Peter Huber Kältemaschinenbau Unistat GL P550w Dynamisches

Peter Huber Kältemaschinenbau Unistat GL P550w Dynamisches — Temperature control units.

The Unistat GL P550w Dynamisches is a water-cooled, CO2 (R-744) refrigerated circulating thermostat with a magnetically coupled stainless-steel pump, covering -50 to 200 °C with 24 kW heating. Its high-pressure pump delivers 201 l/min at zero head and 42 l/min at 5.0 bar, distinguishing it from standard-pressure siblings for high-pressure-drop loops. It serves precise temperature control of externally closed industrial processes.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat GL P550w Dynamisches Current —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...200 °C
Dimensions, Weight & Mechanical Design
Weight 799 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 3100 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Site supply must tolerate only ±5% voltage with ±2% frequency

The datasheet permits ±5% voltage only together with ±2% frequency, and explicitly forbids combinations such as -5% voltage with +2% frequency. Verify the plant supply quality before connection; the unit draws up to 61.5 A and requires a 3x63 A fuse.

Unistat GL P550w Technical Data Sheet, page 2
Electrical
Provide and fit the site plug; cable is supplied without one

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug. The customer must provide and fit the site plug, and hard-wired connection must be performed by an electrician with supply matching the rating plate.

Unistat GL P550w Technical Data Sheet, page 2; Unistat Operation Manual, page 37
Cooling & Media
Insulate internal cooling-water pipes if below dew point

When cooling water temperatures fall below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation and corrosion. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Unistat GL P550w Technical Data Sheet, page 2
Process Limits
External loop pressure drop must match the pump curve

The pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar. Size the external circuit so its pressure drop at the required flow stays within the pump curve; otherwise the required flow cannot be achieved. Maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Unistat GL P550w Technical Data Sheet, page 2
Mounting
Transport and store upright only; empty before moving

Only transport the temperature control unit in an upright position; horizontal transport damages the compressor. Only an emptied unit may be moved, as filled transport risks thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck.

Unistat Operation Manual, page 22
Mounting
Allow 2 hours acclimation before first power-on

When moving between cold and warm environments (or vice versa), allow the unit to acclimate for about 2 hours before turning it on to prevent condensation-induced electrical faults. Leveling feet must be unscrewed/activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 29, page 22
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active under exothermic conditions.

Unistat Operation Manual, page 19
Control & I/O
Choose power-failure restart behavior for unattended operation

After a power outage or when switching on, the Pilot ONE can be set to 'Off/Standby' (default, manual start) or 'ON/Temperature control active' (automatic restart with a few-second INFO to disable). Power Failure Auto resumes temperature control automatically if it was active before the outage.

Unistat Operation Manual, page 20-21
Maintenance
Periodically inspect hoses and replace as needed

Hoses and connections must be periodically checked for leaks and quality and replaced as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors; general maintenance personnel cannot service the refrigerant circuit.

Unistat Operation Manual, page 29; Unistat Operation Manual, Section 1.4.2
Environment
Do not install in an ATEX zone; ambient 5–40 °C

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. The installation room must stay within 5–40 °C ambient. EMC Class A Group 1 equipment is intended for industrial environments, not residential.

Unistat Operation Manual, page 15; Unistat GL P550w Technical Data Sheet, page 2

Selection & Integration Q&A

When should I choose the Unistat GL P550w Dynamisches instead of a sibling like the Unistat 520w Dynamisches?

Choose the P550w when the external circuit has high pressure drop and you need CO2 refrigerant (GL) with water cooling (w). Its pump delivers 42 l/min at 5.0 bar. Siblings such as the 520w are air-cooled or standard-pressure alternatives for lower pressure-drop loops.

What are the requirements for the cooling water supply and the electrical connection?

Cooling water: 5–20 °C, 1–4 bar differential, max 6 bar static, G3/4 male connections. Electrical: 400 V 3~ 50 Hz, max 61.5 A, 3x63 A fuse; cable supplied without plug. Voltage tolerance ±5% only with ±2% frequency.

What mounting and ambient limits apply from the manual?

Install upright only, on level ground with leveling feet activated and casters braked. Ambient 5–40 °C, no ATEX zone. Allow 2 h acclimation after temperature changes. Transport only when emptied; casters are not for transport.

What maintenance is required and what accessories are not included in base delivery?

Periodically check hoses and replace as needed; shut off cooling water during downtimes. Refrigerant work by approved contractors. Base delivery excludes the site power plug and a drip tray, which the responsible body must install.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat GL P550w Technical Data Sheet
Datasheet ·98,6 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF