Peter Huber Kältemaschinenbau Unistat P530w Dynamisches

Peter Huber Kältemaschinenbau Unistat P530w Dynamisches — Temperature control units with Noise level 62 dB.

The Unistat P530w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pressure pump and Pilot ONE controller, covering -55 to 250 °C with 12 kW heating and up to 21 kW cooling. Its 'P' designation targets applications with high pressure drop, distinguishing it from the 510w/515w/520w/525 siblings. It serves as a dynamic heating/cooling source for externally closed or open circuits in industrial processes.

Noise level
62 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P530w Dynamisches Current 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 62 dB(A)
Control output POKO ONE
Expansion volume 19.9 l
Filling volume 8.2 l
Heating capacity 12 kW
Pump connection M38x1,5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Temperature stability 0.01 K
Alarm message mfr optic, acoustic, relay
Dimensions (W x D x H) mfr 730 x 804 x 1738 mm
Delivery capacity pressure pump max. mfr 201 l/min; 5,3 bar
Pump characteristics mfr Pressure pump, variable speed, magnetically coup
Control input mfr ECS ONE
Dimensions, Weight & Mechanical Design
Weight 439 kg
Electrical, Power & Drive Specifications
Current consumption 31.8 A
Power supply 400V 3~ 50Hz
Fuse mfr 3x32A
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0.01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 21 kW (250 °C)
200 °C: 21 kW · 100 °C: 21 kW · 20 °C: 21 kW · 0 °C: 16 kW · -20 °C: 9 kW · -40 °C: 3 kW · -50 °C: 1 kW
Cooling water connection G3/4 male
Cooling water consumption 830 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior

Control, installation and operation

Integration & Operating Notes

Electrical
Site supply must be sized for 31.8 A and 3x32 A fusing

The 400 V three-phase installation, cable cross-section and protection device must be engineered for 31.8 A consumption with a 32 A fuse per phase. The unit arrives with cable but no plug, so the mating plug and upstream breaker must be procured on site.

Unistat P530w Technical Data, page 2
Process Limits
Cooling water intake must stay between 5 and 20 °C

Operating outside the 5–20 °C cooling water intake range reduces cooling power and promotes condensation and corrosion. Pipes must be insulated where water can fall below the condensation point, and the circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar.

Unistat P530w Technical Data, page 2
Cooling & Media
Flammable R-1270 refrigerant dictates area classification

The refrigerant is R-1270 (A3, H220), so the installation area must respect flammability requirements, ignition-source control and ventilation. The unit must not be installed or started inside an ATEX zone; explosion protection requires a separate Ex px cabinet.

Unistat P530w Technical Data, page 1; Operation Manual, Section 1.3.3
Mounting
Install a drip tray and respect upright transport rules

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit where national legislation requires it. Transport and storage must be upright only, and the unit must be emptied before any move; casters are not for transport and carry only 25 % of total mass.

Operation Manual, Section 1.4 / page 17 and page 22
Control & I/O
Choose power-failure behavior before unattended operation

Pilot ONE can be set to 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart after power return, with a brief INFO window to cancel). Select according to whether an unattended restart is acceptable for the process.

Operation Manual, Section 1.8.1 / page 21
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Operation Manual, page 29

Selection & Integration Q&A

When should I choose the P530w over the 510w, 515w, 520w or 525 Dynamisches?

Pick the P530w when the external circuit has a high pressure drop and you need the larger frame with 21 kW cooling at 20 °C and 12 kW heating. The 510w/515w/520w/525 are smaller siblings; confirm their capacity against your load before downsizing.

What must I provide for the pump circuit and cooling water connections?

Pump ports are M38x1.5 male (bore Y, 60°, DIN 3863), so order matching hoses and fittings. Cooling water uses G3/4 male at 830 l/h, 1–4 bar differential and max 6 bar; verify your loop pressure drop against the pump curve (201 l/min at zero head, 42 l/min at 5 bar).

What maintenance and accessories are not included in the base delivery?

The unit ships with cable but no mains plug; the drip tray, cooling water hoses/fittings and heat-transfer fluid are site-supplied. Periodic tasks include hose inspection/replacement, liquefier fin cleaning, strainer cleaning and fluid inspection/replacement; refrigerant work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P530w Technical Data
Datasheet ·98,7 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF