No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P520w Dynamisches

Peter Huber Kältemaschinenbau Unistat P520w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat P520w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel circulation pump, built for the -55…250 °C range on a 400 V 3~ 50 Hz supply. Its 'P' designation marks the high-discharge-pressure pump variant (82 l/min at zero head, 3 bar max) intended for external circuits with high pressure drop. It serves flow-through chemistry and semiconductor processes where heat rejection is by cooling water at 5…20 °C.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P520w Dynamisches Current 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A) (sound pressure level)
Control output POKO ONE (potential-free contact)
Expansion volume 16 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 208 kg (net)
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must stay within 1–6 bar and 5–20 °C intake

Below 1 bar differential pressure the specified heat rejection is not available; above 6 bar the circuit is overstressed. Intake water is specified between 5 °C and 20 °C, and if the water can drop below the condensation point the cooling-water pipes inside the unit must be insulated to prevent condensation.

Unistat P520w Technical Data, page 2
Mounting
Transport and store upright only; move only when emptied

Horizontal transport damages the compressor, and a filled unit may overflow thermal fluid during movement. Casters, if present, are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, Commissioning, page 22
Control & I/O
Decide power-failure restart behavior before unattended operation

Pilot ONE can be set to 'Off/Standby' (default, manual start after power returns) or 'ON/Temperature control active' (automatic restart, with a brief INFO window to abort). Power Failure Auto resumes the process automatically if temperature control was active during the outage — acceptable only where unattended restart is safe.

Unistat Operation Manual, Section 1.8.1, page 21
Process Limits
Process Safety mode requires the compressor set to Permanently ON

In Process safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This keeps heat extraction available under exothermic conditions but only works with the compressor configured to Permanently ON, so the controller setup must be verified before relying on this protection.

Unistat Operation Manual, Setup Mode, page 19
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be closed for short downtimes such as overnight to avoid water damage and unnecessary load on the cooling circuit. Hoses and hose connections must be checked periodically for leaks and quality and replaced as required.

Unistat Operation Manual, Commissioning / Maintenance, page 29
Mounting
Not permitted in ATEX zones or for medical/food contact duty

The unit must not be installed or started inside an ATEX zone without an Ex px cabinet, and use with medical devices (e.g. in vitro diagnostics) or for direct foodstuff temperature control is not permissible. The responsible body must also provide a condensation-water/heat-transfer-fluid drip tray below the unit.

Unistat Operation Manual, Section 1.3.3–1.4, pages 15–17

Selection & Integration Q&A

When should I choose the P520w over the non-P Unistat 520w Dynamisches?

Pick the P520w when the external circuit has high pressure drop — flow-through chemistry or semiconductor applications — because its pump delivers 82 l/min at zero head and still 32 l/min at 2.5 bar. For low-resistance loops the standard 520w pump is sufficient and avoids the higher pump stress.

What must I provide on the cooling-water and hydraulic side to commission this unit?

A G1/2 male cooling-water supply at 5–20 °C intake, 1–6 bar differential pressure and about 492 l/h at 15 °C, plus M30x1.5 male pump connections requiring ball socket per DIN 3863 and sleeve nut per DIN 3870. Insulate cooling-water pipes if water can fall below the condensation point.

What site and duty-cycle limits apply to installing the P520w?

Ambient must stay 5–40 °C, the unit must not sit in an ATEX zone, and it must be transported and stored upright and moved only when emptied. Allow about 2 hours acclimation after a cold-to-warm move before power-on, and provide a drip tray below the unit.

What maintenance and consumables are not included in the base delivery?

The site plug, mating hoses and DIN 3863/3870 fittings, heat transfer fluid, and the drip tray are not part of the unit. Periodic tasks include checking and replacing hoses, cleaning liquefier fins and the dirt trap, and refrigerant work by an approved contractor; cooling water should be shut off overnight.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P520w Technical Data
Datasheet ·74,1 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF