Peter Huber Kältemaschinenbau Unistat 545w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 545w Dynamisches

Peter Huber Kältemaschinenbau Unistat 545w Dynamisches — Temperature control units.

The Unistat 545w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C. It is the largest of the Unistat 5xxw series, distinguished by 24 kW heating and 35 kW cooling with Thermooil. It serves externally closed or open circuits requiring high heating and cooling capacity with precise Pt100 control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 545w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 479 kg
Electrical, Power & Drive Specifications
Current consumption 58 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 1630 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ 50 Hz with 3x63 A fuse

The unit draws up to 58 A per phase; upstream protection and cable cross-section must be sized for this. Three-phase devices below 63 A are supplied with cable but without a plug, so the final connection must be provided on site by an electrician.

Datasheet, Technical Specifications, page 2
Electrical
Voltage and frequency tolerances are interdependent

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed. Site power quality must be verified before connection to avoid invalidating warranty and safe operation.

Datasheet, Technical Specifications, page 2
Cooling & Media
Cooling water intake must stay between 5 °C and 20 °C

Out-of-range cooling water reduces cooling capacity and risks condensation; pipes must be insulated if water temperature can fall below the dew point. The circuit requires 1630 l/h at 15 °C with 1–4 bar differential pressure and max 6 bar static pressure.

Datasheet, Technical Specifications, page 2
Mounting
Unit must not be installed in an ATEX zone

Without an Ex px cabinet the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.

Operation Manual, Section 1.3.3, page 15-16
Mounting
Transport upright and allow 2 hours acclimation before power-on

Horizontal transport damages the compressor; only an emptied unit may be moved. When changing from a cold to a warm environment (or vice versa), allow about 2 hours acclimation before turning on to prevent condensation-induced electrical faults.

Operation Manual, Commissioning, page 22 and 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active.

Operation Manual, Setup Mode, page 19
Control & I/O
Power failure auto-restart behavior must be configured

After a power outage or when switching on, the response can be set via Pilot ONE: 'Off/Standby' (default, temperature control started only by manual input) or 'ON/Temperature control active' (temperature control always started after turning on). Choose according to whether unattended restart is acceptable for the process.

Operation Manual, Section 1.8.1, page 21
Maintenance
Shut off cooling water supply even for short downtimes

Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Operation Manual, Commissioning, page 29
Maintenance
Coolant circuit work only by approved refrigeration contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Refrigerant R-452A has GWP 2141 and 7.7 t CO2 equivalent, which affects leak detection, F-gas reporting and room ventilation obligations.

Operation Manual, Section 1.4.2, page 19; Datasheet, page 1

Selection & Integration Q&A

When should I choose the Unistat 545w Dynamisches instead of a smaller sibling like the 510w or 525 Dynamisches?

Choose the 545w when you need the highest heating capacity (24 kW) and cooling capacity (35 kW with Thermooil) in the 5xxw series, or when your process requires the larger pump delivery (196 l/min max). Smaller siblings like the 510w or 525 have lower capacities and may suffice for less demanding loads; the 545w is the largest water-cooled model in this series.

What are the requirements for the cooling water supply and electrical connection when integrating this unit?

Cooling water: G3/4 male connection, 1630 l/h at 15 °C, 1–4 bar differential pressure, max 6 bar static, intake 5–20 °C. Electrical: 400 V 3~ 50 Hz, max 58 A, 3x63 A fuse; cable supplied without plug, final connection by electrician. Leakage current >3.5 mA requires protective earthing and RCD selection.

What are the mounting and ambient limits for the Unistat 545w Dynamisches?

Ambient 5–40 °C, IP20 enclosure (no wet or dusty environments), not in ATEX zones. Transport upright only; allow 2 hours acclimation before power-on. Leveling feet must be activated and parking brakes engaged before commissioning. Weight 479 kg and dimensions 730x804x1738 mm require floor loading and access route verification.

What maintenance is required and what accessories are not included in the base delivery?

Periodic checks: hoses and connections for leaks/quality, cleaning liquefier fins, cleaning strainer/dirt trap, heat transfer fluid inspection/replacement, function test of emergency stop. Cooling water supply must be shut off even overnight. Not included: final power plug (cable supplied without plug), drip tray (must be installed by responsible body), and any external Pt100 sensor or interface cables.