Peter Huber Kältemaschinenbau Unichiller 100T-H
Peter Huber Kältemaschinenbau Unichiller 100T-H — Recirculating chillers.
The Unichiller 100T-H is an air-cooled, CFC/HCFC-free recirculating chiller with a Pilot ONE controller, rated for -20 to 100 °C and 9 kW cooling at 0 °C. It is the high-capacity, non-condenser ('-H') variant in the Unichiller series, distinguished by its 15 kW peak at 15 °C and 3 kW at -20 °C. It serves externally closed process loops requiring stable temperature control with G1 1/4 male pump connections.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 100T-H Current | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Expansion volume | 7 l |
| Heating capacity | 2 kW |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 284 kg |
| Current consumption | 16 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 15 °C) at 0 °C: 9 kW · at -20 °C: 3 kW |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The 100T-H is a free-standing, air-cooled refrigeration machine with a 600x790x1614 mm footprint and 284 kg net weight; the '-H' suffix denotes the high-capacity configuration within the Unichiller series.
- Unlike sibling models such as the 030T (69 dB(A)) and 030Tw (61 dB(A)), this unit is specified for higher cooling duty rather than low-noise operation; no noise figure is stated for this model.
- The hydraulic circuit uses a fixed-speed C3 pump with G1 1/4 male connections and a 7 l expansion volume; minimum filling capacity is 8.36 l.
- The unit is classified to DIN 12876 with safety class III/FL and adjustable overtemperature protection.
Thermal behavior and capacity
- Cooling capacity is strongly temperature-dependent: 15 kW at 15 °C, 9 kW at 0 °C, and 3 kW at -20 °C. Selection must use the duty-point capacity, not the headline figure.
- Rated output data are valid at 20 °C room temperature; capacity may drop as ambient temperature rises toward the 40 °C limit.
- A 2 kW heater extends the upper range to 100 °C, enabling both heating and cooling from one unit.
- The pump delivers 96 l/min at zero head and 24 l/min at 5.0 bar (max delivery pressure 5.6 bar), so external circuit pressure drop must be matched to the pump curve.
Control, installation and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection.
- Power-failure response is configurable (OFF/Standby, ON, Power Failure Auto); Process Safety mode releases maximum cooling when the overtemperature trigger is reached.
- The unit requires a 400 V 3~ 50 Hz supply, 16 A current consumption, and 3x25 A fusing; it is delivered with a cable but without a plug.
- IP20 rating and EMC class A group 1 restrict installation to dry, industrial indoor locations with at least 300 lx illuminance and below 2000 m altitude.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow servicing of the refrigeration circuit and condenser. The 284 kg weight and 600x790x1614 mm dimensions also demand verification of floor loading and doorway clearance before siting.
The 400 V 3~ 50 Hz supply must be sized for 16 A current consumption with 3x25 A fusing. Three-phase devices under 63 A are delivered with a cable but without a plug, so the customer must procure and fit a suitable plug. Voltage and frequency tolerances are only valid in combination; -5% voltage with +2% frequency is not allowed.
The unit is designed exclusively for externally closed applications. The external loop pressure drop must be matched to the pump curve (96 l/min at 0 bar, 24 l/min at 5.0 bar) to avoid starving flow at the duty point. Minimum filling capacity is 8.36 l including the 7 l expansion volume.
The refrigerant is R-449A (GHS H280, GWP 1397, CO2 equivalent 5.6 t). Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning contractors. The air-cooled condenser requires unobstructed airflow and periodic cleaning to avoid capacity loss and high-pressure trips.
Selection & Integration Q&A
Choose the 100T-H when the duty point requires up to 15 kW at 15 °C or 9 kW at 0 °C in a high-capacity configuration. The 030T/030Tw are lower-capacity, low-noise variants (61–69 dB(A)); select those when acoustic limits dominate and cooling demand is lower.
External hoses and adapters must match G1 1/4 male pump connections, and the loop pressure drop must align with the pump curve (96 l/min at 0 bar, 24 l/min at 5.0 bar). Electrically, provide a 400 V 3~ 50 Hz supply with 3x25 A fusing and a site-fitted plug; a Pt100 sensor can be connected for external control.
Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, below 2000 m altitude, and outside ATEX zones. IP20 requires a dry, protected location. Maintain 50–200 cm side clearance for maintenance and activate leveling feet or parking brakes before operation.
The air-cooled condenser needs periodic cleaning; refrigerant circuit work requires certified contractors due to R-449A (GWP 1397). The unit ships with a cable but without a plug, and heat transfer fluid is not included—charge at least 8.36 l. No accessories or cables beyond the power cable are stated in the base delivery.