Peter Huber Kältemaschinenbau Unichiller GL 100w
Peter Huber Kältemaschinenbau Unichiller GL 100w — Recirculating chillers.
The Unichiller GL 100w is a water-cooled recirculating chiller with a 20 kW cooling capacity at 10 °C and a -20…40 °C temperature range, controlled by the Pilot ONE. It is designed for externally closed applications and requires a 400 V three-phase supply and cooling water at 1–6 bar differential pressure. This modification is distinguished within the series by its high capacity and water-cooled condenser, making it suitable for large heat loads in industrial processes.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller GL 100w Current | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Expansion volume | 10 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure pump, speed not variable |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Current consumption | 7 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 20 kW (at 10 °C) |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and positioning
- The GL 100w is a water-cooled chiller using R-744 refrigerant (A1, H280) and a pressure pump with fixed speed. It is intended for externally closed applications only.
- Compared to sibling models such as the Unichiller 030T, the GL 100w offers significantly higher cooling capacity (20 kW at 10 °C) and a wider temperature range down to -20 °C. The water-cooled design eliminates the need for a large air-cooled condenser, but requires a cooling water supply.
- The unit is equipped with the Pilot ONE controller, providing Ethernet, USB (Host and Device), and RS232 interfaces for integration into monitoring or control systems.
Behavior under load and media
- Cooling capacity is rated at 20 kW at 10 °C, but this value is valid only under reference conditions: 20 °C room temperature, 15 °C cooling water inlet, and 1 bar differential pressure. Higher cooling water temperatures reduce capacity and may require increased flow.
- The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so water quality must be compatible to avoid corrosion.
- The unit operates with a fixed-speed pressure pump; pump speed is not adjustable. The pump connection is G1 1/4 AG male, and the cooling water connections are G1/2 male.
Installation, connection, and control
- Electrical supply: 400 V 3~ 50 Hz, max current 7 A, with a 3x16 A fuse required. The unit is delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%.
- Cooling water must be supplied at 1–6 bar differential pressure and 5–20 °C inlet temperature. Pipes must be insulated if below the condensation point. Electrical lines directly below the unit must be protected against liquid ingress.
- The Pilot ONE controller allows configuration of power-failure response (OFF/Standby, ON, or Power Failure Auto) and includes a Process Safety mode that releases maximum cooling capacity when the overtemperature trigger is reached.
- The unit must be installed indoors with at least 300 lx illuminance and up to 2000 m altitude. Wall clearance of 50–200 cm per side is required for maintenance. Leveling feet must be activated and parking brakes applied before operation.
Integration & Operating Notes
The chiller requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse. The unit is delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%; certain combinations are not allowed.
The cooling water inlet temperature must be kept between 5 °C and 20 °C. Higher temperatures reduce cooling capacity and may require increased flow. Pipes must be insulated if below the condensation point. Cooling water differential pressure must be 1–6 bar.
A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit must be installed indoors with at least 300 lx illuminance and up to 2000 m altitude. Leveling feet must be activated and parking brakes applied before operation.
The unit is designed exclusively for externally closed applications. It is not explosion-protected and must not be installed or started up inside an ATEX zone. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.
Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The unit uses R-744 refrigerant (A1, H280), requiring appropriate handling precautions. Cooling water supply must be shut off even for short downtimes such as overnight.
The Pilot ONE controller allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Selection & Integration Q&A
Choose the GL 100w when you need high cooling capacity (20 kW at 10 °C) and a water-cooled condenser, and when a cooling water supply is available. The 030T is air-cooled and has lower capacity, making it suitable for smaller loads or where cooling water is not available.
Cooling water must be supplied at 1–6 bar differential pressure, 5–20 °C inlet, with G1/2 male connections. Electrical supply is 400 V 3~ 50 Hz with a 3x16 A fuse; the unit comes with a cable but no plug, so a suitable plug must be provided. Voltage tolerance is ±5% with frequency tolerance ±2%.
Install indoors only, with at least 300 lx illuminance and up to 2000 m altitude. Maintain 50–200 cm wall clearance per side for maintenance. Ambient temperature must be 5–40 °C. The unit is not explosion-protected and must not be installed in an ATEX zone.
Refrigerant circuit work must be performed by approved refrigeration contractors. Cooling water supply must be shut off during downtimes. The base delivery includes a cable but no plug; a suitable plug must be provided on site. No other accessories or consumables are specified in the provided documents.