Peter Huber Kältemaschinenbau Unichiller 075w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 075w

Peter Huber Kältemaschinenbau Unichiller 075w — Recirculating chillers.

The Unichiller 075w is a water-cooled recirculating chiller with a 48 l/min pump and a -20…40 °C working range, distinguished within the series by its water-cooled refrigeration circuit and G1/2 cooling-water connection. It serves externally closed process loops requiring stable temperature control without air-side heat rejection.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 075w Current —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Maximum delivery flow rate 48 l/min (also 3, 4 bar pump pressure)
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 300 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 240 l/h (cooling water 15°C, flow -20°C)
cooling water 15°C, flow -10°C: 270 l/h · cooling water 15°C, flow 0°C: 330 l/h
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The Unichiller 075w is a water-cooled refrigerating machine (CFC- and HCFC-free, refrigerant R-449A) built for externally closed applications only. Unlike air-cooled siblings such as the 030T/030Tw variants, this model rejects heat to a cooling-water circuit via G1/2 male connections, so it needs no large air-side clearance but does need a cooling-water supply. The pump is a fixed-speed E1 type (not speed-adjustable) with G1 1/4 AG male connections, delivering up to 48 l/min at a maximum pressure of 3.4 bar. The 48 l expansion volume and 18 l minimum fill define the hydraulic buffer available to the loop.

Behavior under load and media

Cooling-water consumption depends on the required flow temperature at a 15 °C cooling-water intake: 270 l/h at -10 °C, 330 l/h at 0 °C, and 240 l/h at -20 °C. Cooling power falls and cooling-water flow demand may rise as the cooling-water temperature increases. The cooling-water circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) constrain acceptable water quality. Pump flow decreases as external pressure drop rises, so the external circuit must be sized against the 3.4 bar maximum.

Installation, control and operation

The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached rather than switching off all poles. The unit is IP20, safety class I/NFL, and must not be installed in an ATEX zone.

Integration & Operating Notes

Cooling & Media
Cooling water must stay within 5–20 °C intake and 0.5–6 bar

The chiller is designed for cooling-water intake between 5 °C and 20 °C with a minimum differential pressure of 0.5 bar and a maximum of 6 bar. If intake is below the condensation point, internal cooling-water pipes must be insulated to prevent condensate.

Unichiller 075w — Technical Data Sheet, page 2
Mounting
Allow 50–200 cm wall clearance per side for maintenance

Free-standing models require a wall clearance of 50 to 200 cm on either side to permit maintenance access. The unit must be used indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m.

Unichiller Operation Manual, page 25
Electrical
Provide a plug: three-phase units under 63 A ship without one

The 400 V 3~ 50 Hz supply must respect ±5% voltage with simultaneous ±2% frequency tolerance, but the combination of -5% voltage and +2% frequency is not permitted. Three-phase devices drawing less than 63 A are supplied with a cable but no plug, so the installer must provide a suitable plug and connection.

Unichiller 075w — Technical Data Sheet, page 2
Process Limits
Only externally closed loops are permitted

The unit is designed exclusively for externally closed applications; open baths or vented circuits are outside its intended use. Models with 'P' in the model name are the ones suited to high-pressure-drop applications, so verify loop pressure drop against the 3.4 bar pump maximum.

Unichiller Operation Manual, page 34
Operation
Shut off cooling water even for overnight downtime

The cooling-water supply must be closed even for short downtimes such as overnight to prevent water damage and corrosion. Before startup, activate parking brakes on casters and/or unscrew leveling feet so the unit cannot move during operation.

Unichiller Operation Manual, pages 28 and 31
Safety
Select Process Safety for exothermic reactions

In Process Safety mode the controller releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles. Choose this mode where cutting cooling would cause runaway temperatures; otherwise the power-failure response (OFF/Standby, ON, or Power Failure Auto) governs restart behavior.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work requires certified contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a GWP of 1397 and a CO2 equivalent of 1.3 t, which must be considered for leak reporting and environmental compliance.

Unichiller Operation Manual, page 17; Technical Data Sheet, page 1

Selection & Integration Q&A

When should I choose the Unichiller 075w instead of an air-cooled sibling like the 030T or 030Tw?

Pick the 075w when a cooling-water circuit is available and air-side heat rejection or fan noise is undesirable; it is water-cooled with G1/2 connections. Choose an air-cooled 030T/030Tw when no cooling water is available, accepting their 69 dB(A) or 61 dB(A) noise levels.

What cooling-water supply and external circuit must be provided for this chiller?

Provide cooling water at 5–20 °C intake, 0.5 bar minimum differential and 6 bar maximum, through G1/2 male connections. Size the external loop against the pump's 48 l/min and 3.4 bar maximum, and use water compatible with copper, 1.4401 stainless, PA, PPE, PTFE and EPDM.

What installation and duty-cycle limits does the manual impose on the 075w?

Install indoors only, at ≥300 lx and up to 2000 m altitude, with 50–200 cm wall clearance per side for maintenance. Ambient must stay within 5–40 °C. Shut off cooling water even overnight, and never install in an ATEX zone.

What is not included in base delivery and what service access is required?

Three-phase units under 63 A ship with a cable but no plug, so the installer supplies the plug. Refrigerant-circuit work is restricted to approved refrigeration contractors. The 18 l minimum fill and 48 l expansion volume must be respected; no accessories or consumables are listed in the verified facts.