water / heat transfer fluid drip tray must be installed below the unit.
Integration & Operating Notes
The unit's leakage current exceeds 3.5 mA, so residual current devices must be chosen with this in mind to avoid nuisance tripping. The electrical installation must also provide a 400 V three-phase supply with a 3x16 A fuse and a suitable plug on site, as the delivered cable has no plug.
Operating outside this range reduces cooling power or causes condensation. The cooling water circuit requires a differential pressure of at least 1 bar and no more than 6 bar, with G1/2 male connections. Pipes must be insulated if below the condensation point, and the circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible.
Without an Ex px cabinet the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable. Ambient temperature must remain between 5 °C and 40 °C.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss. This setting is critical for safe operation and must be verified before commissioning.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger, ensuring heat extraction under exothermic conditions.
Selection & Integration Q&A
The 515w offers 6 kW heating and a water-cooled condenser, positioned between the 510w and 520w. Choose the 515w when its heating capacity and water cooling match your load; select the 510w for lower capacity or the 520w for higher capacity, depending on your process requirements.
Cooling water: G1/2 male connections, differential pressure 1–6 bar, inlet temperature 5–20 °C. Electrical: 400 V 3~ 50 Hz, 3x16 A fuse, cable without plug, leakage current >3.5 mA so choose RCD accordingly.
Ambient temperature must be 5–40 °C. Do not install in an ATEX zone. Unit must be level, with drip tray underneath. Allow 2 hours acclimation after moving between cold and warm environments before power-on.
Periodically check hoses and connections for leaks; shut off cooling water during downtimes. The power cable is delivered without a plug. A drip tray and suitable cooling water hoses are not included and must be provided on site.