No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P650w Dynamisches

Peter Huber Kältemaschinenbau Unistat P650w Dynamisches — Temperature control units.

The Unistat P650w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and DN32 PN10/40 flanges, covering -60 to 200 °C with 48 kW heating. It is the high-capacity water-cooled modification in the Unistat 5xx/6xx series, distinguished by its 100 A three-phase supply and 5000 l/h cooling-water demand. It serves as a central temperature source for large externally closed or open processes requiring dynamic temperature control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat P650w Dynamisches Current —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Expansion volume 28 l
Filling volume 28 l (min)
Heating capacity 48 kW
Pump connection flange DN32 PN10/40 EN1092-1
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 2500 kg (net)
Electrical, Power & Drive Specifications
Current consumption 100 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection flange DN32 PN10/40 EN1092-1
Cooling water consumption 5000 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Control, installation and operation

Integration & Operating Notes

Electrical
Supply must be sized for 100 A with 3x125 A fusing

The 400 V 3~ 50 Hz supply must deliver up to 100 A and be protected with 3x125 A fuses; voltage tolerance is ±5% with simultaneous frequency tolerance ±2%, but not in the unfavorable combination. Because current exceeds 63 A, the unit arrives without cable and plug, so the installer must provide a hard-wired connection by an electrician.

Datasheet, Specifications; Operation Manual, Preparations for Operation and Application/Power Connection
Cooling & Media
Cooling water must be 5–20 °C at 3–6 bar and ~5000 l/h

The facility water system must supply 5–20 °C water at 3–6 bar differential pressure and about 5000 l/h at 15 °C. Higher cooling-water temperature reduces cooling power and may require increased flow, so design the water supply with margin. Insulate cooling-water pipes when water temperature falls below the condensation point.

Datasheet, Specifications and Installation notes
Mounting
Transport upright and empty; casters are not for transport

The 2500 kg unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Operation Manual, Commissioning: transport, positioning, ambient and installation conditions
Process Limits
Expansion vessel must stay unlocked for closed applications

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. Verify the correct state before filling and venting.

Operation Manual, Preparations for Operation
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this requires the compressor to be set to Permanently ON. Configure the Pilot ONE accordingly when exothermic reactions or glass breakage risk must be mitigated.

Operation Manual, Setup Mode: overtemperature protection
Maintenance
Shut off cooling water even for overnight downtimes

Shut off the cooling water supply to the unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling circuit. Periodically check hoses and connections for leaks and quality and replace as required. Refrigerant-circuit work must be carried out by approved refrigeration/air-conditioning contractors.

Operation Manual, Commissioning and Maintenance

Selection & Integration Q&A

When should I choose the Unistat P650w Dynamisches instead of an air-cooled sibling such as the Unistat 510w or 520w Dynamisches?

Choose the P650w when a large cooling-water supply (5–20 °C, 3–6 bar, ~5000 l/h) and a 400 V/100 A feed are available and high heating capacity (48 kW) over -60 to 200 °C is needed. Air-cooled siblings avoid water infrastructure but differ in capacity and cooling concept.

What adjacent system parts must be provided for the P650w, and what are the interface requirements?

Provide DN32 PN10/40 EN1092-1 mating flanges and gaskets for both pump and cooling-water connections, a hard-wired 400 V 3~ 50 Hz supply with 3x125 A fusing (no cable or plug supplied above 63 A), and a cooling-water system delivering 5–20 °C at 3–6 bar and ~5000 l/h. Account for leakage current >3.5 mA in earthing and RCD selection.

What mounting and environmental limits apply to the P650w?

Install indoors, not in an ATEX zone, with 5–40 °C ambient and IP20 protection. Transport upright and empty; use an industrial truck, not casters. Activate leveling feet and parking brakes. Allow ~2 h acclimation after moving between cold and warm environments before power-on.

What maintenance and consumables are required, and what is not included in base delivery?

Base delivery excludes power cable and plug (above 63 A), mating flanges/gaskets, and heat transfer fluid. Plan periodic hose/connection inspection and replacement, cooling-water shutoff for overnight downtimes, cleaning of liquefier fins and hat-type strainer, and refrigerant service by approved contractors (R-507, 26.5 kg).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P650w Technical Data
Datasheet ·139,2 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF