No photo available yet
Peter Huber Kältemaschinenbau Unistat P510w Dynamisches
Peter Huber Kältemaschinenbau Unistat P510w Dynamisches — Temperature control units with Noise level 60 dB.
The Unistat P510w Dynamisches is a water-cooled process temperature control system with a magnetically coupled stainless-steel pump, covering -50 to 250 °C with 6 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from the standard 510w. It serves as a circulation source for externally closed or open process loops requiring deep cooling and dynamic temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P510w Dynamisches Current | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 60 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...250 °C |
| Weight | 182 kg |
| Current consumption | 13 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 355 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled condenser (suffix 'w') with G1/2 male cooling-water connections; requires 1 bar minimum differential pressure and must not exceed 6 bar.
- 'P' variant is intended for applications with high pressure drop; the pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar, so loop pressure drop must be checked against the pump curve.
- Sibling models in the series include the Unistat 510w, 515w, 520w and 525 Dynamisches; this P510w differs by its high-pressure-drop pump configuration.
- Magnetically coupled stainless-steel pump with adjustable speed; maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
- Safety class III / FL; for FL-rated flammable fluids, fire point must be ≥65 °C and ethanol use requires supervised operation.
Behavior under load and media
- Temperature range -50 to 250 °C with 6 kW heating capacity; cooling capacity falls sharply at low temperatures (5.3 kW at 0 °C to 0.25 kW at -50 °C with ethanol), so deep-cooling duty must be sized from the curve, not the nominal figure.
- Cooling water intake must stay between 5 °C and 20 °C; pipes must be insulated if water can drop below dew point.
- Refrigerant R-452A (GWP 2141, CO2 equivalent 3.1 t) imposes F-gas leak-monitoring obligations.
- Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off all poles; this requires the compressor set to Permanently ON.
Installation, control and operation
- Pilot ONE controller with POKO ONE control output; interfaces Ethernet, USB (Host and Device) and RS232; external Pt100 sensor connection; display resolution 0.01 K.
- 400 V 3~ 50 Hz supply with 3x16 A fuse; three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be sourced and fitted on site.
- Leakage current exceeds 3.5 mA, affecting RCD selection and earthing requirements.
- Voltage and frequency tolerances are ±5 % and ±2 % but must not combine in the same direction (e.g. -5 % voltage with +2 % frequency is not allowed).
- IP20 enclosure; ambient 5–40 °C; noise 60 dB(A); weight 182 kg.
- Power-interruption response configurable via Pilot ONE: 'Off/Standby' (default) or 'ON/Temperature control active' with a brief INFO window to disable automatic restart.
Integration & Operating Notes
The unit requires a 400 V three-phase 50 Hz supply protected at 3x16 A. Three-phase devices below 63 A are delivered with cable but without a plug, so the site must source and fit a suitable plug. Leakage current exceeds 3.5 mA, which affects RCD type selection and earthing; voltage and frequency tolerances of ±5 % and ±2 % must not combine in the same direction.
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid choice at low temperatures. For FL-rated flammable fluids, the fire point must be ≥65 °C; if ethanol is used, only supervised operation is permitted. Low level protection via pressure sensor requires thorough venting and correct fill level, otherwise circulation and temperature control are interrupted.
Selection & Integration Q&A
Choose the P510w when the external circuit has high pressure drop: the pump delivers 33 l/min at 2.5 bar versus 119 l/min at zero head. The standard 510w suits low-pressure-drop loops; the 520w offers higher capacity if cooling demand exceeds the P510w's 5.3 kW at 0 °C.
Cooling water: 5–20 °C intake, 1 bar min differential pressure, 6 bar max, G1/2 male connections, 355 l/h at 15 °C. Electrical: 400 V 3~ 50 Hz, 3x16 A fuse, cable supplied without plug; leakage current >3.5 mA affects RCD selection.
Periodic hose/connection inspection, cleaning of the hat-type strainer and liquefier fins, and heat transfer fluid replacement are required; refrigerant work only by approved contractors. Not included: power plug, heat transfer fluid, cooling water hoses, drip tray, and the optional gas detection sensor or processing unit.