Peter Huber Kältemaschinenbau Unistat 540w Dynamisches
Peter Huber Kältemaschinenbau Unistat 540w Dynamisches — Temperature control units.
The Unistat 540w Dynamisches is a water-cooled, free-standing temperature control unit in the Unistat 5xx series, covering -55 to 250 °C with 24 kW heating and a magnetically coupled stainless-steel pump. Its 55 A three-phase draw and 1480 l/h cooling-water demand distinguish it from smaller siblings, suiting high-capacity process temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 540w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 472 kg |
| Current consumption | 55 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1480 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 5xx chassis with water-cooled refrigeration machine (R-452A, 3.6 kg, GWP 2141, 7.7 t CO2e) and CFC/HCFC-free circuit.
- Magnetically coupled stainless-steel pump with M38x1.5 male connection (bore shape Y, 60°, DIN 3863); speed adjustable.
- Within the Dynamisches series, the 540w is the highest-capacity water-cooled variant listed; siblings 510w, 515w, 520w and 525 Dynamisches are smaller. The 'w' suffix denotes water cooling; no suffix would be air cooled.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB host/device and RS232; external Pt100 sensor input.
Behavior and media
- Working range -55 to 250 °C; heating capacity 24 kW. No cooling-capacity curve points were supplied, so capacity at intermediate temperatures cannot be stated.
- Heat transfer fluid must have kinematic viscosity ≤100 mm²/s. Safety class III/FL requires FL-rated fluid with fire point ≥65 °C; ethanol only under supervision.
- Cooling water: 5–20 °C intake, 1–4 bar differential pressure, max 6 bar, 1480 l/h at 15 °C water / 0 °C flow; G3/4 male connection. Cooling power falls as water temperature rises. Insulate cooling-water pipes when below condensation point. Circuit materials include copper, 1.4401, MS, PA, PPE, PTFE, EPDM.
Installation, control and operation
- 400 V 3~ 50 Hz, max 55 A, 3x63 A fuse; delivered with cable but without plug (three-phase <63 A). Leakage current >3.5 mA affects RCD selection. Voltage tolerance ±5% only with simultaneous frequency ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
- IP20; ambient 5–40 °C; not for ATEX zones; not for medical devices or direct foodstuff temperature control.
- Pilot ONE offers Process Safety mode (releases maximum cooling at OT trigger, requires compressor Permanently ON) and configurable power-failure response (Off/Standby or auto-restart).
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
Integration & Operating Notes
The unit requires a 400 V 3~ 50 Hz supply with a 3x63 A fuse and draws up to 55 A. Cable cross-section, upstream protection and the site transformer must be sized for this; the unit arrives with a cable but no plug, so a suitable plug must be provided on site.
Leakage current exceeds 3.5 mA, which affects the choice of residual-current protection and earthing design. The electrical installer must select RCD type and rating accordingly rather than assuming a standard 30 mA device is suitable.
The unit is designed for cooling water intake between 5 °C and 20 °C; cooling power drops as water temperature rises. Differential pressure must be 1–4 bar and static pressure must not exceed 6 bar. Consumption is 1480 l/h at 15 °C water and 0 °C flow, so the supply pump and piping must sustain that flow.
When cooling water temperatures fall below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation, corrosion and electrical risk. The cooling circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
Safety class III/FL requires heat transfer fluid with a fire point of at least 65 °C; ethanol is permitted only under supervised operation. Maximum permissible kinematic viscosity is 100 mm²/s, which also bounds pump performance and the choice of fluid at low temperature.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. The wrong state at commissioning is a direct damage risk, so verify before filling.
The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults. Transport upright only and empty.
Pilot ONE can be set to Off/Standby (default, manual start after power returns) or ON/Temperature control active (automatic restart, with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable for the process; Process Safety mode additionally requires the compressor set to Permanently ON.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be checked periodically for leaks and quality and replaced as required.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The circuit is factory-tested tight to 10⁻⁴ Pa m³/s or better, but refrigerant handling, leak monitoring and shipping classification (UN 3358, 3.6 kg charge) remain regulated tasks.
Selection & Integration Q&A
Pick the 540w when the process needs the highest heating capacity (24 kW) and the -55…250 °C range at the largest scale in this series; the smaller siblings suit lower loads. The 540w is water-cooled ('w'), so it also requires a 1480 l/h cooling-water supply, unlike air-cooled variants.
A 400 V 3~ 50 Hz supply protected at 3x63 A with a site-fitted plug (cable supplied, plug not), RCD design allowing >3.5 mA leakage, and a cooling-water supply of 5–20 °C at 1–4 bar differential, max 6 bar, delivering 1480 l/h through a G3/4 male connection.
Install indoors, IP20, ambient 5–40 °C, never in an ATEX zone; transport upright and empty, allow ~2 h acclimation after temperature change, activate leveling feet/parking brakes, and provide a drip tray. Not permitted for medical-device or direct foodstuff temperature control.
Base delivery excludes the power plug, mating ball sockets/sleeve nuts per DIN 3863/3870, heat transfer fluid and cooling-water hoses. Plan periodic hose/connection inspection and replacement, liquefier-fin and strainer cleaning, fluid inspection/replacement, and refrigerant work only by approved contractors.