Peter Huber Kältemaschinenbau Unistat 610w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 610w Dynamisches

Peter Huber Kältemaschinenbau Unistat 610w Dynamisches — Temperature control units with Noise level 62 dB.

The Unistat 610w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -60 to 200 °C. It is distinguished within the Unistat 5xx/6xx series by its 6 kW heating capacity and 26 l expansion volume, serving closed or open external circuits requiring precise dynamic temperature control.

Noise level
62 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 610w Dynamisches Current 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 62 dB(A)
Control output POKO ONE
Expansion volume 26 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 360 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 400 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unistat 610w Dynamisches is a water-cooled (w suffix) free-standing unit in the Unistat 6xx family, using refrigerant R-452A (2.55 kg charge, 5.5 t CO2e) and a magnetically coupled stainless-steel pump with M30x1.5 male connections. Compared to sibling models such as the Unistat 510w (60 dB(A)) and 520w (61 dB(A)), this 610w delivers 6 kW heating and a 26 l expansion volume, with a noise level of 62 dB(A). The Pilot ONE controller with POKO ONE output provides Ethernet, USB host/device, and RS232 interfaces, plus a Pt100 external sensor input.

Thermal behavior and media constraints

Cooling capacity is strongly temperature-dependent: 2.6 kW at -40 °C falls to only 0.05 kW at -60 °C, so usable load near the lower limit is minimal. The published capacities are valid only at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure. The pump delivers 82 l/min at zero head but 29 l/min at 1.25 bar, so external circuit pressure drop directly reduces flow. Heat transfer fluid viscosity must not exceed 50 mm²/s, and the unit is rated safety class III/FL with adjustable overtemperature protection per DIN 12876.

Installation, connection and control

Electrical supply is 400 V 3~ 50 Hz with max current 15.5 A and 3x16 A fuse; three-phase units below 63 A are supplied with cable but without plug. Cooling water must be 5–20 °C, 1–6 bar differential pressure, with G1/2 male connections; consumption is 400 l/h at 15 °C. The unit is IP20 and must operate between 5 and 40 °C ambient, excluding ATEX zones. The Pilot ONE controller supports process safety mode (requires compressor permanently ON), power failure auto-restart, and adjustable overtemperature limits. For externally closed applications the expansion vessel must remain unlocked; for open applications it must be locked.

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ with 3x16 A fuse and no plug included

The unit draws up to 15.5 A per phase and requires a 3x16 A fuse; three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug and breaker must be provided on site. Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction (e.g., -5% voltage with +2% frequency is not allowed).

Datasheet, page 2; Operation Manual, Section 1.8
Cooling & Media
Cooling water must be 5–20 °C and 1–6 bar differential pressure

Cooling water intake temperature must be between 5 °C and 20 °C, with a minimum differential pressure of 1 bar and maximum 6 bar. Consumption is 400 l/h at 15 °C. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate pipes if condensation can occur.

Datasheet, page 2
Mounting
Do not install in ATEX zone; IP20 requires dry, 5–40 °C ambient

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. With IP20 protection, it must be operated between 5 °C and 40 °C ambient and cannot be placed in wet or hot environments. A drip tray for condensation water or heat transfer fluid must be installed below the unit.

Operation Manual, Section 1.3.3 and 1.4
Control & I/O
Process safety mode requires compressor set to permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power failure auto setting can resume temperature control automatically after an outage if it was active.

Operation Manual, Section 1.7.2 and page 20

Selection & Integration Q&A

When should I choose the Unistat 610w over the 510w or 520w siblings?

Choose the 610w when you need 6 kW heating and a 26 l expansion volume for larger external circuits; the 510w and 520w have lower heating capacities and smaller expansion volumes. The 610w also has a higher noise level (62 dB(A)) than the 510w (60 dB(A)) and 520w (61 dB(A)).

What are the requirements for the cooling water supply and the pump connections?

Cooling water must be 5–20 °C, 1–6 bar differential pressure, with G1/2 male connections and 400 l/h consumption at 15 °C. Pump connections are M30x1.5 male with Y-shaped bore per DIN 3863; use ball socket and sleeve nut per DIN 3863/3870 for leak-free hoses.

What are the ambient and duty-cycle limits for installation?

The unit is IP20 and must operate between 5 and 40 °C ambient, excluding ATEX zones. It must be transported and stored upright only, and allowed to acclimate for about 2 hours when moving between cold and warm environments before power-on. Leveling feet must be activated before commissioning.

What maintenance is required and what accessories are not included?

Periodically check hoses and connections for leaks and replace as needed; shut off cooling water even overnight. The unit is supplied with cable but without a plug, so a suitable plug must be provided on site. A drip tray must be installed below the unit. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 610w Technical Data
Datasheet ·74,8 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF