No photo available yet
Peter Huber Kältemaschinenbau Unistat P530w Dynamisches
Peter Huber Kältemaschinenbau Unistat P530w Dynamisches — Temperature control units with Noise level 62 dB.
The Unistat P530w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pressure pump and Pilot ONE controller, covering -55 to 250 °C with 12 kW heating and up to 21 kW cooling. Its 'P' designation targets applications with high pressure drop, distinguishing it from the 510w/515w/520w/525 siblings. It serves as a dynamic heating/cooling source for externally closed or open circuits in industrial processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P530w Dynamisches Current | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 62 dB(A) |
| Control output | POKO ONE |
| Expansion volume | 19.9 l |
| Filling volume | 8.2 l |
| Heating capacity | 12 kW |
| Pump connection | M38x1,5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Temperature stability | 0.01 K |
| Alarm message mfr | optic, acoustic, relay |
| Dimensions (W x D x H) mfr | 730 x 804 x 1738 mm |
| Delivery capacity pressure pump max. mfr | 201 l/min; 5,3 bar |
| Pump characteristics mfr | Pressure pump, variable speed, magnetically coup |
| Control input mfr | ECS ONE |
| Weight | 439 kg |
| Current consumption | 31.8 A |
| Power supply | 400V 3~ 50Hz |
| Fuse mfr | 3x32A |
| Controller | Pilot ONE |
| Display resolution | 0.01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 21 kW (250 °C) 200 °C: 21 kW · 100 °C: 21 kW · 20 °C: 21 kW · 0 °C: 16 kW · -20 °C: 9 kW · -40 °C: 3 kW · -50 °C: 1 kW |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 830 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing water-cooled unit (suffix 'w'), 730 x 804 x 1738 mm, 439 kg, IP20 enclosure, ambient 5–40 °C.
- 'P' variant is intended for applications with high pressure drop; the pressure pump delivers up to 201 l/min at zero head and 5.3 bar max, with variable speed and magnetic coupling in stainless steel.
- Sibling models 510w, 515w, 520w and 525 Dynamisches share the series architecture but differ in size/performance; the P530w is the larger frame in this set.
- Refrigerant is R-1270 (A3, H220), a flammable natural refrigerant; the cooling circuit is factory-tested leak-tight to 10⁻⁴ Pa·m³/s or better.
Thermal behavior
- Working range -55 to 250 °C, stability 0.01 K, display resolution 0.01 K.
- Cooling capacity is strongly temperature-dependent: 21 kW at 250/200/100/20 °C, 16 kW at 0 °C, 9 kW at -20 °C, 3 kW at -40 °C and 1 kW at -50 °C. Low-temperature duty must be checked against these points.
- Heating capacity 12 kW; heat-transfer fluid charge minimum 8.2 l with 19.9 l expansion volume.
- Cooling water: G3/4 male connection, 830 l/h consumption at 15 °C water/0 °C flow, 1 bar minimum differential pressure, 6 bar maximum, intake 5–20 °C.
Control, installation and operation
- Pilot ONE controller with POKO ONE control output and ECS ONE control input; interfaces Ethernet, USB host/device and RS232; external Pt100 sensor connection.
- Electrical: 400 V 3~ 50 Hz, 31.8 A current consumption, 3x32 A fuse; three-phase units below 63 A are delivered with cable but without a plug.
- Voltage/frequency tolerances ±5 %/±2 % must not be combined in the same direction.
- Pump connection M38x1.5 male (bore shape Y, 60°, DIN 3863); matching hoses/fittings required.
- Safety class III/FL; alarm via optic, acoustic and relay; leakage current >3.5 mA requires verified earthing and RCD selection.
- EMC class A group 1 for industrial environments.
Integration & Operating Notes
The 400 V three-phase installation, cable cross-section and protection device must be engineered for 31.8 A consumption with a 32 A fuse per phase. The unit arrives with cable but no plug, so the mating plug and upstream breaker must be procured on site.
Operating outside the 5–20 °C cooling water intake range reduces cooling power and promotes condensation and corrosion. Pipes must be insulated where water can fall below the condensation point, and the circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar.
The refrigerant is R-1270 (A3, H220), so the installation area must respect flammability requirements, ignition-source control and ventilation. The unit must not be installed or started inside an ATEX zone; explosion protection requires a separate Ex px cabinet.
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit where national legislation requires it. Transport and storage must be upright only, and the unit must be emptied before any move; casters are not for transport and carry only 25 % of total mass.
Pilot ONE can be set to 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart after power return, with a brief INFO window to cancel). Select according to whether an unattended restart is acceptable for the process.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Selection & Integration Q&A
Pick the P530w when the external circuit has a high pressure drop and you need the larger frame with 21 kW cooling at 20 °C and 12 kW heating. The 510w/515w/520w/525 are smaller siblings; confirm their capacity against your load before downsizing.
Pump ports are M38x1.5 male (bore Y, 60°, DIN 3863), so order matching hoses and fittings. Cooling water uses G3/4 male at 830 l/h, 1–4 bar differential and max 6 bar; verify your loop pressure drop against the pump curve (201 l/min at zero head, 42 l/min at 5 bar).
The unit ships with cable but no mains plug; the drip tray, cooling water hoses/fittings and heat-transfer fluid are site-supplied. Periodic tasks include hose inspection/replacement, liquefier fin cleaning, strainer cleaning and fluid inspection/replacement; refrigerant work requires approved contractors.