Peter Huber Kältemaschinenbau Unistat 645w Dynamisches
Peter Huber Kältemaschinenbau Unistat 645w Dynamisches — Temperature control units.
Water-cooled (w) Unistat 645w Dynamisches is a free-standing temperature control unit with magnetically coupled stainless-steel pump, covering -60 to 200 °C with 36 kW heating and 45 kW cooling at 0 °C. It is distinguished within the Unistat 5xx/6xx series by its 460 V 3~ 60 Hz supply and 5000 l/h cooling-water demand. It serves as a high-capacity source for externally closed process circuits requiring dynamic temperature ramps.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 645w Dynamisches Current | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Alarm output | optic, acoustic, relay |
| Control output | POKO ONE |
| Expansion volume | 30 l |
| Filling volume | 30 l |
| Heating capacity | 36 kW |
| Pump connection | flange DN32 PN10/40 EN1092-1 |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | flange DN32 PN10/40 EN1092-1 |
| Cooling water consumption | 5000 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The Unistat 645w Dynamisches is a water-cooled (suffix 'w') free-standing unit; other models in the series (Unistat 510w, 515w, 520w, 525 Dynamisches) differ in cooling capacity and noise level (56–61 dB(A) for the listed siblings).
- The 'Dynamisches' designation indicates a configuration intended for dynamic temperature changes; the manual warns that very dynamic, high temperature changes recurring at short intervals can cause premature aging/fatigue of the plate heat exchanger (this warning does not apply to models without a plate heat exchanger, but the 645w is not listed among those exceptions).
- Pump is magnetically coupled, stainless steel, with adjustable speed; pump and cooling-water connections are flange DN32 PN10/40 EN1092-1.
- Controller is Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB (Host and Device), RS232; external sensor connection Pt100.
Behavior under load and media
- Temperature range -60...200 °C. Cooling capacity falls sharply at low temperatures: 45 kW at 0 °C down to 6 kW at -60 °C with ethanol, so the process heat load at the lowest operating temperature must be matched to this curve.
- Heating capacity 36 kW.
- Pump delivers max. 343 l/min and 300 l/min at 1.0 bar; system pressure drop must be checked against this curve.
- Maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid choice at low temperatures.
- Filling/expansion volume 30 l.
- Safety class III / FL; for FL-rated (flammable) heat transfer fluid, the fluid must have a fire point ≥ 65 °C; if ethanol is used, only supervised operation is permitted.
- Refrigerant R-507 (A1, H280), GWP 3985.
Installation, connection and control
- Requires 460 V 3~ 60 Hz supply; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
- Three-phase devices drawing more than 63 A are delivered without cable and without plug; external branch circuit protection according to UL 489 is required and must be specified from the electrical schematics.
- Cooling water: 5000 l/h at 15 °C water / 0 °C flow, 3 bar differential pressure, 6 bar pressure; intake temperature must stay between 5 °C and 20 °C; pipes must be insulated if below the condensation point. Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Ambient temperature 5...40 °C; protection rating IP20, so dry protected environment required.
- Alarm output: optic, acoustic, relay.
- Pilot ONE controller supports Process Safety mode (requires compressor set to Permanently ON) and Power Failure Auto (resumes temperature control after power outage if it was active).
- Expansion vessel must remain unlocked for externally closed applications; must be locked for externally open (bath) applications.
- Unit must not be installed or started within an ATEX zone; use with medical devices or direct foodstuff temperature control is not permissible.
Selection & Integration Q&A
Choose the 645w when you need 36 kW heating and up to 45 kW cooling at 0 °C with a 460 V 3~ 60 Hz supply and 5000 l/h cooling water. Smaller siblings suit lower capacities or different supply voltages; verify each sibling's electrical and cooling-water data before substituting.
Provide a 460 V 3~ 60 Hz supply with UL 489 branch protection, cooling water at 5000 l/h and 3–6 bar with 5–20 °C intake, flange DN32 PN10/40 EN1092-1 mating flanges, and a drip tray. Power cable and plug are not included if current exceeds 63 A.
Install in a dry, protected environment at 5–40 °C ambient, IP20, not in an ATEX zone. Level the unit, engage parking brakes, and allow 2 h acclimation after temperature changes. Avoid very dynamic high-temperature changes recurring at short intervals to prevent plate heat exchanger fatigue.
Base delivery excludes power cable/plug for >63 A, UL 489 breaker, mating flanges, drip tray, and heat transfer fluid. Periodic tasks include hose inspection/replacement, fluid inspection/replacement and circuit rinsing, cleaning liquefier fins and strainer, and refrigerant work only by approved contractors.