No photo available yet
Peter Huber Kältemaschinenbau Unistat P615w Dynamisches
Peter Huber Kältemaschinenbau Unistat P615w Dynamisches — Temperature control units with Noise level 62 dB.
The Unistat P615w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -60 to 200 °C. Its 12 kW heating capacity and dynamic pressure-rated pump distinguish it within the Unistat 5xx series for high-pressure-drop loops. It serves as a precise heating/cooling source for externally closed industrial processes requiring three-phase power and cooling water.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P615w Dynamisches Current | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 62 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Weight | 358 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 426 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P615w is a water-cooled (w) free-standing Unistat 5xx model with a 12 kW heater and a magnetically coupled stainless-steel pump, rated for -60 to 200 °C.
- The 'P' suffix denotes suitability for applications with high pressure drop; the pump delivers up to 82 l/min at zero head and 36 l/min at 2.5 bar, with a maximum delivery pressure of 3 bar.
- Compared to sibling models (Unistat 510w, 515w, 520w, 525 Dynamisches), the P615w offers higher heating capacity and a pressure-oriented pump curve; noise level is 62 dB(A).
- Cooling is via a water-cooled condenser; the refrigerant R-452A (A1, H280, GWP 2141) is factory-charged, with 5.5 t CO2 equivalent.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 9.5 kW at 0 °C with ethanol, falling to 1.2 kW at -60 °C. Low-temperature loads must be checked against this curve.
- The pump curve must be evaluated against the external circuit pressure drop; high-impedance loops reduce flow to 36 l/min at 2.5 bar.
- The unit is rated for heat transfer fluids with a fire point ≥65 °C when FL-rated; ethanol is permitted only under supervised operation.
- The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used.
Installation, control and operation
- Electrical supply: 400 V 3~ 50 Hz, max current 25.5 A, fuse 3x32 A. Leakage current exceeds 3.5 mA, requiring appropriate RCD selection. A cable is supplied without a plug.
- Cooling water: G1/2 male connections, 426 l/h at 15 °C water, minimum differential pressure 3 bar, maximum pressure 6 bar; intake temperature 5–20 °C.
- The Pilot ONE controller with POKO ONE control output provides Ethernet, USB (host and device) and RS232 interfaces; display resolution 0.01 K; external Pt100 sensor connection.
- Safety class III/FL; protection rating IP20; ambient temperature 5–40 °C; weight 358 kg; dimensions 630x704x1520 mm.
- The unit must not be installed in ATEX zones; transport upright only; allow 2 h acclimation after temperature changes; expansion vessel must remain unlocked for externally closed applications.
Integration & Operating Notes
The unit requires a 400 V 3~ 50 Hz supply with max current 25.5 A and a 3x32 A fuse. A cable is supplied without a plug, so the site must provide and fit a suitable plug. Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for this higher leakage.
Cooling water differential pressure must be at least 3 bar and must not exceed 6 bar; flow is about 426 l/h at 15 °C water. Intake temperature must stay between 5 °C and 20 °C. Pipes must be insulated if water can drop below the condensation point. Shut off cooling water even for short downtimes.
Transport and store the unit upright only; horizontal transport damages the compressor. Only move an emptied unit. Casters are not for transport and carry only 25% of total mass; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Incorrect setting can damage the application or cause fluid loss.
In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Choose power-interruption behavior (Off/Standby or ON/Temperature control active) according to whether unattended restart is acceptable.
Check periodically for leaks and the quality of hoses and hose connections and replace as required. Heat transfer fluid must be inspected and replaced per manual; circuit rinsing may be needed. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Choose the P615w when your application has a high pressure drop and requires the higher heating capacity (12 kW) and pressure-oriented pump (36 l/min at 2.5 bar). The 510w and 520w have lower heating capacities and different pump curves; select them for lower heat loads or lower impedance loops.
Cooling water: G1/2 male connections, 426 l/h at 15 °C, min differential pressure 3 bar, max pressure 6 bar, intake 5–20 °C. Electrical: 400 V 3~ 50 Hz, max 25.5 A, 3x32 A fuse, cable without plug; leakage current >3.5 mA requires suitable RCD.
The base delivery excludes a mains plug (cable supplied without plug), cooling water hoses and fittings (M30x1.5 male pump connections require matching ball-socket fittings per DIN 3870), heat transfer fluid, and a drip tray (must be installed by the responsible body). Periodic hose replacement and fluid inspection are required.